Abstract
Accuracy of implant positioning and reconstruction of the mechanical leg axis are major requirements for achieving good long-term results in total knee arthroplasty (TKA). The purpose of the present study was to determine whether image-free computer navigation technology has the potential to improve the accuracy of component alignment in TKA cohorts of experienced surgeons immediately and constantly. One hundred patients with primary arthritis of the knee underwent the unilateral total knee arthroplasty. The cohort of 50 TKAs implanted with conventional instrumentation was directly followed by the cohort of the very first 50 computer-assisted TKAs. All surgeries were performed by two senior surgeons. All patients received the Zimmer NexGen™ total knee prosthesis (Zimmer Inc., Warsaw, IN, USA). There was no variability regarding surgeons or surgical technique, except for the use of the navigation system (StealthStation® Treon plus®, Medtronic Inc., Minnesota, MI, USA). Accuracy of implant positioning was measured on postoperative long-leg standing radiographs and standard lateral X-rays with regard to the valgus angle and the coronal and sagittal component angles. In addition, preoperative deformities of the mechanical leg axis, tourniquet time, age, and gender were correlated. Statistical analyses were performed using the SPSS 15.0 (SPSS Inc., Chicago, IL, USA) software package. Independent t-tests were used, with significance set at P < 0.05 (two-tailed) to compare differences in mean angular values and frontal mechanical alignment between the two cohorts. To compute the rate of optimally implanted prostheses between the two groups we used the χ2 test. The average postoperative radiological frontal mechanical alignment was 1.88° of varus (range 6.1° of valgus–10.1° of varus; SD 3.68°) in the conventional cohort and 0.28° of varus (range 3.7°–6.0° of varus; SD 1.97°) in the navigated cohort. Including all criteria for optimal implant alignment, 16 cases (32%) in the conventional cohort and 31 cases (62%) in the navigated cohort have been implanted optimally. The average difference in tourniquet time was modest with additional 12.9 min in the navigated cohort compared to the conventional cohort. Our findings suggest that the experienced knee surgeons can improve immediately and constantly the accuracy of component orientation using an image-free computer-assisted navigation system in TKA. The computer-assisted technology has shown to be easy to use, safe, and efficient in routine knee replacement surgery. We believe that navigation is a key technology for various current and future surgical alignment topics and minimal-invasive lower limb surgery.
Similar content being viewed by others
References
Bach CM, Steingruber IE, Peer S, Nogler M, Wimmer C, Ogon M (2001) Radiographic assessment in total knee arthroplasty. Clin Orthop Relat Res 385:144–50
Bathis H, Perlick L, Luring C, Kalteis T, Grifka J (2003) CT-based and CT-free navigation in knee prosthesis implantation. Results of a prospective study. Unfallchirurg 106(11):935–40
Berry DJ (2004) Computerassisted knee arthroplasty is better than a conventional jig-based technique in terms of component alignment. J Bone Joint Surg Am 86:2573
Brown GA, Firoozbakhsh K, DeCoster TA, Reyna JR Jr, Moneim M (2003) Rapid prototyping: the future of trauma surgery? J Bone Joint Surg 85(Suppl 4):49–55
Bolognesi M, Hofmann A (2005) Computer navigation versus standard instrumentation for TKA: a single-surgeon experience. Clin Orthop Relat Res 440:162–9
Caillouette JT, Anzel SH (1990) Fat embolism syndrome following the intramedullary alignment guide in total knee arthroplasty. Clin Orthop Relat Res 251:198–9
Chauhan SK, Scott RG, Breidahl W, Beaver RJ (2004) Computerassisted knee arthroplasty versus a conventional jig-based technique. A randomised, prospective trial. J Bone Joint Surg Br 86:372–7
Decking R, Markmann Y, Fuchs J, Puhl W, Scharf HP (2005) Leg axis after computer-navigated total knee arthroplasty: a prospective randomized trial comparing computer-navigated and manual implantation. J Arthroplasty 20(3):282–8
Delp SL, Stulberg SD, Davies B, Picard F, Leitner F (1998) Computer assisted knee replacement. Clin Orthop Relat Res 345:49–56
Dennis DA, Channer M, Susmann MH, Stringer EA (1993) Intramedullary versus extramedullary tibial alignment systems in total knee arthroplasty. J Arthroplasty 8:43
Ecker ML, Lotke PA, Windsor RE, Cella JP (1988) Long-term results after total condylar knee arthroplasty. Significance of radiolucent lines. Clin Orthop Rel Res 216:151–8
Fortheine F, Ohnsorge JAK, Schkommodau E, Radermacher K (2004) CT-based planning and individual template navigation in TKA. In: Stiehl JB, Konermann WH, Haaker RG Navigation and robotics in total joint and spine surgery. Springer, Berlin, pp 336–42
Haaker RG, Stockheim M, Kamp M, Proff G, Breitenfelder J, Ottersbach A (2005) Computer-assisted navigation increases precision of component placement in total knee arthroplasty. Clin Orthop Relat Res (433):152–9
Hafez MA, Chelule KL, Seedhom BB, Sherman KP (2006) Computer-assisted total knee arthroplasty using patient-specific templating. Clin Orthop Relat Res 444:184–92
