Abstract
Methods of computing orthogonal X, Y and Z voltages representing the equivalent heart dipole activity are described. These components were obtained as linear functions of voltages from the limb and chest electrodes of the conventional 12-lead system. The coefficients of these linear functions are given and a close relationship with the Frank lead system is demonstrated. Two methods were used to obtain transformation coefficients—lead vector approach and statistical least-squares fit. ECG data from 92 individuals (healthy and with cardiac disorders) were used to obtain adequate coefficient values.
Similar content being viewed by others
References
Burger, H. C., van Milaan, J. B. andKlip, W. (1959) Comparison of three different systems of vectorcardiography.Am. Heart J.,57, 723–729.
Burger, H. C., van Brummelen, A. G. W. andvan Herpen, G. (1962) Compromise in vectorcardiography II. Alteration of coefficients as a means of adapting one lead system to another.,64, 666–678.
Dower, G. E. (1971)Polarcardiography, Charles C. Tomas, Springfield, Ill., 225–241.
Dower, G. E., Machado, H. B. andOsborne, J. A. (1980) On deriving the electrocardiogram from the vectorcardiographic leads.Clin. Cardiol.,3, 87–95.
Dower, G. E. (1984) The ECGD: A Derivation of the ECG from VCG leads.J. Electrocardiol.,17, 189–192.
Frank, E. (1956) An accurate clinically practical system for spatial vectorcardiography.Circulation,13, 737–749.
Frank, E. (1954) Image surface of homogenous torso.Am. Heart J.,47, 757–770.
Horan, L. G., Flowers, N. C. andBrody, D. A. (1965) The interchangeability of vectorcardiographic systems.,70, 365–376.
Kornreich, F. (1973) The missing waveform information in the orthogonal electrocardiogram (Frank leads).Circulation,48, 984–995.
Kornreich, F. andRautaharju, P. M. (1981) The missing waveform information in the standard 12 lead electrocardiogram.J. Electrocardiol.,14, 341–350.
Levkov, C., Michov, G., Ivanov, R. andDaskalov, I. K. (1984) Subtraction of 50 Hz interference from the electrocardiogram.Med. & Biol. Eng. & Comput.,22, 371–373.
Levkov, C. L. (1982) Amplification of biosignals by body potential driving.,20, 248–250.
Macfarlane, P. W., Macfarlane, D. K., Podolski, M. andLawrie, T. D. V. (1984) The ECG analysis program for the MINGOCARE system.Electromedica,52, 126–136.
Pipberger, H. V., Bialek, S. M., Perloff, J. K. andSchnaper, H. W. (1961) Correlation of clinical information in the standard 12-lead ECG and in a corrected orthogonal 3-lead ECG.Am. Heart J.,61, 34–43.
Rautaharju, P. (1976) A simple procedure for positioning precordial ECG and VCG electrodes using electrode locator.J. Electrocardiol.,9, 35–40.
Wolf, H. K., Rautaharju, P. M., Unite, V. C., Steward, J. (1976) Evaluation of synthesized standard 12 leads and Frank vector leads.Adv. Cardiol.,16, 87–97.
Author information
Authors and Affiliations
Rights and permissions
About this article
Cite this article
Levkov, C.L. Orthogonal electrocardiogram derived from the limb and chest electrodes of the conventional 12-lead system. Med. Biol. Eng. Comput. 25, 155–164 (1987). https://doi.org/10.1007/BF02442844
Received:
Accepted:
Issue Date:
DOI: https://doi.org/10.1007/BF02442844