Abstract
The system described is based on the employment of a radiosonde energised from the outside in such a way that it will not be subject to the limitations arising from the use of batteries. The main sources of error, consisting in the thermal drift of the electronic components and particularly of the pressure transducer, have been eliminated by transmitting to the outside the information regarding temperature. Thus the data concerning the pressure can be corrected after reception by a simple manual calculation or by automatic computing.
Similar content being viewed by others
References
Atkinson, J. R., Shurtleff, D. B. andFoltz, E. L. (1967) Radiotelemetry for the measurement of intracranial pressure.J. Neurosurg. 27, 428–432.
Beks, J. W. F., Bosch, D. A. andBrock, M. (Eds.) (1976)Intracranial pressure III, Springer-Verlag, Berlin.
Brazzoduro, G. andMacellari, V. (1974) Realizzazione di un prototipo di radiosonda per la misura della pressione del fluido cerebrospinale.Ann. Ist. Super. Sanità 10, 268–274.
Brock, M. andDiefenthaler, K. (1972) A modified equipment for the continuous telemetric monitoring of epidural or subdural pressure. InIntracranial pressure, Springer-Verlag, Berlin, 21–26.
Brock, M. andDietz, H. (Eds.) (1972)Intracranial pressure, Springer-Verlag, Berlin.
Cohadon, F., La Balme, M., Castel, J. P. andVandendriessche (1974) ICP microprobes series microfet.Intracranial pressure II, Springer-Verlag, Berlin, 375–376.
Coroneos, N. J., McDowall, D. G., Gibson, R. M., Pickerodt, V. W. A. andKeaney, N. P. (1973) Measurement of extradural pressure and its relationship to other intracranial pressuresJ. Neurol. Neurosurg. Psychiat. 36, 514–522.
De Rougement, J., Benabid, A. L. andBarge, M. (1974) Intracranial pressure measurement by epidural technique—A simple solution. InIntracranial pressure II, Springer-Verlag, Berlin, 384–385.
Dorsch, N. W. C. andSymon, L. (1974) The validity of extradural measurement of the intracranial pressure. InIntracranial pressure II, Springer-Verlag, Berlin, 403–408.
Foroglou, G., Zander, E., Faure, R. andBesse, R. (1974) Telemetric measurement of intracranial pressure with an electromagnetic detector. InIntracranial pressure II, Springer-Verlag, Berlin, 377.
Handa, H., Yoneda, S., Matsuda, M. andHanda, J. (1974) A miniature transducer for continuous monitoring of intracranial pressure. InIntracranial pressure II, Springer-Verlag, Berlin, 378–380
Harper, C. A. (Ed.) (1970)Handbook of material and processes for electronics, MacGraw-Hill, New York, 1–35.
Jacobson, S. A. andRothballer, A. B. (1966) Prolonged measurement of experimental intracranial pressure using a subminiature absolute pressure transducer.J. Neurosurg. 25, 603–608.
Lundberg, N., Ponten, U. andBrock, M. (Eds.) (1975)Intracranial pressure II, Springer-Verlag, Berlin.
Mackay, S. R. (1970)Bio-medical telemetry, Wiley, New York, 101.
Majors, R., Schettini, A., Mahig, J. andNevis, A. H. (1972) Intracranial pressure measured with the coplanar pressure transducer.Med. & Biol. Eng. 10, 724–733.
Numoto, M., Wallman, J. K. andDonaghy, R. M. P. (1973) Pressure indicating bag for monitoring intracranial pressure.J. Neurosurg. 39, 784–787.
Schettini, A., McKay, L., Majors, R., Mahig, J. andNevis, A. H. (1971) Experimental approach for monitoring surface brain pressure.34, 38–47.
Sundbärg, G. andNoines, H. (1972) Simultaneous recording of the epidural and ventricular fluid pressure. InIntracranial pressure, Springer-Verlag, Berlin, 46–50.
Turner, J. M., Gibson, R. M., McDowall, D. G. andNahhas, F. (1974) Further experiences with extradural pressure monitoring. InIntracranial pressure II, Springer-Verlag, Berlin, 397–402.
Author information
Authors and Affiliations
Rights and permissions
About this article
Cite this article
Barbaro, V., Macellari, V. Intracranial pressure monitoring by means of a passive radiosonde. Med. Biol. Eng. Comput. 17, 81–86 (1979). https://doi.org/10.1007/BF02440957
Received:
Accepted:
Issue Date:
DOI: https://doi.org/10.1007/BF02440957