Introduction

Although cementless sockets seem to be the primary option for young patients, there are not many reports presenting long-term survival data for cementless sockets in patients under 50 years [15]. Nearly all describe long-term results using rigid metal-backed acetabular systems.

In 1967 Robert Mathys Sr. developed the uncoated cementless RM elastic monoblock acetabular component (Fig. 1). This socket was based on the philosophy that the elastic modulus of the polyethylene RM socket (approximately 1,000 N/mm2), in contrast to rigid metal shells (approximately 105.00 N/mm2), mimics the elastic properties of acetabular bone (approximately 500–600 N/mm2). The resulting physiological distribution of articular forces protects the acetabular bone and provides optimal conditions for ingrowth, and subsequent long-term component fixation. The purpose of this retrospective study was to evaluate the effect of the cementless elastic monoblock socket on acetabular osteolysis and long-term survival for aseptic loosening in young patients.

Fig. 1
figure 1

Cementless Robert Mathys elastic monoblock socket

Materials and methods

During the period from 1990 to 1997, 158 hip arthroplasties were performed on 131 patients less than 50 years old. The mean age was 42.4 years (range 16–50); 67 were female and 64 were male. The indications for hip arthroplasty in this cohort are presented in Table 1.

Table 1 Preoperative diagnosis

A posterolateral approach was used in all patients, with direct full weight-bearing while allowing crutches during the first six weeks after surgery. The cementless RM monoblock (Mathys Ltd., Bettlach, Switzerland) socket was used in all patients. This monoblock socket is made from ultrahigh molecular weight polyethylene (UHMWPE) sterilised by gamma irradiation in air (Chirulen, ISO 5834/2 from GUR 1120, MediTECH, Vreden, Germany) with a heat pressed titanium coating using two anchoring pegs and additional screw fixation for primary stability. In 99 hips a CLS Spotorno (Zimmer Ltd., Warsaw, IN, USA) femoral stem was used, 38 hips received an isoelastic RM (Mathys Ltd., Bettlach, Switzerland) stem, 16 hips a Wagner SL stem (Zimmer Ltd., Warsaw, IN, USA) and in five hips a Wagner cone stem (Zimmer Ltd., Warsaw, IN, USA) was implanted. All femoral heads were 28 mm with metal on PE articulation in 58 hips and ceramic on PE in 100 hips. All patients received perioperative antibiotics and deep vein thrombosis prophylaxis.

At the latest outpatient visit, all patients were clinically evaluated using the Harris hip score (HHS) and the Merle D’Aubigné/Postel questionnaires (Fig. 2). We defined an excellent to good clinical score as HHS of 100–80 and Merle D’Aubigné/Postel of 18–14.

Fig. 2
figure 2

Study inclusions

Radiographic evaluation

Observations and measurements were based on standardised anteroposterior pelvic and lateral hip radiographs made early in the postoperative period and at the latest follow-up visit. Correction of magnification was attained for all measurements using the femoral head. All radiographs were evaluated by two authors (DP, JB) and disagreements were resolved through consensus.

Socket migration was evaluated using the method described by Massin et al. [6]. Acetabular component migration of >3 mm in the vertical or horizontal plane and change of cup inclination of >8° were classified as radiographically loose [7]. Wear was measured using methods described by Kang et al. [8]. Acetabular osteolysis was evaluated using the acetabular zones described by DeLee and Charnley [9]. It was deemed significant when progressive, measuring >2 mm and occupying more than 50% of the acetabular zones. Heterotopic ossification (HO) was graded according to Brooker et al. [10].

Statistical analysis

We classified failure as revision of the acetabular component due to septic or aseptic loosening and PE wear. Survival analysis was calculated using the Kaplan-Meier method. We performed both best and worst case scenario analysis during the survival analysis. A logistic regression model was performed to evaluate the effect of patient age, gender, stem type, diagnosis, previous surgery and acetabular inclination on acetabular revision. Cox’s proportional hazards model was used to examine the survival rates for different patient and component factors.

Results

The mean follow-up period was 13.2 years (range ten to 18). During this period, four patients (five hips) died without any relationship to the performed arthroplasty. Clinical and radiographic analysis showed no complications of their arthroplasty. Four patients were lost to follow-up, mainly due to emigration.

Acetabular survivorship analysis

During the study inclusion period 158 hip arthroplasties were performed in 131 patients. Acetabular revision had been performed on 20 patients by the latest follow-up. During our latest follow-up two patients were scheduled for acetabular revision. Our survival analysis was performed on the basis of 22 (14%) acetabular revisions. The reasons for revision and additional data are listed in Table 2.

