Introduction

Photometry should be regarded as the most often applied detection technique in the analytical practice because almost all analytes are colored or could be selectively converted into colored species for easy detection using optical methods. Many of them have been transferred to the sensor format. A vast number of optical sensors and biosensors reported and reviewed in the analytical literature [15] foster the detection of a broad spectrum of analytes ranging from simple molecules (including gases, pH, anions and cations), to organic compounds encompassing low molecular-weight sugars, metabolites and toxins and large biomolecules as well including peptides, proteins, enzymes, immunocompounds, DNA etc. The operation of the most popular fiber-optic chemical sensors (FOCS) is based on analyte-dependent changes of optical properties such as absorbance, diffuse reflectance or fluorescence of their receptor part [15]. The chemical sensing and signal processing capability of the chemo-optical interface is achieved by the incorporation of a given reagent. The reagent layer usually contains a dye (e.g., chromoionophores or acid–base indicators) which reversibly reacts with the analyte contained in the sample. The main role of the optical fibers is the transmission of light from the external light source to the sensing layer and from the sensing layer to the detection system. The advantages of such assembly are the ability of remote sensing and the probe format of FOCS. However, the use of optical sensors in flow measurements demands additional adaptation of FOCS to a flow-through format or the FOCS mounting into dedicated flow cells of optimized geometry.

The main goal of this work is the development of a fibreless, highly compact flow-through optosensing device containing the three main components of optical chemosensors, that is, an immobilized chemical that optically responds to the analyte (chemoreceptor), compatible light source and optical detector, in a miniaturized format using extremely low-budget optoelectronics (light emitting diodes, LEDs) and signal transducing/recording equipment (ordinary voltmeter). The device is based on the concept of paired-emitter-detector-diode (PEDD) developed for batchwise analytical measurements by Diamond’s group [6]. According to this concept, two LEDs are paired together, one used in a conventional way as a light source whereas the second one as a light detector, resulting in a complete absorbance detector. Tymecki et al. [7] demonstrated that the electromotive force generated by PEDDs can be exploited as analytical signal. Hence, the coupling of PEDDs to pH meters [7] or ordinary voltmeters [8] leads to integrated, low-cost photometers. As demonstrated previously [9], ordinary, low-budget multimeters can play a double role, that is, working as an electric source for detector powering and simultaneously as a signal-reading instrument. The reported PEDD-based devices have been applied so far to batchwise (bio)analytical cuvette measurements [912] with potential use as flow-through cells in flow injection analysis [713] and chromatographic separations [1417].

There has been however little effort as of yet towards the integration of optoelectronic components with chemosensitive layers for design of dedicated optrodes. The sensoric operation of the integrated device is the result of photometrically monitored changes in the optical properties of the chemoreceptor. A survey of literature revealed two recent examples of the usage of PEDDs in combination with sensitive films, yet merely applied to textile-based wearable [18] and gas [19] pH sensing. To the best of our knowledge, however, no flow-through PEDD-optrode for assays of aqueous samples in flow analysis or meso/microfluidic devices has been reported so far.

In this communication, a universal device for flow-through chemical optosensing is presented and characterized. The integrated chemosensor features four assets: (a) designed as a dedicated detector for flow analysis and related approaches thereof including microchip-based platforms; (b) compatible with optically chemosensitive materials obtained from conventional large-scale methodologies, that is, physical or chemical deposition onto thin transparent plastic films like polyester foil, (c) miniaturized and tailorable to a hand-held fashion and (d) cost-effective and easy to fabricate.

Materials and methods

Chemicals and reagents

4-(Pyrrol-1-yl)benzoic acid (Pyr-BAc) was obtained from Aldrich (Germany). All other reagents used in this work were of analytical grade. All solutions used in this work were prepared using doubly distilled water.

