Multiphase acquisitions in pediatric abdominal-pelvic CT are a common practice and contribute to unnecessary radiation dose

  • Original Article
  • Published:
Pediatric Radiology Aims and scope Submit manuscript

Abstract

Background

Many patients at our pediatric hospital have had a contrast-enhanced CT of the abdomen and pelvis performed by an outside imaging facility before admission. We have noticed that many of these exams are multiphase, which may contribute to unnecessary radiation dose.

Objective

To determine the frequency of multiphase acquisitions and radiation dose indices in contrast-enhanced CTs of the abdomen and pelvis performed by outside imaging facilities in patients who were subsequently transferred to our pediatric hospital for care, and compare these metrics to contrast-enhanced CTs of the abdomen and pelvis performed internally.

Materials and methods

A retrospective analysis was performed of contrast-enhanced CTs of the abdomen and pelvis from outside imaging facilities uploaded to our picture archiving and communication system (PACS) between January 1, 2012, and December 31, 2015. CT images and dose pages were reviewed to determine the number of phases and dose indices (CT dose index–volume [CTDIvol], dose-length product, size-specific dose estimate). Exams for abdominal or pelvic mass, trauma or urinary leak indications were excluded. Data were compared to internally acquired contrast-enhanced CTs of the abdomen and pelvis by querying the American College of Radiology (ACR) Dose Index Registry. This review was institutional review board and HIPAA compliant.

Results

There were 754 contrast-enhanced CTs of the abdomen and pelvis from 104 outside imaging facilities. Fifty-three percent (399/754) had 2 phases, and 2% (14/754) had 3 or more phases. Of the 939 contrast-enhanced CTs of the abdomen and pelvis performed internally, 12% (115) were multiphase exams. Of 88% (664) contrast-enhanced CTs of the abdomen and pelvis from outside imaging facilities with dose data, CTDIvol was 2.7 times higher than our institution contrast-enhanced CTs of the abdomen and pelvis (939) for all age categories as defined by the ACR Dose Index Registry (mean: 9.4 vs. 3.5 mGy, P<0.0001). The majority (74%) of multiphase exams were performed by 9 of 104 outside imaging facilities.

Conclusion

Multiphase acquisitions in routine contrast-enhanced CT of the abdomen and pelvis exams at outside imaging facilities are more frequent than those at a dedicated pediatric institution and contribute to unnecessary radiation dose. A contrast-enhanced CT of the abdomen and pelvis exam from an outside imaging facility with two passes may have as much as four times to six times the dose as the same exam performed with a single pass at a pediatric imaging center. We advocate for imaging facilities with high multiphase rates to eliminate multiple phases from routine contrast-enhanced CT of the abdomen and pelvis exams in children.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Fig. 1
Fig. 2
Fig. 3

Similar content being viewed by others

References

  1. Children’s Healthcare of Atlanta. https://med.emory.edu/clinical/childrens_healthcare.html. Accessed 9 Nov 2017

  2. Agarwal S, Jokerst C, Siegel MJ et al (2015) Pediatric emergency CT scans at a children’s hospital and at community hospitals: radiation technical factors are an important source of radiation exposure. AJR Am J Roentgenol 205:409–413

    Article  Google Scholar 

  3. Snow A, Milliren CE, Graham DA et al (2017) Quality of pediatric abdominal CT scans performed at a dedicated children’s hospital and its referring institutions: a multifactorial evaluation. Pediatr Radiol 47:391–397

    Article  Google Scholar 

  4. Smith-Bindman R, Wang Y, Yellen-Nelson TR et al (2017) Predictors of CT radiation dose and their effect on patient care: a comprehensive analysis using automated data. Radiology 282:182–193

    Article  Google Scholar 

  5. American College of Radiology ACR Appropriateness Criteria. https://acsearch.acr.org/docs/69467/Narrative/. Accessed 28 Jan 2018

  6. Huda W, Mettler FA (2010) Volume CT dose index and dose-length product displayed during CT: what good are they? Radiology 258:236–242

    Article  Google Scholar 

  7. Strauss KJ (2014) Dose indices: everybody wants a number. Pediatr Radiol 44:450–459

    Article  Google Scholar 

  8. ACR Dose Index Registry. https://www.acr.org/Quality-Safety/National-Radiology-Data-Registry/Dose-Index-Registry. Accessed 11 Feb 2016

  9. COCIR CT Manufacturers’ CTDIvol Reference Phantom Size FAQ. http://www.cocir.org/fileadmin/5_Initiatives/COCIR_CT_Manufacturers_Reference_phantom_size_FAQ_13102014.pdf. Accessed 11 Feb 2016

