An energy-time analysis of ventricular fibrillation and defibrillation thresholds with internal electrodes

  • Published:
Medical and biological engineering Aims and scope Submit manuscript

Abstract

The paper describes experimentation with the intracardiac production and termination of ventricular fibrillation, the ultimate goal being the design of an automatic implantable defibrillation device. Fibrillation and defibrillation thresholds were identified by means of increasing electrical-shock energies on dogs via internal electrodes. A maximum energy, called VF type I (VF1), has been demonstrated and is defined as the highest electrical-shock energy which can evoke a resolvable and non-lesional fibrillation. Energies of a much higher order of magnitude produced a lesional and therefore unresolvable type of ventricular fibrillation (VF11). The optimal defibrillation threshold is identified as the smallest shock energy which terminated VF1 on all attempts. The high defibrillation energies (i.e 16–40 J) reported here and elsewhere are due, in part, to the relatively low resistivity of the blood compartment, which tends to short-circuit the electrical shock. In conclusion, a VF1 window is mapped-out on an energy-time diagram, and a myocardial synchronous response threshold is proposed as being equivalent to both the maximum VF1 energy and the optimal defibrillation threshold.

Sommaire

Ce mémoire décrit des expériences avec la création et la terminaison intracardiaques de fibrillation des ventricules (VF), l'objet final étant l'étude d'un dispositif de défibrillation implantable automatique. Les seuils de fibrillation et de défibrillation ont été identifiés par application d'énergies de choc électrique croissantes sur des chiens par l'intermédiaire d'électrodes internes. Une énergie VF1 maximum a été démontrée et définie comme étant le plus grand choc électrique qui puisse évoquer une fibrillation résoluble et non-lésionnelle qu'on appelle VF type 1 (VF1). Des énergies d'une grandeur bien plus importante ont produit une VF lésionnelle, donc non résoluble (soit VF11). Le seuil de défibrillation optimum est défini comme étant la plus petite énergie de choc qui ait terminé VF1 à chaque tentative. Les fortes énergies de défibrillation (16–40 J) dont on fait état ici et ailleurs sont dûes, en partie, à la résistivité relativement basse du compartiment sanguin qui tend à court-circuiter le choc électrique. En conclusion, une ‘fenêtre’ VF1 est tracée sur un schéma énergietemps, et on propose un seuil de réponse synchrone myocardiaque comme étant équivalent à l'énergie VF1 maximum et au seuil de défibrillation maximum.

Zusammenfassung

Die Arbeit beschreibt Experimente mit Erzeugung und Abstellen von Ventrikelfaserbildung im Herzinneren. Das Ziel besteht letztlich darin, eine automatische, einpflanzbare Vorrichtung für die Entfaserung zu konstruieren. Es werden die Schwellen für Faserbildung und Entfaserung durch wachsende Elektroschockenergien an Hunden über interne Elektroden bestimmt. Eine maximale Energie von VF1 wurde dargestellt und wird als die höchste Elektroschockenergie bezeichnet, die eine auflösbare und keine Verletzungen herbeiführende Faserbildung, die man VF Typ 1 (VF1) nennt, herbeiführt. Energien einer viel höheren Grö\enordnung erzeugten eine verletztenden und daher unauflösliche VF (d.h. VF11). Die optimale Entfaserungsschwelle wird als die geringste Schockenergie bestimmt, die bei allen Versuchen VF1 beendigte. Die hohen Entfaserungsenergien (d.h. 16–40 J), von denen hier und anderswo berichtet wird, sind teilweise auf den relativ niedrigen spezifischen Widerstand der Blutzelle zurückzuführen, die dazu neigt, den Elektroschock kurzzuschließen. Also Schlußfolgerung ist ein VF1-‘Fenster’ auf einem Energie-Zeitdiagramm dargestellt, und es wird angeregt, daß eine myokardiale synchrone Reaktionschwelle sowohl der maximalen VF1-Energie und der optimalen Entfaserungsschwelle entspricht.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Similar content being viewed by others

References

  • Bouvrain, Y., Saumont, R., Fabiato, A., Brunet, J. P., Coumel, P. andLesigne, C. (1966) Etude expérimentale des chocs utilisés pour la défibrillation transthoracique.Arch. Mal. Coeur 4, 495–504.

    Google Scholar 

  • Bouvrain, Y., Saumont, R. andBerthier, Y. (1967) Etude clinique de la fibrillation ventriculaire. Résultats des chocs électriques transthoraciques.Ibid 2, 213–231.

    Google Scholar 

  • Coumel, P., Brunet, J. P., Fabiato, A., Lesigne, C. andSaumont, R. (1965a) Conditions de déclenchement de la fibrillation ventriculaire par les chocs électriques.Réanimation et organes artificiels 2, 61–73.

