Abstract
We report our experience with the use of covered stents for the management of coarctation of the aorta. From December 2001 to March 2004, nine patients (seven males; median age, 31 years; mean weight, 65 ± 15 kg) underwent implantation. Indications included critical or atretic native coarctation (n = 4), patients >50 years of age (n = 2), associated patent ductus arteriosus (n = 1) or adjacent aneurysm (n = 1), and the presence of a circumferential fracture within a previously implanted stent (n = 1). The covered balloon-expandable Cheatham-Platinum stent and the self-expandable stent graft Braile were employed. Adequate implantation was observed in all patients. Gradients were reduced from 54 ± 14 to 3 ± 8 mmHg and the coarctation site increased from 2.4 ± 2.9 to 15.9 ± 4.3 mm. The patent ductus arteriosus was immediately closed and the aneurysm excluded. Two patients >35 years with aneurysmal ascending aorta and metallic aortic prosthesis had aneurysm formation at follow-up, with one undergoing aneurysm exclusion using a Braile stent. Although covered stents are useful in the management of selected patients with coarctation, aneurysm formation may still occur in patients with markers of aortic wall weakness. Refinements in the deployment technique and/or the stent design are needed to eliminate this risk.
Similar content being viewed by others
References
ZR Bulbul E Bruckheimer JC Love JT Fahey WE Hellenbrand (1996) ArticleTitleImplantation of balloon-expandable stents for coarctation of the aorta: implantation data and short-term results Cathet Cardiovasc Diagn 39 36–42 Occurrence Handle8874943 Occurrence Handle1:STN:280:DyaK2s%2FivV2gsw%3D%3D
JP Cheatham (1999) ArticleTitleStents and Amplatzers: what's an interventionalist to do? Cathet Cardiovasc Interv 47 39–40 Occurrence Handle1:STN:280:DyaK1MzhtFyrsA%3D%3D
JP Cheatham (2001) ArticleTitleStenting of coarctation of the aorta Cathet Cardiovasc Interv 54 112–125 Occurrence Handle10.1002/ccd.1249 Occurrence Handle1:STN:280:DC%2BD3MrgtlKitg%3D%3D
JP Cheatham (2003) NuMED Cheatham platinum stents: role in the management of congenital heart defects PS Rao MJ Kern (Eds) Catheter Based Devices Lippincott Williams & Wilkins Philadelphia 353–368
JV Giovanni Particlede (2001) ArticleTitleCovered stents in the treatment of aortic coarctation J Interv Cardiol 14 187–190 Occurrence Handle12053302
RS Dieter (2001) ArticleTitleTransluminal endovascular stent grafting of aortic dissections and aneurysms: a concise review of the major trials Clin Cardiol 24 358–363 Occurrence Handle11346242 Occurrence Handle1:STN:280:DC%2BD3M3ltVChsw%3D%3D Occurrence Handle10.1002/clc.4960240503
C Duke E Rosenthal SA Qureshi (2003) ArticleTitleThe efficacy and safety of stent redilatation in congenital heart disease Heart 89 905–912 Occurrence Handle12860870 Occurrence Handle1:STN:280:DC%2BD3szjtVyjsQ%3D%3D
WD Edwards DS Leaf JE Edwards (1978) ArticleTitleDissecting aortic aneurysm associated with congenital bicuspid aortic valve Circulation 57 1022–1025 Occurrence Handle639201 Occurrence Handle1:STN:280:DyaE1c7ks1ygsA%3D%3D
ME Fawzy M Awad H Hegazy et al. (2003) ArticleTitleLong-term follow-up (15-year) results of balloon coarctation angioplasty in the adolescent and adult patient J Am Coll Cardiol 41 488 Occurrence Handle10.1016/S0735-1097(03)82667-0
SE Fletcher MR Nihill RG Grifka MP O'Laughlin CE Mullins (1995) ArticleTitleBalloon angioplasty of native coarctation of the aorta: midterm follow-up and prognostic factors J Am Coll Cardiol 25 730–734 Occurrence Handle7860921 Occurrence Handle10.1016/0735-1097(94)00437-U Occurrence Handle1:STN:280:DyaK2M7msVykuw%3D%3D
T Forbes D Matisoff J Dysart S Aggarwal (2003) ArticleTitleTreatment of coexistent coarctation and aneurysm of the aorta with covered stent in a pediatric patient Pediatr Cardiol 24 289–291 Occurrence Handle12522654 Occurrence Handle10.1007/s00246-002-0262-9 Occurrence Handle1:STN:280:DC%2BD3s3mt12muw%3D%3D
J Gunn T Cleveland P Gaines (1999) ArticleTitleCovered stent to treat co-existent coarctation and aneurysm of the aorta in a young man Heart 82 351 Occurrence Handle10455088 Occurrence Handle1:STN:280:DyaK1Mzoslagtg%3D%3D
MA Hamdan S Maheshwari JT Fahey WE Hellenbrand (2001) ArticleTitleEndovascular stents for coarctation of the aorta: initial results and intermediate-term follow-up J Am Coll Cardiol 38 1518–1523 Occurrence Handle11691533 Occurrence Handle10.1016/S0735-1097(01)01572-8 Occurrence Handle1:STN:280:DC%2BD3MnktVOrsw%3D%3D
DA Harrison PR McLaughlin C Lazzam M Connelly LN Benson (2001) ArticleTitleEndovascular stents in the management of coarctation of the aorta in the adolescent and adult: one year follow up Heart 85 561–566 Occurrence Handle11303011 Occurrence Handle10.1136/heart.85.5.561 Occurrence Handle1:STN:280:DC%2BD3M3ms1Wjsg%3D%3D
