Abstract
The centre of rotation of the hip can be displaced in hip dysplasia and revision arthroplasty. This study examined the effect of artificial femoral head load after acetabular component displacement in total hip arthroplasty. Sixteen total hip arthroplasty models of human cadaver specimens were reconstructed, and under different acetabular component position, the load around the femoral head was evaluated by strain gages. The results showed that the load was higher in the same specimens when the cup was moved 2 mm inward or upward, especially after the cup was moved more than 6 mm, and the load had an increasing effect in the inward group. In the upward group, an increasing effect happened at 8 mm upward displacement, but the stress value decreased from 4 mm to 6 mm upward displacement. In the same moving distance, the stress of inward displacement is obviously higher than upward displacement. Altogether, the results suggested that for both inward displacement and upward displacement of the acetabular cup, the load around the femoral head increased gradually, while the distance of the inward displacement and the superior displacement was increased. The greater the displacement, the bigger the loading contact stress. The upward displacement caused less stress change on the femoral head. The stress of the 6 mm upward position was lower than nearby positions; perhaps this site represented a stress buffering zone.
Similar content being viewed by others
References
Udomkiat P, Dorr LD, Wan Z (2002) Cementless hemispheric porous-coated sockets implanted with pressfit technique without screws: average ten-year follow- up. J Bone Joint Surg Am 84:1195–1200
Patil S, Bergula A, Chen PC, Colwell CW Jr, D'Lima DD (2003) Polyethylene wear and acetabular component orientation. J Bone Joint Surg Am 85:56–63
Wan Z, Boutary M, Dorr LD (2008) The influence of acetabular component position on wear in total hip arthroplasty. J Arthroplasty 23:51–56
Witzel U, Rieger W, Effenberger H (2008) Three-dimensional stress analysis of threaded cups—a finite element analysis. Int Orthop 32:195–201
D'Lima DD, Urquhart AG, Buehler KO, Walker RH, Colwell CW Jr (2000) The effect of the orientation of the acetabular and femoral components on the range of motion of the hip at different head-neck ratios. J Bone Joint Surg Am 82:315–321
Widmer KH, Zurfluh B (2004) Compliant positioning of total hip components for optimal range of motion. J Orthop Res 22:815–821
Delimar D, Bićanić G, Pećina M, Korzinek K (2004) Acetabular roof reconstruction with pedicled iliac graft: early clinical experience. Int Orthop 28:319–320
Delimar D, Cicak N, Klobucar H, Pećina M, Korzinek K (2002) Acetabular roof reconstruction with pedicled iliac graft. Int Orthop 26:344–348
Erceg M (2009) The influence of femoral head shift on hip biomechanics: additional parameters accounted. Int Orthop 33:95–100
Delp SL, Wixson RL, Komattu AV, Kocmond JH (1996) How superior placement of the joint center in hip arthroplasty affects the abductor muscles. Clin Orthop Relat Res 328:137–146
Gore DR, Murray MP, Gardner GM, Sepic SB (1977) Roentgenographic measurements after Muller total hip replacement: Correlations among roentgenographic measurements and hip strength and mobility. J Bone Joint Surg Am 59:948–953
Bay BK, Hamel AJ, Olson SA, Sharkey NA (1997) Statically equivalent load and support conditions produce different hip joint contact pressures and periacetabular strains. J Biomech 30:193–196
Dearborn JT, Harris WH (1999) High placement of an acetabular component inserted without cement in a revision total hip arthroplasty. Results after a mean of ten years. J Bone Joint Surg Am 81:469–480
Charnley J (1970) Total hip replacement by low-friction arthroplasty. Clin Orthop 72:7–21
Johnston RC, Brand RA, Crowninshield RD (1979) Reconstruction of the hip: A mathematical approach to determine optimum geometric relationships. J Bone Joint Surg Am 61:639–652
Bicanic G, Delimar D, Delimar M, Pecina M (2009) Influence of the acetabular cup position on hip load during arthroplasty in hip dysplasia. Int Orthop 32:397–402
Poss R, Walker P, Spector M, Reilly DT, Robertson DD, Sledge CB (1998) Strategies for improving fixation of femoral components in total hip arthroplasty. Clin Orthop Relat Res 235:181–194
Hirakawa K, Mitsugi N, Koshino T, Saito T, Hirasawa Y, Kubo T (2001) Effect of acetabular cup position and orientation in cemented total hip arthroplasty. Clin Orthop Relat Res 388:135–142
Delp SL, Maloney W (1993) Effects of hip center location on the moment-generating capacity of the muscles. J Biomech 26:485–499
Olson SA, Bay BK, Chapman MW, Sharkey NA (1995) Biomechanical consequences of fracture and repair of the posterior wall of the acetabulum. J Bone Joint Surg Am 77:1184–1192
Acknowledgments
We thank Michael D. Brodt and Daniel J. Wood (Washington University School of Medicine) for critical reading of this paper. We are grateful to Ziji Zhang, Zibo Yang, Zihong Lin, Xing Duan, Yuanchen Ma (Sun Yat-sen University) for their technical assistance in preparation of specimens, and Bo Bai and Nansheng Yu (Guangzhou Medical School) for excellent help in preparing this study.
Conflict of interest statement
This work was supported by China Science and Technology Grant of Guangdong (2007B031002003).
Author information
Authors and Affiliations
Corresponding author
Additional information
This study was approved by the Medical Ethical Board of Sun Yat-sen University.
Rights and permissions
About this article
Cite this article
Zhang, Z., Fu, M., Kang, Y. et al. Upward and inward displacements of the acetabular component increase stress on femoral head in single endoprothesis models. International Orthopaedics (SICOT) 34, 479–483 (2010). https://doi.org/10.1007/s00264-009-0788-z
Received:
Revised:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s00264-009-0788-z