    Google Scholar 

  • Coumel, P., Brunet, J. P., Fabiato, A. andSaumont, R. (1965b) Période vulnérable et fibrillation ventriculaire.Semaine des Hôpitaux 41, 2760.

    Google Scholar 

  • Dessertenne, F. (1964) La fibrillation ventriculaire. Actualités cardiologiques et angéiologiques internationales.4.

  • Ewy, G. A., Ewy, M. D., Nuttall, Q. J. andNuttall, W. W. (1972) Canine transthoracic resistance.J. Appl. Physiol 32, 91–94.

    Google Scholar 

  • Fabiato, A., Coumel, P., Gourgon, R. andSaumont R. (1967) Le seuil de réponse synchrone des fibres myocardiques. A Application à la comparaison expérimentale de l'efficacité des différentes formes de choc électrique de défibrillation.Arch. Mal. Coeur 60, 527.

    Google Scholar 

  • Fabiato, A. (1970) Désynchronisation et resynchronisation dans le syncitium fonctionnel myocardique. Aspects électrophysiologiques et mécaniques. D.Sc. thesis, Université Paris-Sud.

  • Guyton, A. C. andSatterfield, J. (1951) Factors concerned in electrical defibrillation of the heart, particularly through the unopened chest.Amer. J. Physiol. 167, 81.

    Google Scholar 

  • Hoffman, B. F., Suckling, E. E. andMcBrooks, C. (1954) Vulnerability of the dog ventricle and effects of defibrillation.Circ. Res. 3, 147–151.

    Google Scholar 

  • Mirowski, M., Mower, M. M., Gott, V. L., Brawley, R. K. andDenniston, R. (1973a) Transvenous automatic defibrillator. Preliminary clinical tests of its defibrillating subsystem.Trans. Amer. Soc. Artif. Int. Organ. 18, 520–525.

    Google Scholar 

  • Mirowski, M., Mower, M. M., Gott, V. L. andBrawley, R. K. (1973b) Feasibility and effectiveness of low-energy catheter defibrillation in man.Circulation 47, 79–85.

    Google Scholar 

  • Saumont, R., Brunet, J. P., Fabiato, A., Coumel, P. andBouvrain, Y. (1965a) Exploration du cycle cardiaque par chocs dirigés.Arch. Mal. Coeur 2, 246–258.

    Google Scholar 

  • Saumont, R., Fabiato, A., Brunet, J. P., Coumel, P., Lesigne, C. andBouvrain, Y. (1965b) Exploration du cycle cardiaque par chocs dirigés en courant alternatif.Ibid. 9, 1296–1310.

    Google Scholar 

  • Saumont, R. andGourgon, R. (1969) La défibrillation par choc électrique. Colloques de l'inserm Arrêt cardiaque et circulatoire,1, 87–101.

    Google Scholar 

  • Saumont, R. andBouvrain, Y. (1974) Defibrillation cardiaque automatique.La Nouvelle Presse Méd. 3, 10.

    Google Scholar 

  • Saumont, R., Lenne, W., Lesigne, C., Levy, B., Bardou, A., andBirkui, P. (1974) Colloque biologie et fonctions du myocarde 7–8. Délégation Générale à la Recherche Scientifique et Technique.

  • Schuder, J. C., Gold, J. H., andStoeckle, H. (1972) A multielectrode-time sequential laboratory defibrillator.Trans. Amer. Soc. Artif. Int Organs 18, 514–518.

    Google Scholar 

  • Schuder, C., Stoeckle, H., Gold, J. H., West, J. A. andKeskar, P. Y. (1970) Experimental ventricular defibrillation with an automatic and completely implanted system.Ibid. 16, 207–212.

    Google Scholar 

  • Schuder, J. C., Stockle, H., West, J. A. andKeskar, P. Y. (1973) Relationship between electrode geometry and effectiveness of ventricular defibrillation in the dog with catheter having one electrode in right ventricle and other electrode in superior vena cava, or external jugular vein, or both.Cardiovascular Res. 7, 629–637.

    Article  Google Scholar 

  • Surawicz, B. (1971) Ventricular fibrillation.Amer. J. Cardiol. 28, 268–287.

    Article  Google Scholar 

Download references

Author information

Authors and Affiliations

Authors

Rights and permissions

Reprints and permissions

About this article

Cite this article

Lesigne, C., Levy, B., Saumont, R. et al. An energy-time analysis of ventricular fibrillation and defibrillation thresholds with internal electrodes. Med. & biol. Engng. 14, 617–622 (1976). https://doi.org/10.1007/BF02477038

Download citation

  • Received:

  • Accepted:

  • Issue Date:

  • DOI: https://doi.org/10.1007/BF02477038

Keywords

Navigation