Sensing film preparation

The Prussian Blue (PB) film used in this study as an optochemically sensitive component of the sensor [20] was obtained by soaking a 100 μm-thick transparent plastic foil (Lumocolor® universal photocopy film, Staedtler®,) into a solution of 0.1 mol L−1 K3Fe(CN)6 in 1 mol L−1 HCl saturated with Pyr-BAc as additive. In a strongly acidic environment the hexacyanoferrate complex slowly decomposes forming the PB film (so-called “insoluble form”, that is, (Fe4[Fe(CN)6]3). Oxidative polymerization of Pyr-BAc also takes place, leading to the formation of a mixed inorganic–organic composite material. The chemical process for film deposition is initiated and accelerated by UV irradiation. The preparation protocol and mechanism of the nonelectrochemical deposition of the composite material on nonconducting supports have been reported in detail elsewhere [21].

Apparatus and analytical procedure

The construction of the sensor device presented in this work capitalizes on LEDs (product no. OSHR53E1A2M from Optosupply, Hongkong). The LED emitter was powered with a constant-current source consisting of a simple home-made circuitry with light intensity modulation. The electromotive force (EMF) signal generated by the sensor was recorded using an ordinary multimeter from Voltcraft (model VC820, Germany) connected via an RS-232 interface to data storage PC. The use of low internal resistance voltmeter offers significant enhancement of sensitivity [810].

The miniaturized flow-through sensor was integrated in a sequential injection (SI) fluid handling system composed of a 1-mL syringe pump (Micro BU 2030, Crison Alella, Spain) working as a pressure-driven liquid-driver, an eight-port multiposition selection valve (Crison) and a 45 position rack autosampler (Micro Sampler Crison). The flow network fosters automatic metering of well-defined volumes of solutions at will and delivery to the flow-through optrode. The software package AutoAnalysis 5.0 (Sciware, Palma de Mallorca) based on dynamic link libraries was employed for automated control of the instrumentation including the multimeter and data acquisition as well. The flow manifold is schematically illustrated in Fig. 1a.

Fig. 1
figure 1

Schematic illustration of A PEDD-based chemosensor integrated in a sequential injection (SI) manifold, B close-up of the assembled optrode, C expanded components of the sensor and D magnified sketch of the flow-through optosensor

The analytical procedure involves the sequential aspiration into the holding coil (0.8 mm i.d., 1.5 m-long) of 200 μL of regenerator solution and 200 μL of sample separated by 50 μL carrier (0.2 mol L−1 KH2PO4/K2HPO4, pH 7) working as segment spacer. Thereafter, the stacked segments were delivered by pressure-driven flow to the optosensor at ≤0.8 mL min−1, impinging on the chemoreceptor surface to generate the transient signal, which resembles the typical readout in flow injection assays, followed by backward flowing of liquids to waste. The sequence of segments delivered to the flow-through sensor cell and flow profile are schematically shown in Fig. 1a. The SI setup incorporating the PEDD sensor has been proven suitable for implementation of automated assays of both reducing and oxidizing species without manifold reconfiguration (see below) provided that suitable regeneration/conditioning solutions (namely, 10 mmol L−1 H2O2 in 0.3 mol L−1 KCl at pH 2 for reductants and 10 mmol L−1 ascorbic acid in 0.2 mol L−1 KH2PO4/K2HPO4 buffer at pH 7 for oxidants) are placed at the peripheral ports of the multiposition selection valve.

Results and discussion

Flow-through sensor

For construction of the optosensing device, two identical and commercially available low-budget LEDs (5 mm diameter, with flat front) have been used. The sketch and close-up of the prototype sensor are illustrated in Fig. 1b, c and d. The flow inlet and outlet are drilled through the plastic body of the LED detector (see Fig. 1c and d), generating the so-called flow-cell within LED [22]. The second LED operating as a light emitter is utilized without any mechanical modification. A piece of plastic foil bearing the sensing film is located on the top of the flat bulb of the LED emitter, acting as essential component of the flow-though cell. An elastic o-ring made of silicone rubber is located between the LED detector and film and defines the thickness as well as the inner volume of the resulting optical cell. The sealing of the flow-through cell is accomplished by clamping of LEDs with Lego bricks.