Ischii Y, Ohmori G, Bechthold J, Gustilo RB (1995) Extramedullary versus intramedullary alignment guides in total knee arthroplasty. Clin Orthop Relat Res 318:167–75
Jeffery RS, Morris RW, Denham RA (1991) Coronal alignment after total knee replacement. J Bone Joint Surg Br 73:709–14
Jenny JY, Clemens U, Kohler S, Kiefer H, Konermann W, Miehlke RK (2005) Consistency of implantation of a total knee arthroplasty with a non-image-based navigation system: a case-control study of 235 cases compared with 235 conventionally implanted prostheses. J Arthroplasty 20(7):832–9
Kalairajah Y, Cossey AJ, Verrall GM, Ludbrook GG, Spriggins AJ (2006) Are systemic emboli reduced in computerassisted knee surgery? A prospective, randomised, clinical trial. J Bone Joint Surg Br 88-B:198–202
Kolettis GT, Wixson RL, Peruzzi WT, Blake MJ, Wardell S, Stuhlberg SD (1994) Safety of 1-stage bilateral total knee arthroplasty. Clin Orthop Relat Res 309:102–9
Laskin RS (2001) The genesis total knee prosthesis: a 10-year follow-up study. Clin Orthop Relat Res (388):95–102
Loer I, Plitz W (2003) Tibial malalignment of mobile-bearing prostheses: a simulator study. Orthopade 32:296–304
Maestro A, Harwin SF, Sandoval MG, Vaquero DH, Murcia A (1998) Influence of intramedullary versus extramedullary alignment guides on final total knee arthroplasty component position: a radiographic analysis. J Arthroplasty 13:552–8
Mahaluxmivala J, Bankes MJK, Nicolai P, Aldam CH, Allen PW (2001) The effect of surgeon experience on component positioning in 673 press fit condylar posterior cruciate—sacrificing total knee arthroplasties. J Arthroplasty 16:635–40
McGrory JE, Trousdale RT, Pagnano MW, Nigbur M (2002) Preoperative hip to ankle radiographs in total knee arthroplasty. Clin Orthop Relat Res 404:196–202
Mont MA, Urquhart MA, Hungerford DS, Krackow KA (1997) Intramedullary goniometer can improve alignment in knee arthroplasty surgery. J Arthroplasty 12(3):332–6
Moreland JR (1988) Mechanics of failure of total knee arthroplasty. Clin Orthop Relat Res 226:49–64
Patel DV, Ferris BD, Aichroth PM (1991) Radiological study of alignment after total knee replacement. Short radiographs or long radiographs?. Int Orthop 15:209–10
Perlick L, Bathis H, Lerch K, Luring C, Tingart M, Grifka J (2005) Revision total knee arthroplasty: a comparison of postoperative leg alignment after computer-assisted implantation versus the conventional technique. Knee Surg Sports Traumatol Arthrosc 13(3):167–73
Petersen TL, Engh GA (1988) Radiographic assessment of knee alignment after total knee arthroplasty. J Arthroplasty 3:67–72
Ranawat CS, Boachie-Adjei O (1988) Survivorship analysis and results of total condylar knee arthroplasty. Clin Orthop Relat Res 226:6–13
Rand JA, Coventry MB (1988) Ten-year evaluation of geometric total knee arthroplasty. Clin Orthop Rel Res 232:168–73
Reed SC, Gollish J (1997) The accuracy of femoral intramedullary guides in total knee arthroplasty. J Arthroplasty 12:677–82
Ritter MA, Faris PM, Keating EM, Meding JB (1994) Postoperative alignment of total knee replacement. Its effect on survival. Clin Orthop 153–6
Rodriguez JA, Bhende H, Ranawat CS (2001) Total condylar knee replacement: a 20-year follow-up study. Clin Orthop Relat Res (388):10–7
Ryd L, Lindtsrand A, Stentstrom A, Selvik G (1990) Porous coated anatomic tricompartmental tibial components. The relationship between position and micromotion. Clin Orthop Relat Res 251:189–97
Siu D, Cooke TD, Broekhiven LD, Lam M, Fisher B, Saunders G, Challis TW (1991) A standardized technique for lower limb radiography. Practice, applications, and error analysis. Invest Radiol 26:71–7
Sparmann M, Wolke B, Czupalla H, Banzer D, Zink A (2003) Positioning of total knee arthroplasty with and without navigation support. J Bone Joint Surg Br 85-B:830–5
Stulberg SD, Loan P, Sarin V (2002) Computer-assisted navigation in total knee replacement: results of an initial experience in thirty-five patients. J Bone Joint Surg Am 90–8
Swanson KE, Stocks GW, Warren PD, Hazel MR, Janssen HF (2000) Does axial limb rotation affect the alignment measurements in deformed limbs?. Clin Orthop Relat Res 371:246–52
Teter KE, Bregman D, Colwell CW Jr (1995) Accuracy of intramedullary versus extramedullary tibial alignment cutting systems in total knee arthroplasty. Clin Orthop Relat Res 5 321:106–10
Tew M, Waugh W (1985) Tibiofemoral alignment and the results of knee replacement. J Bone Joint Surg Am 67B:551–6
Vince KG, Insall JN, Kelly MA (1989) The total condylar prosthesis. Ten- to twelve-year results of a cemented knee replacement. J Bone Joint Surg Br 793–7
Wright JG, Treble N, Feinstein AR (1993) Measurement of lower limb alignment using long radiographs. J Bone Joint Surg Br 75:164–5
Acknowledgment
We thank Gudrun Rosenberger, MD from the Psychiatry Neuroimaging Lab, Harvard Medical School, Boston, MA, USA, for her assistance in statistical analysis.
Author information
Authors and Affiliations
Corresponding author
Rights and permissions
About this article
Cite this article
Rosenberger, R.E., Hoser, C., Quirbach, S. et al. Improved accuracy of component alignment with the implementation of image-free navigation in total knee arthroplasty. Knee Surg Sports Traumatol Arthr 16, 249–257 (2008). https://doi.org/10.1007/s00167-007-0420-y
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s00167-007-0420-y