Abstract
Synthetic materials used for medical prostheses and devices, mainly those which will come into contact with blood, have to provide for specific properties. Three in vitro blood-compatibility tests are presented; one is a screening method (m.t.e.g.) whose results show remarkable correlations to in vivo behaviour, the other two (Nosé-blood-chamber, protein adsorption) yield more basis information on synthetic surfaces. Aspects of toxicity and biodegradation should be included in the definition of ‘blood compatibility’. After preselection by in vitro tests, a material has to pass an in vivo examination in animal experiments (shunt, catheter, tube, test chamber) under haemodynamic conditions. Finally, devices or prostheses in their intended shape, such as the total artificial heart, are implanted in animals.
Similar content being viewed by others
References
Affeld, K., Berger, J., Müller, R. andBücherl, E. S. (1974) A new method for an in vitro test of bloodcontact materials.Proc. Europ. Soc. Artif. Organs,1, 26–29.
Brash, J. L. (1979)Hydrophobic polymers as materials for interfacing with blood.Unger, F. (Ed.): Assisted Circulation, Springer-Verlag, 506–519.
Bruck, S. D. (1979)Some current problems and new dimensions of polymeric biomaterials for blood contacting applications.Unger, F. (Ed.): Assisted Circulation, Springer-Verlag, 496–502.
Bücherl, E. S.,et al. (1979) Total artificial heart replacement.Internat. J. Artif. Organs,2, 141.
Carr, S. H., Zuckermann, L. andCaprini, J. A. (1979) In vitro testing of surface thrombogenicity using the thrombelastograph. 80th National Meeting of the American Institute of Chemical Engineers.
Dudley, B., Williams, J. L., Able, K. andMuller, B. (1976) Synthesis and characterization of blood compatible surfaces.Trans. Am. Soc. Artif. Int. Organs,22, 538.
Gott, V. I. andFuruse, A. (1971) Antithrombogenic surfaces, classification and in vivo evaluation.Fed. Proc.,30, 1679.
Grosse-Siestrup, C., Unger, V. andLemm, W. (1978) A long-term ex vivo shunt for blood compatibility tests of biomaterials.Trans. Soc. Biomater.,2, 102.
Grosse-Siestrup, C., Unger, V., Lemm, W. andOlsen, D. B. (1978) Externer arteriovenöser Langzeitshunt mit Testkammer für Ex-vivo-Blutverträglichkeitsuntersuchungen von Polymeroberflächen.Biomed. Technik.,23, 173.
Grosse-Siestrup, C., Krautzberger, W. andBücherl, E. S. (1979) Arteriovenous shunt with percutaneous leads in calves for long-term in vivo testing of blood-contact materials.J. Biomed. Mat. Res.,13, 15–22.
Grosse-Siestrup, C., Warnecke, I., Gerlach, K., Lemm, W. (1979) Vergleichende Untersuchung von Cava-Kathetern aus vier Polymermodifikationen.Biomed. Technik.,24, 295–296.
Kusserow, B. K., Larrow, R. andNichols, J. (1970) Observations concerning prosthesis-induced thromboembolic phenomena made with an in vivo embolus test system.Trans. Am. Soc. Artif. Int. Org.,16, 58.
Kaiser, M. andBücherl, E. S. (1979) In vitro tests of thrombogenicity protein adsorption and hemolytic properties of biomaterials.Proc. Europ. Soc. Artif. Organs,6, 203.
Lemm, W., Unger, V., Grosse-Siestrup, C., Olsen, D. B. andBücherl, E. S. (1977) Blood compatibility tests on biomaterials on the basis of the proteinadsorption-hypothesis. Extensions and comparisons with other methods.Proc. Europ. Soc. Artif. Organs,4, 168.
Lemm, W. andUnger, V. (1978) Proteinadsorption an verschiedenen Kunststoffoberflächen in vitro.Biomed. Technik.,23, (Erg.-Band), 171.
Lemm, W., Unger, V., Kaiser, M., Grosse-Siestrup, C. andBücherl, E. S. (1978) A comparing evaluation of three blood compatibility tests for biomaterials.Proc. Europ. Soc. Artif. Orangs,5, 29.
Lemm, W., Affeld, K. andBücherl, E. S. (1979) Die modifizierte Thromboelastographie, ein geeigneter invitro-Blutverträglichkeitstest für Biomaterialien.Biomed. Technik.,24, 297.
Lyman, D. J., Muir, W. M. andLee, I. J. (1965) The effect of chemical structure and surface properties of synthetic polymers on the coagulation of blood.Trans. Amer. Soc. Artif. Organs,11, 301.
Lyman, D. J., Brash, J. L., Chaikin, S. W., Klein, K. G. andCarini, M. (1968) Protein and platelet interaction with polymer surfaces.Trans. Amer. Soc. Artif. Organs,14, 250.
Lyman, D. J., Metcalf, L. C., Albo, D., Richards, K. F. andLamb, J. (1974) In vivo adsorption of proteins on polymer surfaces.Trans. Amer. Soc. Artif. Organs,20, 474.
Lyman, D. J., Knutson, K., McNeill, B. andShibatani, K. (1975) The relation between polymer morphology and protein adsorption.Trans. Amer. Soc. Artif. Organs,21, 49.
Nosé, Y. andKambic, H. (1973) Development of blood and tissue compatible materials. First Annual Report, Cleveland Clinic Foundation.
Rennekamp, F., Affeld, K. andBücherl, E. S. (1978) A new method for fabricating blood pumps with a seampless inner surface.Proc. Internat. Soc. Artif. Organs,2, 207.
Rennekamp, F., Lemm, W. andBücherl, E. S. (1979) Long term results with seamless blood pumps out of polyurethanes for the replacement of the heart.Proc. Europ. Soc. Artif. Orangs,6, 94.
Unger, V., Lemm, W. andBücherl, E. S. (1976) Kinetics of albumin and fibrinogen adsorption on biomaterials.Proc. Europ. Soc. Artif. Organs,3, 48.
Warnecke, I., Wolff, D. andBücherl, E. S. (1979) Angiographische Methode zur vergleichenden Untersuchung der Thrombogenität von intravenösen Kathetern.Biomed. Technik.,24, 293.
Watson, J. T., Pitlick, F. A. andAltieri, F. D. (1979) Final Report to the devices and technology branch of the NHLB-Institute, 58–68.
Witt, I. (1975) Biochemie der Blutgerinnung und Fibrinolyse. Verlag Chemie, Weinheim, 90–92.
Author information
Authors and Affiliations
Rights and permissions
About this article
Cite this article
Lemm, W., Unger, V. & Bücherl, E.S. Blood compatibility of polymers:in vitro andin vivo tests. Med. Biol. Eng. Comput. 18, 521–526 (1980). https://doi.org/10.1007/BF02443330
Received:
Issue Date:
DOI: https://doi.org/10.1007/BF02443330