A radiological technique for the assessment of wear in prosthetic knee replacements

  • Article
  • Published:
Skeletal Radiology Aims and scope Submit manuscript

Abstract 

Objective. Wear of the polyethylene insert is a well-recognised cause of implant failure in total knee replacements. The purpose of this study was to evaluate a simple, digital fluoroscopic technique for the assessment of wear in knee prostheses.

Design.Fluoroscopic images of knee prostheses were produced both of a phantom and in a patient group. Joint space thickness was measured by reference to a known diameter. Measurements were made to assess repeatability of positioning, inter- and intra-observer variance and the effect of angulation.

Results. Standard phantom images showed small variation between measurements, high inter-reader correlation (Pearson’s correlation coefficient, r=0.98, P<0.001; coefficient of variation=0.53%) and low intra-reader variation (coefficient of variation=0.57%). Inter- and intra-imager variation were low (coefficient of variation=1.05% and 0.88%, respectively). In the patient group, the range of joint space measurements was 1.9–8.9 mm. The coefficient of variation in insert measurements on repeated images was 2.0%. Repeatability of measurements was 0.2 mm with 99% confidence interval.

Conclusions. This technique allows repeatable, precise measurement of insert thickness. The technique may be adapted to any implant where a reliable calibrating distance is present.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Similar content being viewed by others

Author information

Authors and Affiliations

Authors

Additional information

Received: 21 December 1999 Revision requested: 4 April 2000 Revision received: 4 May 2000 Accepted: 11 May 2000

Rights and permissions

Reprints and permissions

About this article

Cite this article

Hide, I., Grainger, A., Wallace, I. et al. A radiological technique for the assessment of wear in prosthetic knee replacements. Skeletal Radiol 29, 583–586 (2000). https://doi.org/10.1007/s002560000240

Download citation

  • Issue Date:

  • DOI: https://doi.org/10.1007/s002560000240

Navigation