Abstract
A total of 18 patients with clinical suspicion of a pancreatic tumor underwent dynamic contrast-enhanced CT and MRI examinations. A fat-suppressed T1-weighted 2D fast-low-angle-shot (FLASH) sequence and a T2-weighted spin-echo (SE) sequence were applied in a transverse orientation using a circularly polarized (CP) body phased-array coil. The FLASH sequence was repeated after Gd-DTPA administration. The highest spatial resolution was 1.37×1.37×3.00 mm3. In two cases a half Fourier single-shot turbo-SE sequence (HASTE) was additionally applied. In a comparison between CT and MRI, pancreatic masses could be demonstrated and characterized with excellent image quality. The fat-saturated 2D FLASH sequence yielded the highest contrast-to-noise ratios after Gd-DTPA administration between pancreas and inflammatory or neoplastic lesion. One non-contour deforming carcinoma could be detected only with MRI and was only retrospectively visible on CT with an element of uncertainty. Magnetic resonance imaging using a CP body phased-array coil and fat-suppressed T1- and T2-weighted FLASH, SE, and turbo-SE sequences offers diagnostic possibilities in improved imaging of the pancreas.
Similar content being viewed by others
References
Becker N, Frentzel-Beyme R, Wagner G (1984) Atlas of cancer mortality in the Federal Republic of Germany. Springer, Berlin Heidelberg New York
Fontham ETH, Correa P (1989) Epidemiology of pancreatic cancer. Surg Clin North Am 69: 551–567
Müller MF, Meyenberger Ch, Bertschinger Ph, Schaer R, Marcinek B (1994) Pancreatic tumors: evaluation with endoscopic US, CT, and MR imaging. Radiology 190: 745–751
Semelka RC, Kroeker MA, Shoenut JP, Kroeker R, Yaffe CS, Micflikier AB (1991) Pancreatic disease: prospective comparison of CT, ERCP, and 1.5-T MR imaging with dynamic gadolinium enhancement and fat suppression. Radiology 181: 785–791
Megibow AJ, Zhou XH, Rotterdam H, Francis IR, Zerhouni EA, Balfe DM, Weinreb JC, Aisen A, Kuhlmann J, Heuken JP, Gatsonis C, McNeil BJ (1995) Pancreatic adenocarcinoma: CT versus MR imaging in the evaluation of resectibility: report of the Radiology Diagnostic Oncology Group. Radiology 195: 327–3326
Müller-Miny H, Dülks A, Kreft B, Sommer T, Gieseke J, Görich J (1996) MRT des Pankreas— Wertigkeit von Standardgegenüber fettunterdrückenden Pulssequenzen bei 0,5T. Fortschr Röntgenstr 164: 19–24
Hommeyer StC, Freeny PC, Crabo LG (1995) Carcinoma of the head of the pancreas: evaluation of the pancreaticoduodenal veins with dynamic CT: potential in improved accuracy in staging. Radiology 196: 233–238
Vellet AD, Romano W, Bach DB, Passi RB, Taves DH, Munk PL (1992) Adenocarcinoma of the pancreatic ducts: comparative evaluation with CT and MR imaging at 1.5T. Radiology 183: 87–95
Imanishi Y, Hou VY, Chako C, Tempany CMC, Herold Ch J, Zerhouni EA (1994) Evaluation of the pancreas by MRI. Eur Radiol 4: 243–253
Semelka RC, Ascher SM (1993) MR imaging of the pancreas. Radiology 188: 593–602
Hagspiel KD, Neidl KF, Hauser M, Duewell S, Marcinek B (1994) Fettunterdrückte MR-Bildsequenzen in der Diagnostik von Neoplasien der Leber und des Pankreas bei 1,5 Tesla. Fortschr Röntgenstr 160: 235–242
Campeau NG, Johnson CD, Felmlee JP, Rydberg JN, Butts RK, Ehmann RL, Riederer StJ (1995) MR imaging of the abdomen with a phased-array multicoil: prospective clinical evaluation. Radiology 195: 769–776
Author information
Authors and Affiliations
Rights and permissions
About this article
Cite this article
Engelhard, K., Hollenbach, H.P. High-resolution MRI of pancreatic masses with a new circularly polarized body phased-array coil. Eur. Radiol. 7, 643–648 (1997). https://doi.org/10.1007/BF02742917
Received:
Revised:
Accepted:
Issue Date:
DOI: https://doi.org/10.1007/BF02742917