  10. McNitt-Gray M (2011) Assessing radiation dose: how to do it right. AAPM Summit on CT Dose. https://www.aapm.org/meetings/2011CTS/documents/McNitt-GrayAssessingRadiationdoseFINAL.pdf. Accessed 7 June 2018

  11. Radiation Dose Reporting. http://www.ucdmc.ucdavis.edu/radiology/RadiationDose.html. Accessed 11 Feb 2016

  12. AAPM Computed Tomography Radiation Dose Education Slides. https://www.aapm.org/pubs/CTProtocols/documents/ToshibaEducationSlides.pdf. Accessed 11 Feb 2016

  13. Seibert JA, Boone JM, Wootton-Gorges SL et al (2014) Dose is not always what it seems: where very misleading values can result from volume CT dose index and dose length product. J Am Coll Radiol 11:233–237

    Article  Google Scholar 

  14. ACR Dose Index Registry Measures. https://www.acr.org/~/media/ACR/Documents/PDF/QualitySafety/NRDR/DIR/DIR-Measures.pdf. Accessed 24 Nov 2016

  15. Boone JM, Strauss KJ, Cody DD et al (2011) Size-Specific Dose Estimates (SSDE) in pediatric and adult body CT examinations. AAPM report No 204. http://www.aapm.org/pubs/reports/RPT_204.pdf. Accessed 11 Feb 2016

  16. Strauss KJ, Goske MJ, Kaste SC et al (2010) Image gently: ten steps you can take to optimize image quality and lower CT dose for pediatric patients. AJR Am J Roentgenol 194:868–873

    Article  Google Scholar 

  17. Guite KM, Hinshaw JL, Ranallo FN et al (2011) Ionizing radiation in abdominal CT: unindicated multiphase scans are an important source of medically unnecessary exposure. J Am Coll Radiol 8:756–761

    Article  Google Scholar 

  18. Nievelstein RA, van Dam IM, van der Molen AJ (2010) Multidetector CT in children: current concepts and dose reduction strategies. Pediatr Radiol 40:1324–1344

    Article  Google Scholar 

  19. Goske MJ, Applegate KE, Boylan J et al (2008) The “image gently” campaign: increasing CT radiation dose awareness through a national education and awareness program. Pediatr Radiol 38:265–269

    Article  Google Scholar 

  20. Kanal KM, Butler PF, Sengupta D et al (2017) U.S. diagnostic reference levels and achievable doses for 10 adult CT examinations. Radiology 284:120–133

    Article  Google Scholar 

  21. Goske MJ, Strauss KJ, Coombs LP et al (2013) Diagnostic reference ranges for pediatric abdominal CT. Radiology 268:208–218

    Article  Google Scholar 

  22. Donnelly LF, Emery KH, Brody AS et al (2001) Minimizing radiation dose for pediatric body applications of single-detector helical CT. AJR Am J Roentgenol 176:303–306

    Article  CAS  Google Scholar 

  23. Kim ME, Orth RC, Fallon SC et al (2015) Performance of CT examinations in children with suspected acute appendicitis in the community setting: a need for more education. AJR Am J Roentgenol 204:857–860

    Article  Google Scholar 

  24. Goske MJ, Applegate KE, Bulas D et al (2012) Image Gently 5 years later: what goals remain to be accomplished in radiation protection for children? AJR Am J Roentgenol 199:477–479

    Article  Google Scholar 

  25. Goske MJ, Applegate KE, Bulas D et al (2011) Image Gently: progress and challenges in CT education and advocacy. Pediatr Radiol 41:461–466

    Article  Google Scholar 

  26. Parikh RA, Wien MA, Novak RD et al (2017) A comparison study of size-specific dose estimate calculation models. Pediatr Radiol 48:56–65

    Article  Google Scholar 

  27. Karmazyn B, Ai H, Klahr P et al (2016) How accurate is size-specific dose estimate in pediatric body CT examinations? Pediatr Radiol 46:1234–1240

    Article  Google Scholar 

  28. Little S, Grattan-Smith D, Johnson B (2012) CT radiation dose delivered by community hospitals and imaging centers. The Society of Pediatric Radiology 55th Annual Meeting

Download references

Acknowledgements

The authors would like to thank Christina A. Rostad, MD, for performing the statistical analysis and for preparing the figures.

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Bradley S. Rostad.

Ethics declarations

Conflicts of interest

None

Rights and permissions

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Rostad, B.S., Applegate, K.E., Kim, T. et al. Multiphase acquisitions in pediatric abdominal-pelvic CT are a common practice and contribute to unnecessary radiation dose. Pediatr Radiol 48, 1714–1723 (2018). https://doi.org/10.1007/s00247-018-4192-y

Download citation

  • Received:

  • Revised:

  • Accepted:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s00247-018-4192-y

Keywords

Navigation