H Kaemmerer (2003) Aortic coarctation and interrupted aortic arch MA Gatzoulis GD Webb PEF Daubeney (Eds) Diagnosis and Management of Adult Congenital Heart Disease Churchill Livingstone Edinburgh, UK 253–264
MS Khan JW Moore (2000) ArticleTitleTreatment of abdominal aortic pseudoaneurysm with covered stents in a pediatric patient Cathet Cardiovasc Interv 50 445–448 Occurrence Handle10.1002/1522-726X(200008)50:4<445::AID-CCD17>3.0.CO;2-1 Occurrence Handle1:STN:280:DC%2BD3cvjsFSnsQ%3D%3D
J Koerselnan H Vries Particlede W Jaarsma et al. (2000) ArticleTitleBalloon angioplasty of coarctation of the aorta: a safe alternative for surgery in adults: immediate and mid-term results Cathet Cardiovasc Interv 50 28–33
SJ Korkola CI Tchervenkov D Shum-Tim N Roy (2002) ArticleTitleAortic rupture after stenting of a native coarctation in an adult Ann Thorac Surg 74 936 Occurrence Handle12238877 Occurrence Handle10.1016/S0003-4975(01)03160-5
J Lindsay SuffixJr (1988) ArticleTitleCoarctation of the aorta, bicuspid aortic valve and abnormal ascending aortic wall Am J Cardiol 61 182–184 Occurrence Handle3276120 Occurrence Handle10.1016/0002-9149(88)91327-6
AG Magee G Brzezinska-Rajszys SA Qureshi et al. (1999) ArticleTitleStent implantation for aortic coarctation and recoarctation Heart 82 600–606 Occurrence Handle10525517 Occurrence Handle1:STN:280:DyaK1Mvlslaiug%3D%3D
BW McCrindle TK Jones WR Moaow et al. (1996) ArticleTitleAcute results of balloon angioplasty of native coarctation versus recurrent aortic obstruction are equivalent. Valvuloplasty and Angioplasty of Congenital Anomalies (VACA) Registry Investigators J Am Coll Cardiol 28 1810–1817 Occurrence Handle8962571 Occurrence Handle1:STN:280:DyaK2s7it1ahug%3D%3D
C Ovaert BW McCrindle D Nykanen et al. (2000) ArticleTitleBalloon angioplasty of native coarctation: clinical outcomes and predictors of success J Am Coll Cardiol 35 988–996 Occurrence Handle10732899 Occurrence Handle10.1016/S0735-1097(99)00646-4 Occurrence Handle1:STN:280:DC%2BD3c7psFSgtQ%3D%3D
JH Palma JA Souza Particlede CM Rodrigues Alves et al. (2002) ArticleTitleSelf-expandable aortic stent-grafts for treatment of descending aortic dissections Ann Thorac Surg 73 1138–1141 Occurrence Handle11996254 Occurrence Handle10.1016/S0003-4975(02)03397-0
CA Pedra CB Pilla SL Braga CA Esteves VF Fontes (2002) ArticleTitleManagement of a large pseudoaneurysm secondary to balloon dilatation for native coarctation of the aorta with coil occlusion after stent implantation in a child Cathet Cardiovasc Interv 56 262–266 Occurrence Handle10.1002/ccd.10185
SA Qureshi M Zubrzycka G Brzezinska-Rajszys A Kosciesza J Ksiazyk (2004) ArticleTitleThe use of covered Cheatham-Platinum stents in aortic coarctation and recoarctation Cardiol Young 14 50–54 Occurrence Handle15237671 Occurrence Handle10.1017/S104795110400109X
CS Roberts WC Roberts (1991) ArticleTitleDissection of the aorta associated with congenital malformation of the aortic valve J Am Coll Cardiol 17 712–716 Occurrence Handle1993792 Occurrence Handle1:STN:280:DyaK3M7jtF2juw%3D%3D Occurrence Handle10.1016/S0735-1097(10)80188-3
CM Rodrigues Alves JH da Fonseca JA Souza Particlede AC Camargo Carvalho E Buffolo (2002) ArticleTitleEndovascular treatment of thoracic disease: patient selection and a proposal of a risk score Ann Thorac Surg 73 1143–1148 Occurrence Handle11996255
M Sadiq NH Malick SA Qureshi (2003) ArticleTitleSimultaneous treatment of native coarctation of the aorta combined with patent ductus arteriosus using a covered stent Cathet Cardiovasc Interv 59 387–390 Occurrence Handle10.1002/ccd.10501
TJ Schlatmann AE Becker (1977) ArticleTitleHistologic changes in the normal aging aorta: implications for dissecting aortic aneurysm Am J Cardiol 39 13–20 Occurrence Handle831420 Occurrence Handle1:STN:280:DyaE2s%2FosV2hsQ%3D%3D
BD Thanopoulos L Hadjinikolaou GN Konstadopoulou et al. (2000) ArticleTitleStent treatment for coarctation of the aorta: intermediate term follow up and technical considerations Heart 84 65–70 Occurrence Handle10862593 Occurrence Handle10.1136/heart.84.1.65 Occurrence Handle1:STN:280:DC%2BD3czitFajtg%3D%3D
C Varma LN Benson J Butany PR McLaughlin (2003) ArticleTitleAortic dissection after stent dilatation for coarctation of the aorta: a case report and literature review Cathet Cardiovasc Interv 59 528–535 Occurrence Handle10.1002/ccd.10548 Occurrence Handle1:STN:280:DC%2BD3szlslyiuw%3D%3D
Author information
Authors and Affiliations
Corresponding author
Rights and permissions
About this article
Cite this article
Pedra, C., Fontes, V., Esteves, C. et al. Use of Covered Stents in the Management of Coarctation of the Aorta. Pediatr Cardiol 26, 431–439 (2005). https://doi.org/10.1007/s00246-004-0814-2
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s00246-004-0814-2