Abstract
Background
There are mechanisms within CT scanners and shielding that can be used to lower dose to patients.
Objective
The objective of our study was to evaluate the dose changes and image quality with and without a breast shield for multidetector CT of the chest with angular tube current modulation.
Materials and methods
An anthropomorphic 5-year-old phantom was scanned with and without bismuth breast shielding and with and without angular tube current modulation. Image quality was assessed by measuring signal-to-noise ratio and contrast-to-noise ratio as well as measuring qualitatively. ANOVA single-factor p-value analysis of paired datasets was used for the statistical analysis.
Results
The breast shielding and angular tube current modulation resulted in a dose reduction to 85% for the breast and 90% overall. Although no qualitative differences were noted with and without the breast shielding used with angular tube current modulation, there were statistically significant differences in the contrast-to-noise ratio and CT numbers.
Conclusions
When used with angular tube current modulation, breast shields influence the contrast-to-noise ratio and CT numbers, and therefore should not be used for chest CT. Greater dose reduction can be obtained by reducing the reference mAs (or reference noise index).
Similar content being viewed by others
References
American Cancer Society (2012) Cancer facts & figures 2012. American Cancer Society, Atlanta
Hurwitz LM, Yoshizumi TT, Reiman RE et al (2006) Radiation dose to the female breast from 16-MDCT body protocols. AJR Am J Roentgenol 186:1718–1722
Ostroumova E, Preston DL, Ron E et al (2008) Breast cancer incidence following low-dose rate environmental exposure: Techa River Cohort, 1956–2004. Br J Cancer 99:1940–1945
Brenner D, Elliston C, Hall E et al (2001) Estimated risks of radiation-induced fatal cancer from pediatric CT. AJR Am J Roentgenol 176:289–296
Fricke BL, Donnelly LF, Frush DP et al (2003) In-plane bismuth breast shields for pediatric CT: effects on radiation dose and image quality using experimental and clinical data. AJR Am J Roentgenol 180:407–411
Mayo JR, Aldrich J, Muller NL et al (2003) Radiation exposure at chest CT: a statement of the Fleischner Society. Radiology 228:15–21
Aldrich JE, Chang SD, Bilawich AM et al (2006) Radiation dose in abdominal computed tomography: the role of patient size and the selection of tube current. Can Assoc Radiol J 57:152–158
Angel E, Yaghmai N, Jude CM et al (2009) Dose to radiosensitive organs during routine chest CT: effects of tube current modulation. AJR Am J Roentgenol 193:1340–1345
Colombo P, Pedroli G, Nicoloso M et al (2004) Evaluation of the efficacy of a bismuth shield during CT examinations. Radiol Med 108:560–568
Diederich S, Lenzen H (2000) Radiation exposure associated with imaging of the chest: comparison of different radiographic and computed tomography techniques. Cancer 89:2457–2460
Heaney DE, Norvill CA (2006) A comparison of reduction in CT dose through the use of gantry angulations or bismuth shields. Australas Phys Eng Sci Med 29:172–178
Hohl C, Wildberger JE, Suss C et al (2006) Radiation dose reduction to breast and thyroid during MDCT: effectiveness of an in-plane bismuth shield. Acta Radiol 47:562–567
McCollough CH (2008) CT dose: how to measure, how to reduce. Health Phys 95:508–517
Kalra MK, Maher MM, Toth TL et al (2004) Strategies for CT radiation dose optimization. Radiology 230:619–628
Preston DL, Mattsson A, Holmberg E et al (2002) Radiation effects on breast cancer risk: a pooled analysis of eight cohorts. Radiat Res 158:220–235
Hrubec Z, Boice JD Jr, Monson RR et al (1989) Breast cancer after multiple chest fluoroscopies: second follow-up of Massachusetts women with tuberculosis. Cancer Res 49:229–234
Davis FG, Boice JD Jr, Hrubec Z et al (1989) Cancer mortality in a radiation-exposed cohort of Massachusetts tuberculosis patients. Cancer Res 49:6130–6136
Boice JD Jr (2001) Radiation and breast carcinogenesis. Med Pediatr Oncol 36:508–513
Yilmaz MH, Albayram S, Yasar D et al (2007) Female breast radiation exposure during thorax multidetector computed tomography and the effectiveness of bismuth breast shield to reduce breast radiation dose. J Comput Assist Tomogr 31:138–142
Hopper KD, King SH, Lobell ME et al (1997) The breast: in-plane x-ray protection during diagnostic thoracic CT–shielding with bismuth radioprotective garments. Radiology 205:853–858
Kalra MK, Dang P, Singh S et al (2009) In-plane shielding for CT: effect of off-centering, automatic exposure control and shield-to-surface distance. Korean J Radiol 10:156–163
Vollmar SV, Kalender WA (2009) Reduction of dose to the female breast as a result of spectral optimisation for high-contrast thoracic CT imaging: a phantom study. Br J Radiol 82:920–929
Wang J, Duan X, Christner JA et al (2011) Radiation dose reduction to the breast in thoracic CT: comparison of bismuth shielding, organ-based tube current modulation, and use of a globally decreased tube current. Med Phys 38:6084–6092
Tack D, De Maertelaer V, Gevenois PA (2003) Dose reduction in multidetector CT using attenuation-based online tube current modulation. AJR Am J Roentgenol 181:331–334
Coursey C, Frush DP, Yoshizumi T et al (2008) Pediatric chest MDCT using tube current modulation: effect on radiation dose with breast shielding. AJR Am J Roentgenol 190:W54–W61
Geleijns J, Want J, McCollough C (2010) The use of breast shielding for dose reduction in pediatric CT: arguments against the proposition. Pediatr Radiol 40:1744–1747
Geleijns J, Salvado Artells M, Veldkamp WJ et al (2006) Quantitative assessment of selective in-plane shielding of tissues in computed tomography through evaluation of absorbed dose and image quality. Eur Radiol 16:2334–2340
Leswick DA, Hunt MM, Webster ST et al (2008) Thyroid shields versus z-axis automatic tube current modulation for dose reduction at neck CT. Radiology 249:572–580
Wang J, Duan X, Christner JA et al (2012) Bismuth shielding, organ-based tube current modulation, and global reduction of tube current for dose reduction to the eye at head CT. Radiology 262:191–198
American Association of Physicists in Medicine (2012) Use of bismuth shielding for the purpose of dose reduction in CT scanning. AAPM, College Park
Siemens Healthcare (2010) Guide to low dose. Siemens Healthcare, Erlangen
Conflicts of interest
None.
Author information
Authors and Affiliations
Corresponding author
Rights and permissions
About this article
Cite this article
Servaes, S., Zhu, X. The effects of bismuth breast shields in conjunction with automatic tube current modulation in CT imaging. Pediatr Radiol 43, 1287–1294 (2013). https://doi.org/10.1007/s00247-013-2687-0
Received:
Revised:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s00247-013-2687-0