Abstract
We studied the effects of tube potentials and exposures on the accuracy of carious diagnosis using Sens-A-Ray direct digital radiography with two types of CCD detectors, one without and one with a scintillator layer. Our second objective was to determine the optimum exposure factors.
Radiographs of 75 extracted premolars with and without clinically observed discolorations or cavities were obtained at exposure settings of 70 kVp and 90 kVp. The observers diagnosed dental caries in radiographs exposed with the two types of detectors. The results were analyzed with Receiver Operating Curve (ROC) technique. The averaged areas under the ROC curves were plotted as functions of exposure to obtain Diagnostic Accuracy (DA) curves.
No statistically significant difference was found in diagnostic accuracy for the two tube potentials. However, there was a statistically significant difference in diagnostic accuracy for the two lesion types, discolorations and cavities. There was also a significant effect on diagnostic accuracy related to exposure: the exposure providing the optimum diagnostic accuracy corresponds to a gray level of about 31 when SVGA graphics are employed, i.e., in the middle of the available range.
To acquire optimum radiographs for diagnosing carious lesions with the Sens-A-Ray system, the operator must select the appropriate exposure parameters.
Similar content being viewed by others
References
Svenson, B., Welander, U., Anneroth, G. and S derfeldt, B.: Exposure parameters and their effects on diagnostic accuracy.Oral Surg Oral Med Oral Pathol. 78: 544–50, 1994.
Webber, R., Benton, P. and Ryge, G.: Diagnostic variations in radiographs.Oral Surg Oral Med Oral Pathol 26: 800–9, 1968.
Arnold, L. The influence of exposure conditions (tubevoltage, filtration and exposure time) on the detection of lesions in the proximal surfaces. The radiographic detection of initial carious lesions on the proximal surfaces of teeth. pp93–147. Diss. The University of Utrecht Dental School. The Netherlands. (Alphen aan den Rijn/Brussel: Stafleu & Tholen, 1983), 1983.
Okano, T., Huang, H.-J. and Nakamura, T.: Diagnostic accuracy on detection of proximal enamel lesions in nonscreen radiographic performance.Oral Surg Oral Med Oral Pathol. 59: 543–7, 1985.
Svenson, B., Gröndahl, H.-G., Petersson, A. and Olving, A.: Accuracy of radiographic caries diagnosis at different kilovoltages and two film speeds.Swed Dent J. 9: 37–43, 1985.
Gröndahl, H.-G.: Some factors influencing observer performance in radiographic caries diagnosis.Swed Dent J. 3: 157–72, 1979.
Price, C. and Ergül, N: A comparison of a film-based and a direct digital dental radiographic system using a proximal caries model.Dentomaxillofac Radiol 26: 45–52, 1997.
White, S. and Yoon, D.: Comparative performance of digital and conventional images for detecting proximal surface caries.Dentomaxillofac Radiol. 26: 32–38, 1997.
Gotfredsen, E., Wenzel, A. and Gröndahl, H.-G.: Observers' use of image enhancement in assessing caries in radiographs taken by four intra-oral digital systems.Dentomaxillofac Radiol. 25: 34–8, 1996.
Møystad, A., Svanæs, D., Risnes, S., Larheim, T. and Gröndahl, H.-G.: Detection of approximal caries with a storage phosphor system. A comparison of enhanced digital images with dental X-ray film.Dentomaxillofac Radiol. 25: 202–6, 1996.
Borg, E. and Gröndahl, H.-G.: On the dynamic range of different X-ray photon detectors in intra-oral radiography. A comparison of image quality in film, charge-coupled device and storage phosphor systems.Dentomaxillofac Radiol. 25: 82–8, 1996.
Price, C.: The effects of beam quality and optical density on image quality in dental radiography.Oral Surg Oral Med Oral Pathol. 62: 580–8, 1986.
De Belder, M., Bollen R. and Duville, R.: A new approach to the evaluation of radiographic systems. J Phot Sc. 19: 126–31, 1971.
Welander U., Nelvig, P., Tronje, G., McDavid, W., Dove S., Mörner, A.-C. and Cederlund, T.: Basic technical properties of a system for direct acquisition of digital intraoral radiographs.Oral Surg Oral Med Oral Pathol. 75: 506–16, 1993.
Welander, U., McDavid, W., Mörner, A., Tronje, G. Tokuoka, O., Fuchihata, H., Nelvig P. and Dove, S.: Absolute measures of image quality for the Sens-A-Ray direct digital intraoral radiography system.Oral Surg Oral Med Oral Pathol Oral Radiol Endod. 80: 345–50, 1995.
Swets, J. and Pickett, R.: Methods from signal detection theory. 1982, Academic Press Inc., New York.
Norusis, M.: SPSS Base System User's Guide. 1990, SPSS, Chicago, IL.
Svenson, B., Welander, U., Shi, X.-Q., Stamatakis, H. and Tronje, G.: A sensitometric study of four dental X-ray films and their diagnostic accuracy.Dentomaxillofac Radiol. 26: 230–5, 1997.
Møystad, A., Svanæs, D., Larheim, T. and Gröndahl, H.-G.: Effect of image magnification of digitized bitewing radiographs on approximal caries detection: an in vitro study.Dentomaxillofac Radiol. 24: 255–9, 1995.
Wenzel, A. and Verdonshot, E.: Some considerations in the evaluation of diagnostic tests in dentistry.Dentomaxillofac Radiol. 23: 179–82, 1994.
Beideman, R., Pettigrew, J. and Green, P.: A follow-up study of a third-party radiographic evaluation system.Oral Surg Oral Med Oral Pathol. 23: 103–8, 1983.
Eliasson, S., Lavstedt, S., Wouters, F. and Östlin, L.: Quality of intraoral radiographs sent by private dental practitioners for therapy evaluation by the Social Insurance Office.Swed Dent J. 14: 81–9, 1990.
Kantor, M., Hunt, R. and Morris, A.: An evaluation of radiographic equipment and procedures in 300 dental offices in the United States.J Am Dent Assoc. 120: 547–50, 1990.
Rushton, V. and Horner, K.: A comparative study of radiographic quality with five periapical techniques in general dental practice.Dentomaxillofac Radiol. 23: 34–45, 1994.
Svenson, B., Eriksson, T., Kronström, M. and Palmqvist, S.: Image quality of intraoral radiographs used by general practitioners in prosthodontic treatment planning.Dentomaxillofac Radiol. 23: 46–8, 1994.
Svenson, B., Eriksson, T., Kronström, M. and Palmqvist, S.: Quality of intraoral radiographs used for prosthodontic treatment planning by general dentists in the Public Dental Health Service.Swed Dent J. 19: 47–54, 1995.
Author information
Authors and Affiliations
Rights and permissions
About this article
Cite this article
Svenson, B., Shi, XQ., Stamatakis, H. et al. Exposure parameters in sens-a-ray direct digital radiography and their effects on diagnostic accuracy of carious lesions. Oral Radiol. 13, 83–91 (1997). https://doi.org/10.1007/BF02492573
Received:
Revised:
Accepted:
Issue Date: