Summary
The work presented here compares the performance of slit with video densitometry. Linearity, precision and detection limits were studied for some test dyes. Slit densitometry was performed using commercial TLC apparatus and video densitometry using a conventional flat-bed scanner with commercial software. Results show that video densitometry has equivalent precision compared with slit densitometry, however, it has a broader linear range and better sensitivity for the compounds studied. Other attributes of video densitometry are highlighted.
References
Essig, S.: Jehle, H.; Kovar, K.A.; Renger, B.1st International Meeting on Imaging Techniques in Planar Chromatography, Jezersko, Slovenia,May 1999.
Vovk, I.; Prosek, M.J. Chromatogr. 1997,779, 329–336.
Vovk, I.; Prosek, M.J. Chromatogr. 1997 768, 329–333.
Petrovic, M.; Kastelan-Macan, M.; Lazaric, K.; Babic, S.J. AOAC Int. 1999,82, 25–30.
Vovk, I.1st International Meeting on Imaging Techniques in Planar Chromatography, Jezersko, Slovenia,May 1999.
Teasdale, P.R.; Hayward, S.; Davison, W.Anal. Chem. 1999,71, 2186–2191.
Summanen, J.; Yrjonen, T.; Hiltunen, R.; Vuorela, H.J. Planar Chromatogr. 1998,11, 421–427.
Essig, S.; Jehle, H.; Kovar, K.A.J. Planar Chromatogr. 1999,12, 63–65.
Ebel, S.; Henkel, T.J. Planar Chromatogr. 2000,13, 248–253.
Diamond, D.; Byrne, L.; Barker, J.; Pennarum-Thomas, G.Trends Anal. Chem. 2000,19, 517–522.
Author information
Authors and Affiliations
Rights and permissions
About this article
Cite this article
Mustoe, S.P., McCrossen, S.D. A comparison between slit densitometry and video densitometry for quantitation in thin layer chromatography. Chromatographia 53 (Suppl 1), S474–S477 (2001). https://doi.org/10.1007/BF02490381
Received:
Accepted:
Issue Date:
DOI: https://doi.org/10.1007/BF02490381