Die unblutige Bestimmung des Herzzeitvolumens mit der Impedanzkardiographie Vergleichende Untersuchungen mit der Thermodilutionsmethode

The noninvasive determination of cardiac output by means of impedance cardiography Comparative evaluation with a thermal dilution technique

  • Originalarbeiten
  • Published:
Basic Research in Cardiology Aims and scope Submit manuscript

Summary

When a high frequency alternating current is passed through the thorax by means of tape-electrodes, the electrical impedance Z is changing synchronously with the cardiac cycle. The relation between the first derivative dZ/dt and the stroke volume is the basis for the application of impedance cardiography (IK).

Determinations of cardiac output (CO) at rest during ergometer exercise using this noninvasive technique were compared with CO values obtained by the thermal dilution method. In 57 subjects simultaneous determinations using both methods were performed at resting conditions. The correlation coefficient was r=0.63. In addition, in 48 of the subjects CO was measured with both methods during supine ergometer exercise. Comparison of the exercise values resulted in a correlation coefficient of r=0.63. In all 48 cases there was a concordant change of the IK and thermodilution CO values.

In is concluded that relative changes of CO are reliably detected by IK. The absolute values can be utilized only approximatively.

Zusammenfassung

Bei der Impedanzkardiographie wird ein hochfrequenter Wechselstrom über bandförmige Elektroden durch die thoraxorgane geleitet. Der Wechselstromwiderstand (Impedanz Z) und die erste Ableitung dZ/dt ändern sich dabei synchron mit der Herzaktion. Die Beziehung zwischen dem Kurvenverlauf von dZ/dt und dem Schlagvolumen ist Grundlage dieses Meßverfahrens.

Die unblutige Bestimmung des HZV mit Hilfe der Impedanzkardiographie in Ruhe und während Ergometerbelastung wurde einem Standardverfahren zur HZV-Bestimmung, der Thermodilutionsmethode, gegenübergestellt. Bei 57 Patienten wurden Simultanbestimmungen des Ruhewertsmmit beiden Methoden durchgeführt. Es ergab sich ein Korrelationskoeffizient von r=0.68. Zusätzlich wurden ei 48 Patienten mit der Impedanzkardiographie und der Thermodilutionsmethode bestimmte Herzzeitvolumina unter Belastungsbedingungen im Steady-state miteinander verglichen (r=0,63). Jede Änderung des HZV, gemessen mit der Thermodilutionsmethode, ging einher mit einer gleichsinnigen Änderung des mit der Impedanzkardiographie gemessenen Werts. Bei allen 48 Patienten konntenrelative Änderungen des HZV zuverlässig mit Hilfe der Impedanzkardiographie erfaßt werden. Absolutwerte können mit der Methode nur approximativ bestimmt werden.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Similar content being viewed by others

Literatur

  1. Atzler, E., G. Lehmann, Arbeitsphysiologie6, 636 (1932).

    Google Scholar 

  2. Baker, L. E., W. V. Judy, L. E. Geddes, F. M. Langley, D. W. Hill, Cardiovasc. Res. Cent. Bull.9, 135 (1971).

    Google Scholar 

  3. Gauer, O. H., Kreislauf des Blutes, in: Herz und Kreislauf (München-Berlin-Wien 1972).

  4. Harley, A., J. C. Greenfield jr., Aerosp. Med.39, 248 (1968).

    PubMed  Google Scholar 

  5. Hiltmann, W. D., W. Kollmeier, W. Stegaru, H. J. Schaumann, Verh. dtsch. Ges. Inn. Med.80, 1181 (1974).

    PubMed  Google Scholar 

  6. Judy, W. V., F. M. Langley, K. D. McCowen, D. M. Stinnett, L. E. Baker, P. C. Johnson, Aerosp. Med.40, 532 (1969).

    PubMed  Google Scholar 

  7. Judy, W. V., F. M. Langley, K. D. McCoven, L. E. Baker, NASA Tech. Rep. Contract No. NAS 9-4500, 301 (1969).

  8. Kubicek, W. G., J. M. Carnegis, R. P. Patterson, D. A. Witsoe, R. H. Mattson, Aerosp. Med.37, 1208 (1966).

    PubMed  Google Scholar 

  9. Kubicek, W. G., A. H. L. From, R. P. Patterson, D. A. Witsoe, A. Castaneda, R. C. Lillehei, R. Ersek, J. Assn. Advan. Med. Instrum.4, 79 (1970).

    Google Scholar 

  10. Kubicek, W. G., R. P. Patterson, D. A. Witsoe, Annals N. Y. Acad. Sciences170, 724 (1970).

    Google Scholar 

  11. Labadidi, Z., D. A. Ehmke, R. E. Durnin, P. E. Leaverton, R. M. Lauer, Circulation41, 651 (1970).

    PubMed  Google Scholar 

  12. Labadidi, Z., D. A. Ehmke, R. E. Durnin, P. E. Leaverton, R. M. Lauer, Pediatrics47, 870 (1971).

    PubMed  Google Scholar 

  13. Luepker, R. V., J. R. Michael, J. R. Warbasse, Amer. Heart J.85, 83 (1973).

    PubMed  Google Scholar 

  14. Moritz, E., W. Kreuzer, F. Kubicek, K. Polzer, F. Schuhfried, Intensivmed.12, 317 (1975).

    Google Scholar 

  15. Nechwatal, W., A. Eversmann, P. Bier, E. König, Klin. Wschr., im Druck.

  16. Nyboer, J., S. Bagno, A. Barnett, R. H. Halsey, Amer. Soc. Clin. Inv.19, 773 (1940).

    Google Scholar 

  17. Pomerantz, M., F. Delgado, B. Eisemann, Surgery70, 865 (1971).

    PubMed  Google Scholar 

  18. Smith, J. J., V. T. Wiedmeier, F. E. Tristani, K. E. Cooper, Fed. Proc.28, 643 (1969).

    Google Scholar 

  19. Thomasson, B., Scand. J. Clin. Lab. Invest.9, 365 (1957).

    PubMed  Google Scholar 

  20. Topham, W. S., H. R. Warner, NASA Tech. Rep. Contract No. NAS 9-4500, 175 (1969).

  21. Van de Water, J. M., P. A. Philips, L. G. Thouin, L. S. Watanabe, R. S. Lappen, Arch. Surg.102, 541 (1971).

    PubMed  Google Scholar 

Download references

Author information

Authors and Affiliations

Authors

Additional information

Mit 6 Abbildungen

Rights and permissions

Reprints and permissions

About this article

Cite this article

Nechwatal, W., Bier, P., Eversmann, A. et al. Die unblutige Bestimmung des Herzzeitvolumens mit der Impedanzkardiographie Vergleichende Untersuchungen mit der Thermodilutionsmethode. Basic Res Cardiol 71, 542–552 (1976). https://doi.org/10.1007/BF01909768

Download citation

  • Received:

  • Issue Date:

  • DOI: https://doi.org/10.1007/BF01909768