Contrary to PEDD-based photometric detectors for measurement of the absorbance in solution [617], the optical pathlength in the PEDD-based sensor herein presented should be minimized, because the device is dedicated to the optosensing of the reactive film. Besides, the miniaturized detector should be designed to provide maximum interaction of the radiation beam from the LED emitter with the foil sensing area, and prevent light loss, aimed at increasing the signal-to-noise ratio. The architecture of the developed PEDD device illustrated in Fig. 1 fulfills the above requirements: The optical pathlength through the solution (determined by the o-ring thickness) is merely 250 μm. The area of the sensor film contacting with solution is 0.07 cm2 equivalent to a 3-mm diameter sensing microzone. Thus, the total internal volume of the developed flow-through cell housing the chemosensor layer is ca 1.7 μL.

The selection of the pair of LEDs for the PEDD is dictated by the optical properties of the sensing film. In this study the model sensing film is made of PB, featuring deep blue color as a result of the wide absorption band at wavelengths >600 nm. The absorbance spectrum of this compound is compatible with the emission spectrum of red LEDs. Thus, a red LED has been selected as a light emitter. An LED can be merely used as a detector for light of equal or higher energy (low wavelengths) than the light emitted by itself [7, 23]. Therefore, for the construction of PEDD optically compatible with PB-based sensing film, two identical red LEDs have been herein employed.

It is worth to notice that the red LED with drilled flow channels (see Fig. 1c and d) playing the role of detector is light sensitive in the entire visible spectrum (λ ≤ 650 nm) [7, 23]. This denotes the universal nature of the miniature PEDD-based flow-through sensor, which might be furnished with a wide variety of sensing films and LED emitters operating in the visible range. As the flow-through chemosensor solely involves mechanical drilling of the LED detector body, construction of any other type of PEDD-based sensor is readily accomplishable by resorting to unmodified LED emitters optically compatible with the sensing film.

The optical sensitivity of the reported sensor exploits the unique spectroelectrochemical properties of PB [20, 21, 24]. In the sensing scheme herein proposed, PB plays a double role: first as immobilized reagent (oxidant), and second as an optical sensoric receptor. Strong reductants cause discoloration of the sensing material. Reduction of PB by ascorbic acid, used in this study as a model analyte, yields Prussian White (PW, namely, K2FeFe(CN)6). The transparent and colorless PW film does not absorb in the Vis–NIR range of spectrum. The PW film can be easily re-oxidized to PB using various oxidants. Hence, hydrogen peroxide was used as a model regeneration agent. This reversible optical sensing scheme based on the PB/PW redox system was integrated with optically compatible LEDs to demonstrate the figures of merit of the sensor.

In the course of preliminary experiments (data not shown), replicate measurements of ascorbic acid using concentrations of analyte and regenerator at the 10 mmol L−1 level both were weakly reproducible. It was found, that the first three to six peaks are less repeatable as a consequence of the usage of a freshly prepared PB film. As mentioned above, the freshly prepared PB film is composed of the so-called “insoluble PB”. Fortunately, after a few red–ox cycles the sensing film material is converted into a more reactive, so-called soluble form (KFeFe(CN)6) [20] and then consecutive measurements are nicely repeatable. The relative standard deviation for three preconditioned PB films of slightly different thickness bearing the KFeIIIFeII(CN)6/K2FeIIFeII(CN)6 redox system varied from 0.7% to 1.2% (n = 12 peaks for each film/sensor tested). The need of pretreatment of the PB film by conditioning cycles of reduction and oxidation before use for the first time is in agreement with earlier findings [21].

Reductant sensing

An important feature in PEDD sensor operation is the ability of tuning optrode’s sensitivity by modulation of the current powering the LED emitter [8]. This PEDD attribute is well illustrated in Fig. 2 (top), where the readouts for a single standard of ascorbic acid (10 mmol L−1) are obtained at increasing currents from 0.7 to 2.0 mA. This finding is significant in the general context of PEDD sensor development, because it offers the ability of optimization of PEDD-based procedures regardless of the transparency of the sensing layer, which is dictated by the thickness of the film.

