Abstract
Several brands of laryngoscopes and commonly used blades are evaluated for differences in direct light intensity and area of illumination. The purpose of this study was to determine which laryngoscope/ blade combination provided the best lighting conditions for tracheal intubation. The direct light intensity was assessed by directing various laryngoscope/ blade combinations towards a multifunction exposure meter (Flash Meter III, Minolta Corp.). Light intensity expressed in Lux was calculated using the formula: lux = 2.5 × 2exposure index. The area of illumination provided by the laryngoscopes was assessed by taking photographs when the laryngoscope blade tips were touching a wall. The widest diameter of bright illumination in the photograph was measured in centimeters (cm). The direct light intensity ranged from 0.4-5.5 × 103lux. The field of illumination ranged from 3.2 to 8.8 cm. When fitted on regular size handles, blades with incandescent bulbs provided more intense light than blades using fibreoptic light sources (P = 0.0078). For blades using incandescent light, the use of a short handle laryngoscope resulted in a decrease in light intensity (P = 0.0117). The results of this study suggest that blades using incandescent light provide greater light intensity than blades using fibreoptics.
Résumé
Plusieurs marques de laryngoscope et de lames utilisées couramment sont évaluées au regard de l’intensité lumineuse directe et de la zone éclairée. L’objectif de cette étude consiste à déterminer quel assemblage laryngoscope/ lame fournit les meilleures conditions d’éclairage en vue de l’intubation trachéale. L’intensité lumineuse directe est évaluée en orientant différentes combinaisons de laryngoscopes et de lames vers un photomètre à fonctions multiples (Flash Meter III, Minolta Corp.). L’intensité lumineuse exprimée en Lux est calculée avec la formule LUX = 2,5 X 2index d’exposition.La zone édairée émise par le laryngoscope est évaluée par photographie au moment où l’extrémité de la lame du laryngoscope touche un mur. Le diamètre le plus grand d’éclairage brillant obtenu sur la photographie est mesuré en centimètres (cm). L’intensité lumineuse directe varie de 0,4 à 5,5 × 103 Lux. Le champ d’éclairage varie de 3,2 à 8,8 cm. Lorsque raccordée à des manches de grandeurs régulières, les lames à ampoules incandescentes produisent une lumière plus intense que les lames qui utilisent une source lumineuse à fibres optiques (P = 0,0078). Pour les lames à ampoules incandescentes, l’utilisation d’un laryngoscope à manche court produit une diminution de l’intensité lumineuse (P = 0,0117). Les résultats de cette étude suggèrent que les lames à ampoule incandescente procurent une intensité lumineuse plus grande que les lames à fibres optiques.
Article PDF
Similar content being viewed by others
References
McIntyre JWR. Laryngoscope design and the difficult adult tracheal intubation. Can J Anaesth 1989; 36: 94–8.
Minolta Corporation, Minolta Flash Meter HI Owner’s manual.
Keitz HAE. Light Calculations and Measurements, 2nd ed. London: Macmillan and Co. Ltd., 1971, 79.
Mason J. The International Center of Photography Encyclopedia of Photography. New York: Crown Publishers Inc, 1984, 181.
Buckmaster D. The Focal Encyclopedia of Photography, Vol 1, 2nd ed. New York: Focal Press Ltd., 1965.
Marks RRD, Hancock R, Charters P. An analysis of laryngoscope blade shape and design: new criteria for laryngoscope evaluation. Can J Anaesth 1993; 40: 262–70.
Rusch Inc., Duluth, Georgia: Product Monograph, 1990.
Welch Allyn Inc., Skaneateles Falls, N.Y.: Product Monograph, 1990.
Perel A, Katz E, Davidson JT. Fiberbronchoscopic retrieval of an aspirated laryngoscope bulb. Intensive Care Med 1981; 7: 143–4.
Stehling LC. Management of the airway.In: Barash PG, Cullen BF, Stoelting RK (Eds.). Clinical Anesthesia, 2nd ed. Philadelphia: J.B. Lippincott Company Inc., 1992, 689.
Buckmaster D. The Focal Encyclopedia of Photography, Vol 2, 2nd ed. New York: Focal Press Ltd., 1965.
Mclntyre JWR. Tracheal intubation and laryngoscope design. Can J Anaesth 1993; 40: 193–6.
Author information
Authors and Affiliations
Rights and permissions
About this article
Cite this article
Tousignant, G., Tessler, M.J. Light intensity and area of illumination provided by various laryngoscope blades. Can J Anaesth 41, 865–869 (1994). https://doi.org/10.1007/BF03011596
Accepted:
Issue Date:
DOI: https://doi.org/10.1007/BF03011596