Abstract
An activation method was used to determine the mercury content of typical laboratory reagents and materials, in particular, filter paper and substances used in the analysis for mercury. Typical values for filtering materials are in the tens of nanogram range but in some types they can be considerably higher. Results for some other reagents also show that a thorough checking for background levels of substances used in analysis is essential. A minor modification of the volatilization method is also described for the analysis of substances rich in sulfur or evolving large quantities of halogen acid.
Similar content being viewed by others
References
K. Irukayama, 3rd Intern. Conf. on Water Pollution Research, 1966, Paper No. 8, Water Pollution Control Federation, Washington, D. C.
G. Lofröth, Methylmercury, Ecological Research Committee, Bull. No. 4, Swedish Nat. Sci. Res. Council. Stockholm, Sweden, March 1969.
L. Kosta, A. R. Byrne,Talanta, 16 (1969) 1297.
H. J. M. Bowen,Analyst, 92 (1967) 124.
J. Ružička, C. G. Lamm,Talanta, 16 (1969) 157.
E. B. Sandell, Colorimetric Determination of Traces of Elements, Interscience, New York, 1959, p. 625.
Author information
Authors and Affiliations
Rights and permissions
About this article
Cite this article
Byrne, A.R., Dermelj, M. & Kosta, L. The mercury content of some laboratory chemicals and materials determined by neutron activation analysis. J. Radioanal. Chem. 6, 325–330 (1970). https://doi.org/10.1007/BF02513961
Received:
Revised:
Issue Date: