Studies in ultrasonic echoencephalography—VIII: The effects on resolution of irregularities in attenuation of an ultrasonic beam traversing cadaver skull

  • Published:
Medical and biological engineering Aims and scope Submit manuscript

Abstract

It is found that the cadaver skull attenuates the ultrasonic beam to a greatly varying degree. Variations in attenuation of up to 60 dB for echoes returned to the transducer may result from quite small movements of the transducer relative to the skull. Such variations in attenuation may introduce another type of false localization of echoing interfaces when receiver sensitivity is restricted.

Sommaire

On a trouvé que le crâne d'un cadavre atténue un faisceau d'ultrasons à un degré qui varie beaucoup. Des variations jusqu'à 60 dB dans l'atténuation sur les échos revenus au capteur peuvent résulter de mouvements très petits de celui-ci par rapport au crâne. De telles variations dans l'atténuation peuvent introduire un type particulier de mauvaise localisation des échos provenant des structures profondes lorsque la sensibilité du réceptuer est réduite.

Zusammenfassung

Es wird beobachtet, daß der Leichenschädel den Ultraschallstrahl zu einem sehr untersuchiedlichen Grad abschwächt. Für die zum Transducer reflektierten Echos können Abweichungen von bis zu 60 dB durch sehr geringe Bewegungen des Transducers auf dem Schädel verursacht werden. Derartige Abweichungen in der Abschwächung können ienen neuen Typ falscher Lokalisation echogebender Grenzflächen erzeugen, wenn die Empfängerempfindlichkeit begrenzt ist.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Similar content being viewed by others

References

  • White, D. N., Clark, J. M. andWhite, M. N. (1967) Studies in ultrasonic echoencephalography—VII: General principles of recording information in ultrasonic B- and C-scanning and the effects of scatter, reflection and refraction by cadaver skull on this information.Med. biol. Engng,5, 3–14.

    Google Scholar 

Download references

Author information

Authors and Affiliations

Authors

Rights and permissions

Reprints and permissions

About this article

Cite this article

White, D.N., Clark, J.M. & White, M.N. Studies in ultrasonic echoencephalography—VIII: The effects on resolution of irregularities in attenuation of an ultrasonic beam traversing cadaver skull. Med. & biol. Engng. 5, 15–23 (1967). https://doi.org/10.1007/BF02478837

Download citation

  • Received:

  • Issue Date:

  • DOI: https://doi.org/10.1007/BF02478837

Keywords

Navigation