Abstract
Purpose
In this paper, we aimed to evaluate the positron emission tomography (PET) performance of, to the best of our knowledge, the third commercially available whole-body integrated PET/magnetic resonance (MR) system.
Methods
The PET system performance was measured following the NEMA standards with and without simultaneous MR operation. PET spatial resolution, sensitivity, scatter fraction, count-rate performance, accuracy of count losses and random corrections, image quality, and time-of-flight (TOF) resolution were quantitatively evaluated. Clinical scans were acquired at the PET/MR system and compared with images acquired at a PET/CT with the same digital detector technology.
Results
Measurement results of essential PET performance were reported in the form of MR idle (MR pulsing). The axial, radial, and tangential spatial resolutions were measured as 2.72 mm (2.73 mm), 2.86 mm (2.85 mm), and 2.81 mm (2.82 mm) FWHM, respectively, at 1 cm radial offset. The NECR peak was measured as 129.2 kcps (129.5 kcps) at 14.7 kBq mL−1 (14.2 kBq mL−1). The scatter fraction at NECR peak was 37.9% (36.5%), and the maximum slice error below NECR was 4.1% (4.5%). Contrast recovery coefficients ranged from 51.8% (52.3%) for 10 mm hot sphere to 87.3% (87.2%) for 37 mm cold sphere. TOF resolution at 5.3 kBq mL−1 was measured at 535 ps (540 ps). With point source, TOF was measured to be 474 ps (485 ps). Clinical scans revealed similar image quality from the PET/MR and the comparative PET/CT system.
Conclusion
The PET performance of the newly introduced integrated PET/MR system is not significantly affected by the simultaneous operation of an MR sequence (2-point DIXON sequence). Measurement results demonstrate comparable performance with other state-of-the-art PET/MR systems. The clinical benefits of high spatial resolution and long axial coverage remain to be further evaluated in specific clinical imaging applications.
Similar content being viewed by others
Data availability
All of NEMA raw data images are stored at the Department of Nuclear Medicine, Zhongshan Hospital, Fudan University, China, and United Imaging Healthcare Co., Ltd., China.
References
Bailey DL, et al. Combined PET/MR: the real work has just started. Summary report of the third international workshop on PET/MR imaging; February 17-21, 2014, Tubingen, Germany. Mol Imaging Biol. 2015;17(3):297–312.
Bailey DL, et al. Combined PET/MR: where are we now? Summary report of the second international workshop on PET/MR imaging April 8-12, 2013, Tubingen, Germany. Mol Imaging Biol. 2014;16(3):295–310.
Bailey DL, et al. Summary report of the first international workshop on PET/MR imaging, March 19-23, 2012, Tubingen, Germany. Mol Imaging Biol. 2013;15(4):361–71.
Bailey DL, et al. Combined PET/MRI: from status quo to status go. Summary report of the fifth international workshop on PET/MR imaging, February 15-19, 2016, Tubingen, Germany. Mol Imaging Biol. 2016;18(5):637–50.
Bailey DL, et al. Combined PET/MRI: multi-modality multi-parametric imaging is here: summary report of the 4th international workshop on PET/MR imaging; February 23-27, 2015, Tubingen, Germany. Mol Imaging Biol. 2015;17(5):595–608.
Abgral R, et al. Clinical utility of combined FDG-PET/MR to assess myocardial disease. JACC Cardiovasc Imaging. 2017;10(5):594–7.
Nakajo K, et al. Diagnostic performance of fluorodeoxyglucose positron emission tomography/magnetic resonance imaging fusion images of gynecological malignant tumors: comparison with positron emission tomography/computed tomography. Jpn J Radiol. 2010;28(2):95–100.
Grueneisen J, et al. Integrated PET/MRI for whole-body staging of patients with primary cervical cancer: preliminary results. Eur J Nucl Med Mol Imaging. 2015;42(12):1814–24.
Shin HW, et al. Initial experience in hybrid PET-MRI for evaluation of refractory focal onset epilepsy. Seizure. 2015;31:1–4.
Miller-Thomas MM, Benzinger TL. Neurologic applications of PET/MR imaging. Magn Reson Imaging Clin N Am. 2017;25(2):297–313.
Vaska P, Cao T. The state of instrumentation for combined positron emission tomography and magnetic resonance imaging. In: Seminars in nuclear medicine. 2013. Elsevier.
Herzog H, Lerche C. Advances in clinical PET/MRI instrumentation. PET Clin. 2016;11(2):95–103.
Herzog H, Van Den Hoff J. Combined PET/MR systems: an overview and comparison of currently available options. Q J Nucl Med Mol Imaging. 2012;56(3):247–67.
Hope TA, et al. Summary of the first ISMRM-SNMMI workshop on PET/MRI: applications and limitations. J Nucl Med. 2019;60(10):1340–6.
Zaidi H, et al. Design and performance evaluation of a whole-body ingenuity TF PET-MRI system. Phys Med Biol. 2011;56(10):3091–106.
Delso G, et al. Performance measurements of the Siemens mMR integrated whole-body PET/MR scanner. J Nucl Med. 2011;52(12):1914–22.
Grant AM, et al. NEMA NU 2-2012 performance studies for the SiPM-based ToF-PET component of the GE SIGNA PET/MR system. Med Phys. 2016;43(5):2334.
Levin CS, et al. Performance of a high sensitivity time-of-flight PET ring operating simultaneously within a 3T MR system. In: EJNMMI physics. 2014. Springer.