Fig. 2
figure 2

Analytical performance of the SI-PEDD-chemosensor for reductant detection. The effect of PEDD powering current (values given at the SIgrams) on the sensor response for 10 mmol L−1 ascorbic acid and the corresponding current–signal dependence (top). Readouts for ascorbic acid determination using the SI-PEDD chemosensor powered at three different currents (given at the SIgrams) and corresponding calibration graphs (middle). Readouts for ascorbic acid determination at three different reaction times (given at the SIgrams) and corresponding calibration graphs (bottom)

Experimental results of sensor calibration performed at three different currents are compiled in Fig. 2 (middle). Both the baseline signal (related to light transmission through the chemosensitive layer) and optrode’s sensitivity are markedly influenced by the LED emitter powering current. High currents (≥1.4 mA) yield calibration graphs with low sensitivity (see Fig. 2 top) because the enhancement effect attainable with the low-resistance EMF-reading instrument is suppressed. Nearly identical results could be obtained using pH meters offering almost currentless conditions for EMF measurement [7, 8]. As to smaller PEDD powering currents (<0.7 mA), which lead to a weak illumination of the LED detector, a low sensitivity is observed because the effect of LED detector discharging by the signal-reading-instrument is stronger than the LED charging by illumination [8]. Intermediate PEDD powering currents (0.7–1.0 mA) are appropriate for monitoring of the chemoreceptor (Fig. 2, top and middle) because the above opposed processes are now of the same magnitude. A linear regression equation within the concentration range of 0.1 to 0.5 mM ascorbic acid (sensitivity, 574 mV/mM; R 2 = 0.999) was found at 0.9 mA (see Fig. 2 middle) with a detection limit (LOD) of 20 μM. Upper quantification limit might be extended up to 1.0 mM by resorting to polynomial regression. The flow-through PEDD sensor features long-term stability with negligible LED corrosion and optrode leaking throughout consecutive assays. Sensor reusability is feasible yet daily recalibration is needed for reliable results because of eventual changes in optical properties of the chemoreceptor.

It is worth to notice that maximum currents commonly recommended for LED driving are more than one order of magnitude higher (ca. 30 mA) than those applied in this work for operation of the optrode under maximum sensitivity (Fig. 2). This observation clearly indicates that the use of more sensitive photodiodes instead of LEDs as detectors in the developed measurement system has no practical use. In other words, the application of LEDs as light detectors suffices for effective operation of the devised PEDD sensor.

Yet another parameter significantly influencing the PEDD sensor response is the reaction time of the sample zone within the sensing film (Fig. 2 bottom). The higher the flow rate of sample delivery to the PB/PW reactive microzone, the higher is the sample throughput, yet at the expenses of decreased sensitivity. In the case of PB sensing, however, saturation of the analytical signal is observed. This effect is caused by the “stoichiometric” nature of the analyte recognition process [20, 21] and limited PB load onto the film surface. The saturation of the signal, which determines the upper concentration level of the dynamic range, is caused by the entire depletion of the receptor phase. As shown in Fig. 2 (bottom), this phenomenon is distinctly observed at low flow rates (≤0.4 mL min−1) and high analyte concentrations (≥1.0 mmol L−1), whereby sample delivery flow rates of 0.8 mL/min were used in acid ascorbic assays.

