Evaluation of a new low-dose biplanar system to assess lower-limb alignment in 3D: a phantom study

  • Scientific Article
  • Published:
Skeletal Radiology Aims and scope Submit manuscript

Abstract

Objective

Knee coronal alignment is routinely assessed on a full-length radiograph of the lower limbs. However, poor positioning of the knee during the procedure affects the accuracy of this kind of measurement, particularly in cases combining knee rotation and flexion. The purpose of this study was to assess the value of a three-dimensional assessment of the hip-knee-ankle (HKA) angle based on a biplanar radiographic system.

Materials and methods

A biplanar slot scanning system was used to take radiographs of three lower-limb synthetic models with similar frontal deviation (5°valgus) but different flexion angulations (0°, 9°, and 18°). Biplane acquisitions were done with lower-limb axial rotations ranging from 20° of internal rotation to 20° of external rotation on each of the lower limb models. Three independent observers performed standard 2D measurements of the HKA angle from each anteroposterior (AP) image and also modeled the lower limb in 3D for each biplane acquisition with dedicated software. The HKA angle was automatically calculated from the 3D models. The results of the 2D and 3D techniques were compared.

Results

Axial rotation provoked 2D HKA measurement errors up to, respectively, 1.4°, 4.7°, and 6.8° for the lower extremities with 0°, 9°, and 18° flexion, while it never affected the 3D HKA measurement for more than 1.5°. Interobserver errors were 0.7° (SD = 0.5°) for the 2D measurements and 0.6° (SD = 0.4°) for the 3D measurements.

Conclusions

The 3D modeling allows for a more accurate evaluation of coronal alignment compared to 2D, eliminating bias due to wrong knee positioning.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Fig. 1
Fig. 2
Fig. 3
Fig. 4

Similar content being viewed by others

References

  1. Brouwer GM, van Tol AW, Bergink AP, Belo JN, Bernsen RM, Reijman M, et al. Association between valgus and varus alignment and the development and progression of radiographic osteoarthritis of the knee. Arthritis Rheum. 2007;56(4):1204–11.

    Article  PubMed  CAS  Google Scholar 

  2. Cerejo R, Dunlop DD, Cahue S, Channin D, Song J, Sharma L. The influence of alignment on risk of knee osteoarthritis progression according to baseline stage of disease. Arthritis Rheum. 2002;46(10):2632–6.

    Article  PubMed  Google Scholar 

  3. Sharma L, Song J, Felson DT, Cahue S, Shamiyeh E, Dunlop DD. The role of knee alignment in disease progression and functional decline in knee osteoarthritis. JAMA. 2001;286(2):188–95.

    Article  PubMed  CAS  Google Scholar 

  4. Mullaji AB, Padmanabhan V, Jindal G. Total knee arthroplasty for profound varus deformity: technique and radiological results in 173 knees with varus of more than 20 degrees. J Arthroplasty. 2005;20(5):550–61.

    Article  PubMed  Google Scholar 

  5. Williot A, Rosset P, Favard L, Brilhault J, Burdin P. Total knee arthroplasty in valgus knee. Orthop Traumatol, Surg Res. 2010;96(S4):S37–42.

    Google Scholar 

  6. Choong PF, Dowsey MM, Stoney JD. Does accurate anatomical alignment result in better function and quality of life? Comparing conventional and computer-assisted total knee arthroplasty. J Arthroplasty. 2009;24(4):560–9.

    Article  PubMed  Google Scholar 

  7. Cooke TD, Sled EA, Scudamore RA. Frontal plane knee alignment: a call for standardized measurement. J Rheumatol. 2007;34(9):1796–801.

    PubMed  Google Scholar 

  8. Sailer J, Scharitzer M, Peloschek P, Giurea A, Imhof H, Grampp S. Quantification of axial alignment of the lower extremity on conventional and digital total leg radiographs. Eur Radiol. 2005;15(1):170–3.

    Article  PubMed  CAS  Google Scholar 

  9. Sled EA, Sheehy LM, Felson DT, Costigan PA, Lam M, Cooke TD. Reliability of lower limb alignment measures using an established landmark-based method with a customized computer software program. Rheumatol Int. 2011;31(1):71–7.

    Article  PubMed  Google Scholar 

  10. Radtke K, Becher C, Noll Y, Ostermeier S. Effect of limb rotation on radiographic alignment in total knee arthroplasties. Arch Orthop Trauma Surg. 2010;130(4):451–7.