Bailey DL, et al. Combined PET/MRI: global warming-summary report of the 6th international workshop on PET/MRI, March 27-29, 2017, Tubingen, Germany. Mol Imaging Biol. 2018;20(1):4–20.
Cao T, Chen S, Pang L. NEMA NU2-2012 performance measurements of the United-Imaging uPMR790: a HD TOF simultaneous PET/MR system. J Nucl Med. 2018;59(supplement 1):1856.
Association, N.E.M., NEMA Standards Publication NU-2 2012 performance measurements of positron emission tomographs. 2012: Rosslyn, VA.
Association, N.E.M., NEMA Standards Publication NU-2 2018 performance measurements of positron emission tomographs. 2018: Rosslyn, VA.
Hoffman EJ, Huang SC, Phelps ME. Quantitation in positron emission computed tomography: 1. Effect of object size. J Comput Assist Tomogr. 1979;3(3):299–308.
Watson CC, Newport D, Casey ME. A single scatter simulation technique for scatter correction in 3D PET. In: Three-dimensional image reconstruction in radiology and nuclear medicine: Springer; 1996. p. 255–68.
Lv Y, et al. Mini EXPLORER II: a prototype high-sensitivity PET/CT scanner for companion animal whole body and human brain scanning. Phys Med Biol. 2019;64(7):075004.
Bunka M, et al. Imaging quality of (44)Sc in comparison with five other PET radionuclides using Derenzo phantoms and preclinical PET. Appl Radiat Isot. 2016;110:129–33.
Cox BL, et al. Development of a novel linearly-filled Derenzo microPET phantom. Am J Nucl Med Mol Imaging. 2016;6(3):199–204.
Zhang Q, et al. Comparison of (68)Ga-PSMA-11 PET-CT with mpMRI for preoperative lymph node staging in patients with intermediate to high-risk prostate cancer. J Transl Med. 2017;15(1):230.
Liu G, et al. Validation of MR-based attenuation correction of a newly released whole-body simultaneous PET/MR system. Biomed Res Int. 2019;2019:8213215.
Amorim BJ, et al. Clinical impact of PET/MR in treated colorectal cancer patients. Eur J Nucl Med Mol Imaging. 2019;46(11):2260–9.
Galgano S, et al. Practical considerations for clinical PET/MR imaging. PET Clin. 2018;13(1):97–112.
Mansi L, Ciarmiello A. Perspectives on PET/MR imaging: are we ready for clinical use? J Nucl Med. 2014;55(4):529–30.
Rausch I, et al. Performance evaluation of the biograph mCT flow PET/CT system according to the NEMA NU2-2012 standard. EJNMMI Phys. 2015;2(1):26.
Vandendriessche D, et al. Performance characteristics of silicon photomultiplier based 15-cm AFOV TOF PET/CT. EJNMMI Phys. 2019;6(1):8.
Rausch I, et al. Performance evaluation of the Vereos PET/CT system according to the NEMA NU2-2012 standard. J Nucl Med. 2019;60(4):561–7.
van Sluis J, et al. Performance characteristics of the digital biograph vision PET/CT system. J Nucl Med. 2019;60(7):1031–6.
Moses WW. Fundamental limits of spatial resolution in PET. Nucl Instrum Methods Phys Res A. 2011;648 Supplement 1:S236–40.
MacDonald LR, Dahlbom M. Parallax correction in PET using depth of interaction information. IEEE Trans Nucl Sci. 1998;45(4):2232–7.
Deller TW, et al. PET imaging stability measurements during simultaneous pulsing of aggressive MR sequences on the SIGNA PET/MR system. J Nucl Med. 2018;59(1):167–72.
Acknowledgments
The authors thank Shanghai Municipal Key Clinical Specialty (shslczdzk03401).
Funding
Supported by National Key Research and Development Plan “Digital Diagnosis and Treatment Equipment Research and Development” Key Specialty (2016YFC0103900), Shanghai Science and Technology Commission “Science and Technology Innovation Action Plan” in the Field of Cooperation in the Field of Industry, Education and Research (19DZ1930700), Shanghai Municipal Key Clinical Specialty(shslczdzk03401).
Author information
Authors and Affiliations
Contributions
HS contributed to the design of this study. SC designed and performed the experiments. SC, YG, and HY were in charge of the preparation of the NEMA phantom and of the acquisition. SC, XC, and LH analyzed the data. SC and TC drafted the manuscript. All authors read and approved the final manuscript.
Corresponding author
Ethics declarations
Ethical approval
All procedures performed in studies involving human participants were in accordance with the ethical standards of the institutional and/or national research committee and with the 1964 Helsinki declaration and its later amendments or comparable ethical standards.
Informed consent
Informed consent was obtained from all individual participants included in the study.
Consent for publication
Not applicable.
Conflict of interest
Xin Chen, Tuoyu Cao, and Lingzhi Hu are employees of United Imaging Healthcare. The other authors working with Zhongshan Hospital have full control of the data and declare that they have no conflict of interest.
Additional information
Publisher’s note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
This article is part of the Topical Collection on Technology
Supplementary information
ESM 1
(DOCX 1057 kb).
Rights and permissions
About this article
Cite this article
Chen, S., Gu, Y., Yu, H. et al. NEMA NU2-2012 performance measurements of the United Imaging uPMR790: an integrated PET/MR system. Eur J Nucl Med Mol Imaging 48, 1726–1735 (2021). https://doi.org/10.1007/s00259-020-05135-9
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s00259-020-05135-9