Skip to main content
Log in

DSC calibration in the study of shape memory alloys

  • Published:
Journal of thermal analysis Aims and scope Submit manuscript

Abstract

The unusual mechanical properties (i.e. shape memory effect and superelasticity) of shape memory alloys (SMA) rely on the thermoelastic martensitic transformation (TMT) which is a first-order solid-solid, non-diffusive phase transition, athermal in character.

Differential scanning calorimetry (DSC) is often used as a convenient method of investigating the thermal properties ofSMAs. The common practice of standard temperature calibration, required for a correct instrument performance, is here critically discussed in relation to the study of both the direct exothermic transformation on cooling, and the reverse endothermic transformation on heating in a NiTiSMA. The DSC results show that, with the standard temperature calibration, the instrument is calibrated on heating but un-calibrated on cooling. A general method is advanced to overcome this problem, intrinsically related to the dynamic character of DSC.

Zusammenfassung

Außergewöhnliche mechanische Eigenschaften (z. B. Form-Memoryeffekte und Superelastizität) von Formmemory-Legierungen (SMA) beruhen auf einer thermoelastischen martensitischen Umwandlung (TMT), bei der es sich um eine athermische, nichtdiffusive Feststoff-Feststoff-Phasenumwandlung erster Ordnung handelt.

DSC wird oft als praktische Methode zur Untersuchung der thermischen Eigenschaften von SMAs benutzt. Die übliche Anwendung der für eine präzise Gerätefunktion erforderlichen Standardtemperaturkalibration wird hier in Bezugnahme sowohl auf die direkte exotherme Umwandlung beim Abkühlen als auch auf die umgekehrte endotherme Umwandlung beim Erhitzen von NiTi SMA diskutiert. Die DSC-Ergebnisse zeigen, daß das Gerät durch diese Standardtemperaturkalibration zwar für das Erhitzen, nicht aber für das Abkühlen kalibriert ist. Es wird eine allgemeine Methode zur Überwindung dieses Problemes entwickelt und in Bezug zum dynamischen Charakter der DSC gesetzt.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Similar content being viewed by others

References

  1. M. Hill and P. Nicholas, Metals and Materials, November (1989) 639.

  2. ‘Thermal Analysis-Techniques and Applications’, E. L. Charsley and B. Warrington Eds., The Royal Society of Chemistry, 1992.

  3. ‘Differential Scanning Calorimetry’, J. L. McNaughton and C. T. Mortimer, The Perkin-Elmer Corporation, Norwalk, Connecticut 06856 USA, L-604.

  4. S. C. Mraw, Rev. Sc. Instrum., 53 (1982) 228.

    Article  CAS  Google Scholar 

  5. ASTM E 967-83, 1984 ASTM Annual Book of Standards.

  6. G. W. Höhne and E. Glöggler, Thermochim. Acta, 151 (1989) 295.

    Article  Google Scholar 

  7. G. W. Höhne, H. K. Cammenga, W. Eysel,et al., Thermochim. Acta 160 (1990) 1.

    Article  Google Scholar 

  8. C. H. Ling and R. Kaplow, Met. Trans. A, 11A (1980) 77.

    CAS  Google Scholar 

  9. C. M. Wayman, Proc. Int. Symp. on Shape Memory Alloys, Eds. C. Youyi, T. Hsu and T. Ko, China Acad. Pub., 1986, p. 59.

  10. S. Miyazaki, Y. Igo and K. Otsuka, Acta Metall., 34 (1986) 2045.

    Article  CAS  Google Scholar 

  11. M. B. Salamon, M. E. Meichle and C. M. Wayman, Phys. Rev. B, 31 (1985) 7306.

    Article  CAS  Google Scholar 

Download references

Author information

Authors and Affiliations

Authors

Rights and permissions

Reprints and permissions

About this article

Cite this article

Airoldi, G., Riva, G., Rivolta, B. et al. DSC calibration in the study of shape memory alloys. Journal of Thermal Analysis 42, 781–791 (1994). https://doi.org/10.1007/BF02546750

Download citation

  • Issue Date:

  • DOI: https://doi.org/10.1007/BF02546750

Keywords

Navigation