Intra-operative results and radiological outcome of conventional and patient-specific surgery in total knee arthroplasty: a multicentre, randomised controlled trial

  • Knee
  • Published:
Knee Surgery, Sports Traumatology, Arthroscopy Aims and scope

Abstract

Purpose

This prospective, double-blind, randomised controlled trial was designed to address the following research questions: firstly, is there a significant difference in outliers in alignment in the frontal and sagittal plane between PSG TKA and conventional TKA. Secondly, is there a significant difference in operation time, blood loss and length of hospital stay between the two techniques. We hypothesise that there will be fewer outliers with PSG TKA and that operation time, blood loss and length of hospital stay can be significantly reduced with PSG.

Methods

A total of 180 patients were randomised for PSG TKA (group 1) or conventional TKA (group 2) in two centres. Patients were stratified per hospital. Alignment of the mechanical axis of the leg and flexion/extension and varus/valgus of the individual prosthesis components were measured on digital, standing, long-leg and standard lateral radiographs by two independent outcome assessors in both centres. Percentages of outliers (>3°) were determined. We compared blood loss, operation time and length of hospital stay.

Results

There was no statistically significant difference in mean mechanical axis or outliers in mechanical axis between groups. No statistically significant difference was found for the alignment of the individual components in the frontal plane nor for the percentages of outliers. There was a statistically significant difference in outliers for the femoral component in the sagittal plane, with a higher percentage of outliers in the group 1 (p = 0.017). No such significant result was found for the tibial component in that plane. All interclass correlation coefficients were good. Blood loss was 100 mL less in group 1 (p < 0.001). Operation time was 5 min shorter in group 1 (p < 0.001). Length of hospital stay was identical with a mean of 3.6 days (p = 0.657).

Conclusions

The results in terms of obtaining a neutral mechanical axis and a correct position of the prosthesis components did not differ between groups. A small reduction in operation time and blood loss was found with the PSG system. Future research should especially focus on cost-effectiveness analysis and functional outcome of PSG TKA.

Level of evidence

I.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Fig. 1
Fig. 2

Similar content being viewed by others

References

  1. Boonen B, Schotanus MG, Kort NP (2012) Preliminary experience with the patient-specific templating total knee arthroplasty. Acta Orthop 83(4):387–393

    Article  PubMed  Google Scholar 

  2. Chareancholvanich K, Narkbunnam R, Pornrattanamaneewong C (2013) A prospective randomised controlled study of patient-specific cutting guides compared with conventional instrumentation in total knee replacement. Bone Joint J 95-B(3):354–359

    Article  PubMed  CAS  Google Scholar 

  3. Conteduca F, Iorio R, Mazza D, Caperna L, Bolle G, Argento G, Ferretti A (2012) Evaluation of the accuracy of a patient-specific instrumentation by navigation. Knee Surg Sports Traumatol Arthrosc. doi:10.1007/s00167-012-2098-z

    Google Scholar 

  4. Daniilidis K, Tibesku CO (2013) Frontal plane alignment after total knee arthroplasty using patient-specific instruments. Int Orthop 37(1):45–50

    Article  PubMed  Google Scholar 

  5. Kurtz SM, Ong KL, Lau E, Widmer M, Maravic M, Gomez-Barrena E, de Pina MF, Manno V, Torre M, Walter WL, de Steiger R, Geesink RG, Peltola M, Roder C (2011) International survey of primary and revision total knee replacement. Int Orthop 35(12):1783–1789

    Article  PubMed  Google Scholar 

  6. Lustig S, Scholes CJ, Oussedik SI, Kinzel V, Coolican MR, Parker DA (2013) Unsatisfactory accuracy as determined by computer navigation of VISIONAIRE patient-specific instrumentation for total knee arthroplasty. J Arthroplast 28(3):469–473

    Article  Google Scholar 

  7. Mason JB, Fehring TK, Estok R, Banel D, Fahrbach K (2007) Meta-analysis of alignment outcomes in computer-assisted total knee arthroplasty surgery. J Arthroplast 22(8):1097–1106

    Article  Google Scholar 

  8. Moreland JR, Bassett LW, Hanker GJ (1987) Radiographic analysis of the axial alignment of the lower extremity. J Bone Joint Surg Am 69(5):745–749

    PubMed  CAS  Google Scholar 

  9. Nam D, Maher PA, Rebolledo BJ, Nawabi DH, McLawhorn AS, Pearle AD (2013) Patient specific cutting guides versus an imageless, computer-assisted surgery system in total knee arthroplasty. Knee 20(4):263–267

    Article  PubMed  Google Scholar 

  10. Ng VY, DeClaire JH, Berend KR, Gulick BC, Lombardi AV Jr (2012) Improved accuracy of alignment with patient-specific positioning guides compared with manual instrumentation in TKA. Clin Orthop Relat Res 470:99–107

    Article  PubMed  Google Scholar 

  11. Noble JW Jr, Moore CA, Liu N (2012) The value of patient-matched instrumentation in total knee arthroplasty. J Arthroplast 27(1):153–155

    Article  Google Scholar 

  12. Nunley RM, Ellison BS, Ruh EL, Williams BM, Foreman K, Ford AD, Barrack RL (2012) Are patient-specific cutting blocks cost-effective for total knee arthroplasty? Clin Orthop Relat Res 470:889–894

    Article  PubMed  Google Scholar 

  13. Nunley RM, Ellison BS, Zhu J, Ruh EL, Howell SM, Barrack RL (2012) Do patient-specific guides improve coronal alignment in total knee arthroplasty? Clin Orthop Relat Res 470:895–902

    Article  PubMed  Google Scholar 

  14. Otten R, van Roermund PM, Picavet HS (2010) Trends in the number of knee and hip arthroplasties: considerably more knee and hip prostheses due to osteoarthritis in 2030. Ned Tijdschr Geneeskd 154:A1534

    PubMed  Google Scholar 

  15. Spencer BA, Mont MA, McGrath MS, Boyd B, Mitrick MF (2009) Initial experience with custom-fit total knee replacement: intra-operative events and long-leg coronal alignment. Int Orthop 33(6):1571–1575

    Article  PubMed  Google Scholar 

Download references

Acknowledgments

The authors want to thank Jorgen A. Wullems for his work in measuring the alignment on digital standing long-leg radiographs.

Conflict of interest

The authors declare that they have no conflict of interest.

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to B. Boonen.

Rights and permissions

Reprints and permissions

About this article

Cite this article

Boonen, B., Schotanus, M.G.M., Kerens, B. et al. Intra-operative results and radiological outcome of conventional and patient-specific surgery in total knee arthroplasty: a multicentre, randomised controlled trial. Knee Surg Sports Traumatol Arthrosc 21, 2206–2212 (2013). https://doi.org/10.1007/s00167-013-2620-y