Abstract
For the correction of losses due to true coincidences summing and edge effect, a simple method which is based on the ratio of a reference single γ-ray energy to that of cascade energies at near and far geometry is developed. The correction factors for several radioactive sources with simple and complex decay schemes are experimentally determined for three types of germanium detectors. It is shown that coincidence summing can be a complex effect and depends on the individual detector, the counting geometry and on the decay scheme of the radionuclide concerned.
Similar content being viewed by others
References
K. DEBERTIN, U. SCHOTZIG, Nucl. Instr. Methods, 158 (1979) 471.
G. J. McCALLUM, G. E. COOTE, Nucl. Instr. Methods, 130 (1975) 189.
K. DEBERTIN, R. G. HELMER, Gamma- and X-ray Spectrometry with Semiconductor Detectors, North Holland, Amsterdam, 1988.
A. ADAMS, R. DAMS, Applied Gamma Ray Spectrometry, Pergamon Press, 1970.
R. J. GEHRKE, R. G. HELMER, R. C. GREENWOOD, Nucl. Instr. Methods, 147 (1977) 405.
J. E. CLINE, IEEE Trans. Nucl. Sci. NS-15, No. 3 (1968) 198.
Author information
Authors and Affiliations
Rights and permissions
About this article
Cite this article
Kafala, S.I. Simple method for true coincidence summing correction. Journal of Radioanalytical and Nuclear Chemistry Articles 191, 105–114 (1995). https://doi.org/10.1007/BF02035990
Received:
Issue Date:
DOI: https://doi.org/10.1007/BF02035990