Abstract
A new method is presented to calculate with improved accuracy the absolute peak efficiency of cylindrical Ge and Ge(Li) detectors for point, disk and cylinder sources, positioned at any source-detector distance. Moreover attention was paid to true-coincidence effects. The method is extensively tested and applied for the analysis of reference materials. The accuracy turned out to be 3% or better.
Similar content being viewed by others
References
U. PROESCH, M. LUTHARDT, L. KOLDITZ, Isotopenpraxis, 15 (1979) 297.
C. C. GROSJEAN, Nucl. Instr. Methods, 17 (1962) 289.
R. L. HEATH, Scintillation Spectrometry, Gamma-Ray Spectrum Catalogue, 2nd ed., USAEC Report IDO-16880, 1964.
D. C. CAMP, A. L. VAN LEHN, Nucl. Instr. Methods, 76 (1969) 192.
R. GRIFFITHS, Nucl. Instr. Methods, 91 (1971) 377.
R. GUNNINK, J. B. NIDAY, Computerized Quantitative Analysis by Gamma-Ray Spectrometry Vol. 1 Description of the Gammanal Program, USAEC Report UCRL-51061, 1972.
J. E. CLINE, A Technique of Gamma-Ray Detector Absolute Efficiency Calibration for Extended Sources, Proc. of the Am. Nucl. Soc. Topical Conference at Mayaguez, Puerto Rico, 1978, CONF-780421, 1979, p. 185.
MASAYASU NOGUCHI, KENJI TAKEDA, HIDEA HIGUCHI, Int. J. Appl. Radiat. Isotopes, 32 (1981) 17.
G. AUBIN, J. BARRETTE, G. LAMOUREUX, S. MONARO, Nucl. Instr. Methods, 76 (1969) 85.
S. N. KAPLANIS, Int. J. Appl. Radiat. Isotopes, 29 (1978) 543.
BIAGIO ARCIPIANI, EDMONDO PEDRETTI, Atomkernenergie/Kerntechnik, 35 (1980) 210.
I. R. WILLIAMS, Nucl. Instr. Methods, 44 (1966) 160.
L. MOENS, J. DE DONDER, LIN XILEI, F. DE CORTE, A. DE WISPELAERE, A. SIMONITS, J. HOSTE, Nucl. Instr. Methods, 187 (2,3) (1981).
D. S. ANDREEV, K. I. EROKHINA, V. S. ZVONOV, I. Kh. LEMBERG, Instr. Expt. Techn., 15 (1972) 1358.
L. MOENS et al., (in preparation).
A. SIMONITS, L. MOENS, F. DE CORTE, A. DE WISPELAERE, A. ELEK, J. HOSTE, J. Radioanal. Chem., 60 (1980) 461.
L. MOENS, F. DE CORTE, A. SIMONITS, A. DE WISPELAERE, A. ELEK, J. HOSTE, (in preparation).
L. MOENS, Doctorate Thesis, Rijksuniversiteit Gent, 1981.
LIN XILEI, Doctorate Thesis, Rijksuniversiteit Gent, 1981.
Author information
Authors and Affiliations
Additional information
Research associate of the “Nationaal Fonds voor Wetenschappelijk Onderzoek”.
Rights and permissions
About this article
Cite this article
Moens, L., De Corte, F., Simonits, A. et al. Calculation of the absolute peak efficiency of Ge and Ge(Li) detectors for different counting geometries. J. Radioanal. Chem. 70, 539–550 (1982). https://doi.org/10.1007/BF02516136
Received:
Issue Date:
DOI: https://doi.org/10.1007/BF02516136