Central venous cannulation: are routine chest radiographs necessary after B-mode and colour Doppler sonography check?

  • Original Article
  • Published:
Pediatric Radiology Aims and scope Submit manuscript

Abstract

Background

After the insertion of a central venous catheter, a chest radiograph is usually obtained to ensure correct positioning of the catheter tip.

Objective

To determine in a paediatric population whether B-mode and colour Doppler sonography after central venous access is useful to evaluate catheter position, thus obviating the need for a postprocedural radiograph.

Materials and methods

A prospective study of 107 consecutive central venous access procedures placed in a paediatric intensive care unit was performed. At the end of the procedure, B-mode and colour Doppler sonography were used to assess catheter position and check for complications. A postprocedural chest radiograph was obtained in all patients.

Results

In 96 patients postprocedural B-mode and colour Doppler sonography showed colour Doppler signals within the vena cava. Among the 11 patients predicted to have a potential complication, there was one pneumothorax and ten malpositions. Chest radiography showed a total of 13 complications—1 pneumothorax and 12 malpositions. The concordance between colour Doppler sonography and chest radiography was 98.1% in the detection of catheter position; sonography had a sensitivity of 84.6% and a specificity of 100%.

Conclusions

The close concordance between B-mode and colour Doppler sonography and chest radiography justifies the more frequent use of sonography to evaluate catheter position because ionizing radiation is eliminated. Chest radiography may then be performed only when there is suspected inappropriate catheter tip position after sonography.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Fig. 1
Fig. 2
Fig. 3

Similar content being viewed by others

References

  1. Aubaniac R (1952) L’injection intraveineuse sous-claviculaire. Presse Med 60:1656–1659

    Google Scholar 

  2. Wilson JN, Grow JB, Demong CV, et al (1962) Central venous pressure in optimal blood volume maintenance. Arch Surg 85:563–577

    PubMed  CAS  Google Scholar 

  3. Johnson EM, Saltzman DA, Suh G, et al (1998) Complications and risks of central venous catheter placement in children. Surgery 124:911–916

    PubMed  CAS  Google Scholar 

  4. Sterzel JP, Green TP, Fuhrman BP, et al (1989) Percutaneous central venous catheterization in a pediatric intensive care unit: a survival analysis of complications. Crit Care Med 17:984–988

    Google Scholar 

  5. Etheridge SP, Berry JM, Krabill KA, et al (1995) Echocardiographic guided internal jugular venous cannulation in children with heart disease. Arch Pediatr Adolesc Med 149:77–80

    PubMed  CAS  Google Scholar 

  6. Giacoia GP (1991) Cardiac tamponade and hydrothorax as a complication of central venous parenteral nutrition in infants. J Parenter Enteral Nutr 15:110–113

    Article  CAS  Google Scholar 

  7. Casado-Flores J, Valdivielso-Serna A, Perez-Jurado L, et al (1991) Subclavian vein catheterization in critically ill children: analysis of 322 cannulations. Intensive Care Med 17:350–354

    Article  PubMed  CAS  Google Scholar 

  8. Girvan DP, deVeber LL, Inwood MJ, et al (1994) Subcutaneous infusion ports in the pediatric patient with hemophilia. J Pediatr Surg 29:1220–1223

    Article  PubMed  CAS  Google Scholar 

  9. Albanese CT, Wiener ES (1993) Venous access in pediatric oncology patients. Semin Surg Oncol 9:467–477

    Article  PubMed  CAS  Google Scholar 

  10. Gauderer MW, Stellato TA (1985) Subclavian Broviac catheter placements. J Pediatr Surg 20:402–405

    Article  PubMed  CAS  Google Scholar 

  11. Sznajder JI, Zveibil FR, Bitterman H, et al (1986) Central venous catheterization: failure and complication rates by three percutaneous approaches. Arch Intern Med 146:259–261

    Article  PubMed  CAS  Google Scholar 

  12. Yerdel MA, Karayalcin K, Aras N, et al (1991) Mechanical complications of subclavian vein catheterization: a prospective study. Int Surg 76:18–22

    PubMed  CAS  Google Scholar 

  13. Bernard RW, Stahl WM (1971) Subclavian vein catheterization. A prospective study. Ann Surg 173:184–190

    Article  PubMed  CAS  Google Scholar 

  14. Yerdel MA, Karayalcin K, Anadol E (1991) Malpositioning of subclavian vein catheters in left and right sided attempts: a prospective study. Neth J Surg 43:178–180

    PubMed  CAS  Google Scholar 

  15. Wiener EJ, McGuire P, Stolar CJ, et al (1992) The CCSG prospective study of venous access devices: an analysis of insertions and causes for removal. J Pediatr Surg 27:155–164

    Article  PubMed  CAS  Google Scholar 

  16. Gladwin MT, Slonim A, Landucci DL, et al (1999) Cannulation of the internal jugular vein: is a postprocedural chest radiography always necessary? Crit Care Med 27:1819–1823

    Article  PubMed  CAS  Google Scholar 

  17. Food and Drug Administration (1989) Precautions necessary with central venous catheters. US Government Printing Office, Washington, pp 15–16

    Google Scholar 

  18. Caridi JG, West JH, Stavropoulos SW, et al (2000) Internal jugular and upper extremity central venous access in interventional radiology: is a postprocedure chest radiograph necessary? AJR 174:363–366

    PubMed  CAS  Google Scholar 

  19. Trerotola SO, Johnson MS, Harrid VJ, et al (1997) Outcome of tunneled hemodialysis catheters placed via the right internal jugular vein by interventional radiologist. Radiology 203:489–495

    PubMed  CAS  Google Scholar 

  20. Lucey B, Varghese JC, Haslam P, et al (1999) Routine chest radiographs after central line insertion: mandatory postprocedural evaluation or unnecessary waste of resources. Cardiovasc Intervent Radiol 2:381–384

    Article  Google Scholar 

  21. Chang TC, Funaki B, Szymski GX (1998) Are routine chest radiographs necessary after image-guided placement of internal jugular central venous access devices. AJR 170:335–337

    PubMed  CAS  Google Scholar 

  22. Guth AA (2001) Routine chest X-rays after insertion of implantable long-term venous catheters: necessary or not? Am Surg 67:26–29

    PubMed  CAS  Google Scholar 

  23. Burn PR, Skewes D, King M (2001) Role of chest radiography after the insertion of a subclavian vein catheter for ambulatory chemotherapy. Can Assoc Radiol J 52:392–394

    PubMed  CAS  Google Scholar 

  24. Janik JE, Cothren CC, Janik JS, et al (2003) Is a routine chest x-ray necessary for children after fluoroscopically assisted venous access? J Pediatr Surg 38:1199–1202

    Article  PubMed  Google Scholar 

Download references

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Cecilia Lanza.

Rights and permissions

Reprints and permissions

About this article

Cite this article

Lanza, C., Russo, M. & Fabrizzi, G. Central venous cannulation: are routine chest radiographs necessary after B-mode and colour Doppler sonography check?. Pediatr Radiol 36, 1252–1256 (2006). https://doi.org/10.1007/s00247-006-0307-y

Download citation

  • Received:

  • Revised:

  • Accepted:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s00247-006-0307-y

Keywords

Navigation