The correction of stereotactic inaccuracy caused by brain shift using an intraoperative ultrasound device

  • Brain Models and Neurosurgery
  • Conference paper
  • First Online:
CVRMed-MRCAS'97 (CVRMed 1997, MRCAS 1997)

Abstract

Cranial stereotactic systems which utilize preoperative computed tomography (CT) or magnetic resonance imaging (MRI) data sets to guide surgery are subject to inaccuracy introduced by the intraoperative movement of the brain (brain shift). Although these systems allow precise navigation initially during a procedure, brain shift resulting from surgical intervention can lead to progressive degradation in accuracy, with the greatest inaccuracy occurring when deep structures are manipulated. One method of addressing this issue is with the use of an intraoperative scanning device such as CT or MRI; however, such scanners are costly and restrict surgical access. We have developed an alternative intraoperative imaging device consisting of an ultrasound unit coupled to a stereotactic system to quantify the degree of brain shift. This system determines the orientation of ultrasound images produced by the device and reformats the pre-operative CT or MRI images to match the ultrasound image. By comparing the position of specific structures on the two images, the amount of shift can be determined. Furthermore, this system is being expanded to include the aquisition of three-dimensional ultrasonic volumes.

We have used this device on a series of patients (n=23) to determine the position of specific intracranial structures (e.g. vessels, sulci, gyri) prior to and after surgery. Cases in which hematoma or tumors were removed had the highest average shift (9.5mm and 7.9 mm, respectively); whereas, implantation of electrodes for the recording of seizures had the least amount of shift (2.9 mm). As the degree of shift of specific intracranial structures can vary greatly, we have classified the structures into three levels-low (0–2.9 mm), moderate (3.0–6.9 mm), and high (>7.0 mm). In addition, increased age and the use of diuretics during surgery lend to an increase in shift. Using these results, a model of response of the brain to surgery is being developed to correct for brain shift in a automated fashion. We conclude that brain shift can be corrected using an inexpensive intraoperative ultrasound device to improve the accuracy and reliability of stereotactic intracranial surgery.

This is a preview of subscription content, log in via an institution to check access.

Access this chapter

Institutional subscriptions

Preview

Unable to display preview. Download preview PDF.

Unable to display preview. Download preview PDF.

References

  1. De la Porte C: Technical Possibilities and Limitations of Stereotaxy. Acta Neurochirurgica. 124(1):3–6, 1993.

    Article  PubMed  Google Scholar 

  2. Ohye C: The New Age of Stereotaxis: Toward the 21 st Century. Stereotactic and Functional Neurosurgery. 62(1–4): 17–25, 1994.

    PubMed  Google Scholar 

  3. Iseki H., Kawamura H., Tanikawa T., Kawabatake H., Taira T., Takakura K., Dohi T., Hata N: An Image-Guided Stereotactic System for Neurosurgical Operations. Stereotactic and Functional Neurosurgery. 63(1–4): 130–8, 1994.

    PubMed  Google Scholar 

  4. Walton L., Hampshire A., Forster DM., Kemeny AA: Stereotactic Localization Using Magnetic Resonance Imaging. Stereotactic and Functional Neurosurgery. 64 Suppl. 1:155–63, 1995.

    PubMed  Google Scholar 

  5. Bucholz R.D., Smith K.R., Baumann C., McDurmont L., Schulz D.: Intraoperative Localization with an Optical Digitizer. Proceedings of the Xth Meeting of the World Society for Stereotactic and Functional Neurosurgery, Ixtapa, Mexico, Oct. 11–15, 1993. Stereotactic and Functional Neurosurgery. 63:100, 1994.

    Google Scholar 

  6. Zamorano L., Nolte L., Kadi A., Jiang Z: Interactive Intraoperative Localization Using an Infrared-Based System. Proceedings of the Xth Meeting of the World Society for Stereotactic and Functional Neurosurgery, Ixtapa, Mexico, Oct. 11–15, 1993. Stereotactic and Functional Neurosurgery. 63:84–88, 1994.

    PubMed  Google Scholar 

  7. Bucholz R.D., Smith K.R: A Comparison of Sonic Digitizers Versus Light Emitting Diode-Based Localization. Interactive Image-Guided Neurosurgery. 179–200, 1994.

    Google Scholar 

  8. Maurer C., Fitzpatrick J: A Review of Medical Image Registration. Interactive Image-Guided Neurosurgery. 17–44, 1994.

    Google Scholar 

Download references

Author information

Authors and Affiliations

Authors

Editor information

Jocelyne Troccaz Eric Grimson Ralph Mösges

Rights and permissions

Reprints and permissions

Copyright information

© 1997 Springer-Verlag Berlin Heidelberg

About this paper

Cite this paper

Bucholz, R.D. et al. (1997). The correction of stereotactic inaccuracy caused by brain shift using an intraoperative ultrasound device. In: Troccaz, J., Grimson, E., Mösges, R. (eds) CVRMed-MRCAS'97. CVRMed MRCAS 1997 1997. Lecture Notes in Computer Science, vol 1205. Springer, Berlin, Heidelberg. https://doi.org/10.1007/BFb0029268

Download citation

  • DOI: https://doi.org/10.1007/BFb0029268

  • Published:

  • Publisher Name: Springer, Berlin, Heidelberg

  • Print ISBN: 978-3-540-62734-0

  • Online ISBN: 978-3-540-68499-2

  • eBook Packages: Springer Book Archive

Publish with us

Policies and ethics