    Article  PubMed  Google Scholar 

  11. Swanson KE, Stocks GW, Warren PD, Hazel MR, Janssen HF. Does axial limb rotation affect the alignment measurements in deformed limbs? Clin Orthop Relat Res. 2000;371:246–52.

    Article  PubMed  Google Scholar 

  12. Brouwer RW, Jakma TS, Brouwer KH, Verhaar JA. Pitfalls in determining knee alignment: a radiographic cadaver study. J Knee Surg. 2007;20(3):210–5.

    PubMed  CAS  Google Scholar 

  13. Illes T, Somoskeoy S. The EOS imaging system and its uses in daily orthopaedic practice. Int Orthop. 2012. doi:10.1007/s00264-012-1512-y.

  14. Deschenes S, Charron G, Beaudoin G, Labelle H, Dubois J, Miron MC, et al. Diagnostic imaging of spinal deformities: reducing patients radiation dose with a new slot-scanning X-ray imager. Spine (Phila Pa 1976). 2010;35(9):989–94.

    Article  Google Scholar 

  15. Kalifa G, Charpak Y, Maccia C, Fery-Lemonnier E, Bloch J, Boussard JM, et al. Evaluation of a new low-dose digital x-ray device: first dosimetric and clinical results in children. Pediatr Radiol. 1998;28(7):557–61.

    Article  PubMed  CAS  Google Scholar 

  16. Than P, Szuper K, Somoskeoy S, Warta V, Illes T. Geometrical values of the normal and arthritic hip and knee detected with the EOS imaging system. Int Orthop. 2012;36(6):1291–97.

    Google Scholar 

  17. Ilharreborde B, Steffen JS, Nectoux E, Vital JM, Mazda K, Skalli W, et al. Angle measurement reproducibility using EOS three-dimensional reconstructions in adolescent idiopathic scoliosis treated by posterior instrumentation. Spine (Phila Pa 1976). 2011;36(20):E1306–13.

    Article  Google Scholar 

  18. Chaibi Y, Cresson T, Aubert B, Hausselle J, Neyret P, Hauger O, et al. Fast 3D reconstruction of the lower limb using a parametric model and statistical inferences and clinical measurements calculation from biplanar X-rays. Comput Methods Biomech Biomed Engin. 2012;15(5):457–66.

    Google Scholar 

  19. Harrison MM, Cooke TD, Fisher SB, Griffin MP. Patterns of knee arthrosis and patellar subluxation. Clin Orthop Relat Res. 1994;309:56–63.

    PubMed  Google Scholar 

  20. Cooke TD, Scudamore RA, Bryant JT, Sorbie C, Siu D, Fisher B. A quantitative approach to radiography of the lower limb. Principles and applications. J Bone Joint Surg Br. 1991;73(5):715–20.

    PubMed  CAS  Google Scholar 

  21. Sato T, Koga Y, Omori G. Three-dimensional lower extremity alignment assessment system: application to evaluation of component position after total knee arthroplasty. J Arthroplasty. 2004;19(5):620–8.

    Article  PubMed  Google Scholar 

  22. Sato T, Koga Y, Sobue T, Omori G, Tanabe Y, Sakamoto M. Quantitative 3-dimensional analysis of preoperative and postoperative joint lines in total knee arthroplasty: a new concept for evaluation of component alignment. J Arthroplasty. 2007;22(4):560–8.

    Article  PubMed  Google Scholar 

  23. Ariumi A, Sato T, Kobayashi K, Koga Y, Omori G, Minato I, et al. Three-dimensional lower extremity alignment in the weight-bearing standing position in healthy elderly subjects. J Orthop Sci. 2010;15(1):64–70.

    Article  PubMed  Google Scholar 

Download references

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Philippe Thelen.

Rights and permissions

Reprints and permissions

About this article

Cite this article

Thelen, P., Delin, C., Folinais, D. et al. Evaluation of a new low-dose biplanar system to assess lower-limb alignment in 3D: a phantom study. Skeletal Radiol 41, 1287–1293 (2012). https://doi.org/10.1007/s00256-012-1438-x

Download citation

  • Received:

  • Revised:

  • Accepted:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s00256-012-1438-x

Keywords

Navigation