Table 2 Revision summary

At ten years four patients (3%) had undergone acetabular revision. At the time of the latest follow-up two patients (1%) had undergone revision for aseptic loosening, two patients (1%) for malposition and four patients (3%) for trauma sequelae. In seven patients (4%) acetabular wear was the primary reason for revision. During seven (4%) femoral revisions the surgeon decided perioperatively to revise the acetabular component for minor to severe wear of the socket.

A worst case survival analysis showed 98% survival (95% confidence interval: 95–100) at ten years and 80% survival (95% confidence interval: 72–89) at 14 years (Fig. 3). Survival for aseptic loosening of the RM cup at ten and 14 years was 99% (95% confidence interval: 98–100) and 98% (95% confidence interval: 96–100) (Fig. 4). Survival analysis for wear showed 99% survival (95% confidence interval: 98–100) at ten years and 86% survival (95% confidence interval: 78–94) at 14 years follow-up. Wear analysis for the two featured articulations showed 80% survival (95% confidence interval: 68–91) for the metal on PE articulations and 95% survival (95% confidence interval: 87–100) for the ceramic on PE articulations at 14 years (Fig. 5).

Fig. 3
figure 3

Kaplan-Meier curve of the overall survival of the RM cup

Fig. 4
figure 4

Kaplan-Meier curve of survival rate for aseptic loosening of the RM cup

Fig. 5
figure 5

Kaplan-Meier curve for wear of the RM cup and articulation type

Log-rank analysis for type of articulation and revision showed a statistical difference (p = 0.009) in favour of the ceramic on PE articulations. The Cox proportional hazards regression analysis showed a 6.8 (1.5–30.5) times higher odds ratio for revision in a metal on PE articulation. A logistic regression model showed no effect on revision for factors including age, stem type, gender, diagnosis, previous surgery and acetabular inclination.

Femoral revision

During the follow-up period, 24 patients (15%) underwent femoral revisions. Sixteen RM stems, six CLS stems and two Wagner stems had been revised prior to the follow-up. During seven of these procedures, acetabular revision was simultaneously performed due to wear of the monoblock RM system. All patients who underwent exclusively a femoral revision remained in the survival analysis for the RM acetabular component.

Clinical results

Of the remaining 130 hip arthroplasties, 102 (79%) had excellent to good HHS. At the latest clinical evaluation, nine patients (7%) showed fair and 19 (15%) had poor results. Only 11 patients (9%) showed poor clinical results using the Merle D’Aubigné/Postel questionnaire. All other patients achieved excellent to good scores on the Merle D’Aubigné/Postel questionnaire.

Radiographic results

Pelvic radiographs were available for all patients who were not lost to follow-up and who were still alive. In Table 3 radiographic parameters are shown comparing the revised group with the non-revised group. At the final follow-up three sockets showed a significant advancement in inclination and also showed significant horizontal and vertical migration. Two patients showed only a significant vertical migration.

Table 3 Radiological evaluation

Two patients were considered as outliers for horizontal and vertical migration (5.0 mm, 10.5 mm and 10.5 mm, 32.5 mm, respectively). The first patient showed excellent clinical scores with significant wear and refused revision due to lack of symptoms. The second patient had a poor clinical score, showed significant wear but refused revision. We analysed the migration patterns for the two articulations used in our study. All reported outliers mentioned above were patients with a ceramic on PE articulation. We found no statistical difference for either horizontal (mean: 1.5 mm; 1.9 mm), vertical (mean: 1.8 mm; 2.3 mm) and inclination (mean: 2.3°; 2.8°) migration for, respectively, the metal on PE and ceramic on PE articulations. Heterotopic ossification, HO was seen in 46 patients (grade 1 = 18, grade 2 = 11, grade 3 = 15, grade 4 = 2). Significant progressive acetabular osteolysis was seen in four patients (3%), mainly cavitary osteolysis found in DeLee and Charnley zone 1 adjacent to the fixation screws.

Complications

In our cohort seven arthroplasties (4%) had one or more dislocations, two of which required simultaneous acetabular and femoral revision and in three cases in isolated femoral stem revisions. There were three infections for which one patient underwent operative débridement. Femoral fractures were seen in six cases: three during implantation and three during the follow-up period. One patient showed sciatic nerve irritation due to a postoperative haematoma; the symptoms resolved following conservative treatment. Surgical treatment was indicated for two patients suffering from grade 4 HO. Two patients had postoperative urinary tract infections which were treated with oral antibiotics.

