Objective. Wear of the polyethylene insert is a well-recognised cause of implant failure in total knee replacements. The purpose of this study was to evaluate a simple, digital fluoroscopic technique for the assessment of wear in knee prostheses.
Design.Fluoroscopic images of knee prostheses were produced both of a phantom and in a patient group. Joint space thickness was measured by reference to a known diameter. Measurements were made to assess repeatability of positioning, inter- and intra-observer variance and the effect of angulation.
Results. Standard phantom images showed small variation between measurements, high inter-reader correlation (Pearson’s correlation coefficient, r=0.98, P<0.001; coefficient of variation=0.53%) and low intra-reader variation (coefficient of variation=0.57%). Inter- and intra-imager variation were low (coefficient of variation=1.05% and 0.88%, respectively). In the patient group, the range of joint space measurements was 1.9–8.9 mm. The coefficient of variation in insert measurements on repeated images was 2.0%. Repeatability of measurements was 0.2 mm with 99% confidence interval.
Conclusions. This technique allows repeatable, precise measurement of insert thickness. The technique may be adapted to any implant where a reliable calibrating distance is present.
Similar content being viewed by others
Author information
Authors and Affiliations
Additional information
Received: 21 December 1999 Revision requested: 4 April 2000 Revision received: 4 May 2000 Accepted: 11 May 2000
Rights and permissions
About this article
Cite this article
Hide, I., Grainger, A., Wallace, I. et al. A radiological technique for the assessment of wear in prosthetic knee replacements. Skeletal Radiol 29, 583–586 (2000). https://doi.org/10.1007/s002560000240
Issue Date:
DOI: https://doi.org/10.1007/s002560000240