Under the selected physical and chemical conditions, the PB-film-based PEDD-optrode was utilized for automatic assays of two ascorbic acid-containing formulated dosage forms, namely, Aspirin + C (Quimica Farmaceutica Bayer, SL, Sant Joan Despí, Barcelona) and Supradyn Active Multivitamin (Quimica Farmaceutica Bayer) with manufacturer’s nominal values of 240 mg/unit and 60 mg/unit, respectively. Drug formulations were dissolved or suspended in distilled water and analyzed with the flow-through optrode without further sample processing. Measurement trueness was estimated by resorting to iodometric titration [25] as a reference analytical method endorsed in pharmacopeias. No significant differences at the 0.05 significance level were encountered between the content in vitamin C as measured with the PEDD-optrode (viz., 223 ± 3 mg/unit and 45 ± 5 mg/unit, expressed as mean ± standard deviation, for aspirin + C and Supradyn Active Multivitamin, respectively) and the titrimetric reference method (viz., 226 ± 1 mg/unit and 48 ± 2 mg/unit for aspirin + C and Supradyn Active Multivitamin, respectively) by resorting to a t test for comparison of two experimental means [26]. A striking feature of the miniaturized flow-through optical sensor herein developed is the feasibility of admitting colored and turbid samples, e.g., Supradyn Active Multivitamin water suspension, while rendering unbiased results.

Oxidant sensing

The reduced form of the sensing film (PW) is chemically unstable and easily regenerated to PB. However, the automated SI procedure fosters the in-line generation of this unstable redox state of the sensing layer reliably via pumping of a minute, well-defined plug of reducing agent of appropriate concentration (e.g., 200 μL of 10 mmol/L ascorbic acid). In consequence, the in situ generated PW film can be used for optical sensing of oxidants. In such approach, the roles of oxidant and reductant are reversed: hydrogen peroxide is recognized by the PW-based sensing film as analyte, whereas ascorbic acid plays the role of sensor regenerator. As previously pinpointed for reductant detection (see Fig. 2 top), the current powering PEDD should be, for oxidant sensing, likewise adjusted within the range of 0.6–2.0 mA (Electronic Supplementary Material Fig. S1 top). The current–signal characteristics for hydrogen peroxide sensing are almost the same. The minor variations in the current for maximal sensitivity are caused by slight differences in thickness of the sensor films used throughout this work.

Figure S1 illustrates the non-linear calibration graphs of the PEDD sensor for hydrogen peroxide within the concentration span of 0.1 to 2.0 mM (LOD ≥ 40 μM) as obtained at varied PEDD powering currents (Electronic Supplementary Material Fig. S1 middle) and reaction times (Electronic Supplementary Material Fig. S1 bottom) using sample perfusion flow rates through the measurement system from 0.27 to 0.8 ml/min. These effects are analogous to those earlier reported in this work for reductant detection (Fig. 2). The resulting SI peaks are however inverted because of decreased film transparency when sensing hydrogen peroxide (Electronic Supplementary Material Fig. S1). The slight differences in peak shapes as compared to reductant assays are a consequence of the electrocatalytic nature of hydrogen peroxide reduction by the PB/PW system [20, 21]. Again, the optrode’s sensitivity is mainly dictated by the PEDD powering current, whereas a decrease in flow rate improves sensor detection capability. In brief, the recordings and corresponding calibration graphs compiled in Electronic Supplementary Material Fig. S1 (middle and bottom) clearly demonstrate that the developed flow-through optrode offers relatively fast detection of hydrogen peroxide in the submillimolar range. It is worth to mention that this mode of hydrogen peroxide sensing has potential use in bioassays exploiting oxidoreductases [27, 28].

Conclusions

The dedicated PEDD-based optrode herein reported for flow-through chemical assays of aqueous samples features several advantages. It is an inexpensive sensor system enabling optosensing measurements using merely an ordinary voltmeter and two extremely low-cost LEDs. The total cost of the sensing device does not actually exceed 0.5 euro. Moreover, this miniaturized, and easily customized sensor is fully compatible with highly integrated systems capitalized on mesofluidic lab-on-a-valve and microfluidic lab-on-a-chip devices.

It should be stressed that the flow-through PEDD-optrode concept presented in this paper is of universal applicability. It can be readily expanded to a vast number of sensing schemes with chemosensitive materials, e.g., plasticized polymeric films containing immobilized indicator dyes or chromoionophores. Only careful selection is required for paired LEDs compatible with the optical properties of the sensing membrane along with the adjustment of the PEDD powering current for maximal optrode sensitivity as defined by the thickness and transparency of the sensing layer.