Intraarticular migration of a broken biodegradable interference screw after anterior cruciate ligament reconstruction

  • Knee
  • Published:
Knee Surgery, Sports Traumatology, Arthroscopy Aims and scope

Abstract

Poly-l-lactic acid biodegradable screws have been used effectively for graft fixation in anterior cruciate ligament (ACL) reconstruction. The overall complication rate associated with the use of this implant is low, although some authors reported complications, such as osteolysis and aseptic effusion of the knee joint. We report a case of a 29-year-old female patient with a failure of a biodegradable interference screw at 22 months after ACL reconstruction using bone–patellar tendon–bone graft. In this illustrated case, the screw broke and migrated into the knee joint. In addition, we performed a detailed review of the medical literature from 1990–2005 to identify possible causes of biodegradable screw failures. We identified six published cases of bioabsorbable interference screw failure with migration into the knee joint. Several authors have reported small diameter of the screw, poor bone quality, bone resorption, and screw divergence as potential causes for intraarticular migration of metallic interference screws. With regard to bioscrews, no specific risk factors for screw breakage and intraarticular migration have been reported. ACL reconstruction with the use of bioabsorbable interference screws for fixation is considered to be reliable. However, we need to be aware of potential problems associated with the use of this implant. Early recognition of bioscrew failure may prevent associated morbidities, such as subsequent cartilage damage.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Fig. 1
Fig. 2

Similar content being viewed by others

References

  1. Barber FA (1999) Tripled semitendinosus-cancellous bone anterior cruciate ligament reconstruction with bioscrew fixation. Arthroscopy 15:360–367

    Article  PubMed  CAS  Google Scholar 

  2. Barber FA, Elrod BF, McGuire DA, Paulos LE (2000) Fixation of patellar tendon autografts. Biomaterials 21:2623–2629

    Article  PubMed  CAS  Google Scholar 

  3. Barrett GR, Papendick L, Miller C (1995) Endobutton endoscopic fixation technique in anterior cruciate ligament reconstruction. Arthroscopy 11:340–343

    Article  PubMed  CAS  Google Scholar 

  4. Benedetto KP, Fellinger M, Lim TE, Passler JM, Schoen JL, Willems WJ (2000) A new bioabsorbable interference screw: preliminary results of a prospective, multicenter, randomized clinical trial. Arthroscopy 16:41–48

    Article  PubMed  CAS  Google Scholar 

  5. Bottoni C, DeBerardino T, Fester E, Mitchell D, Penrod BJ (2000) An intra-articular bioabsorbable interference screw mimicking an acute meniscal tear 8 months after an anterior cruciate ligament reconstruction. Arthroscopy 16:395–398

    Article  PubMed  CAS  Google Scholar 

  6. Bush-Joseph CA, Bach BR Jr (1998) Migration of femoral interference screw after anterior cruciate ligament reconstruction. Am J Knee Surg 11:32–34

    PubMed  CAS  Google Scholar 

  7. Cerullo G, Puddu G (1993) Arthroscopic placement of the interference screw for anterior cruciate ligament reconstruction. Arthroscopy 9:712–713

    Article  PubMed  CAS  Google Scholar 

  8. Doerr AL Jr, Cohn BT, Ruoff MJ, McInerny VK (1990) A complication of interference screw fixation in anterior cruciate ligament reconstruction. Orthop Rev 19:997–1000

    PubMed  Google Scholar 

  9. Karlakki SL, Downes ME (2003) Intra-articular migration of femoral interference screw: open or arthroscopic removal. Arthroscopy 19:19E

    Article  Google Scholar 

  10. Kohn D, Rose C (1994) Primary stability of interference screw fixation. Influence of screw diameter and insertion torque. Am J Sports Med 22:334–338

    Article  PubMed  CAS  Google Scholar 

  11. Lembeck B, Wülker N (2005) Severe cartilage damage by broken poly-l-lactic acid (PLLA) interference screw after ACL reconstruction. Knee Surg Sports Traumatol Arthrosc 13:283–286

    Article  PubMed  Google Scholar 

  12. Lu HH, Cooper JA Jr, Manuel S, Freeman JW, Attawia MA, Ko FK, Laurencin CT (2005) Anterior cruciate ligament regeneration using braided biodegradable scaffolds: in vitro optimization studies. Biomaterials 26:4805–4816

    Article  PubMed  CAS  Google Scholar 

  13. MacDonald P, Arneja S (2003) Biodegradable screw presents as a loose intra-articular body after anterior cruciate ligament reconstruction. Arthroscopy 19: E22–E24

    Article  PubMed  Google Scholar 

  14. Marti C, Imhoff AB, Bahrs C, Romero J (1997) Metallic versus bioabsorbable interference screw for fixation of bone–patellar tendon–bone autograft in arthroscopic anterior cruciate ligament reconstruction. A preliminary report. Knee Surg Sports Traumatol Arthrosc 5:217–221

    Article  PubMed  CAS  Google Scholar 

  15. Martinek V, Friederich NF (1999) Tibial and pretibial cyst formation after anterior cruciate ligament reconstruction with bioabsorbable interference screw fixation. Arthroscopy 15:317–320

    Article  PubMed  CAS  Google Scholar 

  16. Morgan CD, Gehrmann RM, Jayo MJ, Johnson CS (2002) Histologic findings with a bioabsorbable anterior cruciate ligament interference screw explant after 2.5 years in vivo. Arthroscopy 18:E47

    Article  PubMed  Google Scholar 

  17. Rodeo SA, Arnoczky SP, Torzilli PA, Hidaka C, Warren RF (1993) Tendon-healing in a bone tunnel. A biomechanical and histological study in dog. J Bone Joint Surg Am 75-A:1795–1803

    Google Scholar 

  18. Shafer BL, Simonian PT (2002) Broken poly-l-lactic acid interference screw after ligament reconstruction. Arthroscopy 18:E35

    Article  PubMed  Google Scholar 

  19. Sidhu DS, Wroble RR (1997) Intraarticular migration of a femoral interference fit screw. A complication of anterior cruciate ligament reconstruction. Am J Sports Med 25:268–271

    Article  PubMed  CAS  Google Scholar 

  20. Stähelin AC, Weiler A, Rüfenacht H, Hoffmann R, Geissmann A, Feinstein R. (1997) Clinical degradation and biocompatibility of different bioabsorbable interference screws: a report of six cases. Arthroscopy 13:238–244

    Article  PubMed  Google Scholar 

  21. Stähelin AC, Südkamp NP, Weiler A (2001) Anatomic double-bundle posterior cruciate ligament reconstruction using hamstring tendons. Arthroscopy 17:88–97

    Article  PubMed  Google Scholar 

  22. Werner A, Wild A, Ilg A, Krauspe R (2002) Secondary intra-articular dislocation of a broken bioabsorbable interference screw after anterior cruciate ligament reconstruction. Knee Surg Sports Traumatol Arthrosc 10:30–32

    Article  PubMed  CAS  Google Scholar 

Download references

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to M. H. Baums.

Rights and permissions

Reprints and permissions

About this article

Cite this article

Baums, M.H., Zelle, B.A., Schultz, W. et al. Intraarticular migration of a broken biodegradable interference screw after anterior cruciate ligament reconstruction. Knee Surg Sports Traumatol Arthr 14, 865–868 (2006). https://doi.org/10.1007/s00167-006-0049-2

Download citation

  • Received:

  • Accepted:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s00167-006-0049-2

Keywords

Navigation