Discussion

Total hip arthroplasty (THA) is a very successful orthopaedic procedure [11], but there is no consensus in defining the optimal socket fixation method, especially for young patients [12]. Although cementless sockets are often advocated in young patients, patients younger than 50 years are more susceptible to wear, osteolysis, implant loosening and failure because of their increased activity level [1315]. Publications specifically investigating long-term results for a homogeneous group of cementless sockets in patients under 50 years are scarce. McLaughlin and Lee reported a 56% revision rate at ten years using the T-Tap socket in patients aged 50 years or less [1]. Comparable results were published by Utting et al. who reported 55% impending Harris-Galante I revisions at a mean follow-up of 13.6 years for patients with a mean age of 40 years [4]. The same socket was used by Crowther and Lachiewicz and Duffy et al., both with better survival rates of 98 and 88%, respectively, at ten years in young patients [2, 3]. Kim et al. implanted 102 Duraloc Option acetabular components in 73 patients with a mean age of 38 years [5]. They reported a cumulative acetabular survival of 99% at 11 years.

All these publications, however, concern long-term results of modular acetabular systems.

In a review of the Finnish Arthroplasty Register in 2008, Mäkelä et al. raised concerns about wear-related revisions for modular cementless sockets [16]. The authors emphasised this problem because of the high proportion of reported liner/wear-related problems. In 2002 Young et al. reported a comparative study between modular and monoblock systems, both using PE sterilised by gamma irradiation in air [17]. In that study the monoblock system demonstrated a lower mean true wear rate and significantly (p = 0.01) less osteolysis. Liner-shell conformity, optimisation of clearance, increased polyethylene thickness, absence of a locking mechanism and no liner-shell micromotions (backside wear) are factors in favour of monoblock systems.

The PE that had been gamma irradiated in air, which we used in our study, showed a mean articular wear of 0.11 mm/year. These values correspond well with articular wear in the literature on cementless modular sockets (0.08 and 0.18) [18, 19], monoblock sockets (0.05–0.17) [20, 21] and even cemented sockets in young patients (0.06 and 0.12) [22, 23]. In our study, approximately one third of the articulations used were metal on PE. In accordance with other authors [21, 24, 25], this resulted in an increased risk for revision for PE wear with a survival of 80% (95% confidence interval: 68–91) at 14 years compared with a superior long-term survival rate of 95% (95% confidence interval: 86.8–100) for a ceramic on PE articulation. The ceramic on PE articulations showed a non-significantly higher mean migration rate mainly due to the several outliers found in our analysis.

Seven sockets were revised during femoral revision surgery. Regional acetabular PE wear was estimated to range from minor to severe during femoral revision surgery although during preoperative planning there was no intention to perform an acetabular revision. The decision to revise the acetabular component was simplified due to the relative ease with which the socket can be revised. Nearly all known monoblock sockets are made by fixing a PE liner into a rigid metal shell during fabrication. In cases of an impending revision, the all poly construction of the RM socket has the advantage over these metal-shelled monoblock sockets due to its relatively easy revision method [26, 27]. Using an acetabular reamer and sufficient irrigation, the socket can be removed with the least amount of acetabular bone stock damage.

Almost all instances of progressive periacetabular osteolysis seen on the pelvic radiographs were cavitary osteolytic lesions mainly found in DeLee and Charnley zone 1. This could be explained by the adjacent screw which can act as a pathway for PE particles.

The osteolysis rates found in our study of the RM socket compare favourably to data reported on cementless acetabular components [17, 28] and are comparable to the known low osteolysis rates for monoblock sockets [21, 29].

The concept of a coated elastic monoblock socket has remained unique. The advantage of this concept is the osseointegration potential of the titanium coating which does not affect the elastic properties of the socket. The elastic modulus of the socket permits transmission of physiological articular stresses and thereby reduces acetabular stress shielding and the development of acetabular osteolysis.

In 2008 Ihle et al. reported the longest follow-up for an RM socket [24]. They presented data on 93 consecutive RM sockets at a mean follow-up of 19 years. Cumulative acetabular survival analysis for any reason showed an excellent survival rate of 83% (95% confidence interval: 73–90) at 20 years. Regression analysis showed a fourfold risk for acetabular revision in younger patients.

A worst case survival analysis of the 158 hip arthroplasties reported in our study showed the somewhat lower survival of 80% (95% confidence interval: 72–89) at 14 years. The patients presented in our study, however, had a lower mean age (42 vs 52) and there were more metal on PE articulations which resulted in more wear and revisions.

This study presents long-term follow-up data relating to a cementless elastic monoblock socket in a large cohort of young patients with a nearly complete clinical and radiological follow-up. A drawback of this study is the retrospective design, two different articulation types and the different stem types used which both interfere with determining the survival of the RM socket alone. Although this study has limitations, our report shows good clinical outcome, low osteolysis rates and excellent long-term survival rates for aseptic loosening of a cementless elastic socket in young patients.

Ongoing tribology developments and knowledge concerning acetabular bone adaptations behind acetabular implants will further lower wear and osteolysis rates and could optimise survival rates of cementless sockets.