Abstract
Objective: To compare the heat and moisture exchange efficiency of 4 commonly used ventilator circuit filers—DAR Hygroster (DHS), DAR Hygobac (DHC), Pall Ultipor BB 50 (PUBC) and Intersurgical filtatherm (IFT).Design: Prospective randomized study.Setting: Intensive care unit of a university teaching hospital.Patients and Participants: 80 patients requiring post-operative mechanical ventilation were entered into the study. A total of 40 patients were studied after 1 h and 40 after 24 h. Due to technical errors 6 of the 1 h studies and 2 of the 24 h studies were excluded from analysis.Interventions: Mean wet and dry ceramic-platinum thermal probes were inserted in the inspiratory side of a divice which separated inspiratory and expiratory gas flows on the patient side of the ventilator circuit filter. Humidification efficiency was calculated with reference to these wet and dry bulb temperatures. Tracheal temperature was measured by a similar thermal probe in the tracheal tube.Measurements and results: Absolute humidity was greatest in group DHS (30.4±3.5 versus 22.9±1.5 mg/l at 1 h and 27.3±3.5 versus 20.7±2.4 mg/l at 24 h). The differences in absolute humidity reached statistical significance as follows:p=0.0001 for DHS versus all others at 1 h and for DHS and DHC versus PUBC and IFT at 24 h.Conclusion: Better humidification capabilities were obtained with groups DHS and DHC than PUBC and IFT.
Article PDF
Similar content being viewed by others
References
Martin C, Perrin G, Gevaudan M, Saux P, Gouin F (1990) Heat and moisture exchangers and vaporizing humidifiers in the ICU. Chest 97:144–149
Branson RD (1987) Laboratory evaluation of moisture output of senen airway heat and moisture exchangers. Respir Care 37:741–745
Shelly MP, Lloyd GM, Park GR (1988) A review of the mechanisms and methods of humidification of inspired gases. Intensive Care Med 14:1–9
Chalon J, Patel C, Ramanathan S, Capan L, Tang C-K, Turndorf H (1979) Humidity and the anaestetised patient. Anaesthesilogy 50:195–198
Forbes AR (1973) Humidification and mucous flow in the intubated trachea. Br J Anaesth 45:874
Forbes AR (1974) Temperature, humidity and mucous flow in the intubated trachea. Br J Anaesth 46:29–34
Noguchi H, Takumi Y, Rochi O (1973) A study of the humidification in tracheostomized dogs. Br J Amaesth 45:844–847
Stone DR, Downs JB, Paul WL, Perkins HA (1981) Adult body temperature and heated humidification during general anaesthesia. Anaesth Analg 60:736–741
Graff TD, Benson DW (1969) Systemic and pulmonary changes with inhaled humid atmospheres: clinical application. Anaesthesiology 30:199
Marfatia S, Donahoe JB, Hendren WH (1975) Effect of dry and humidified gases on the respiratory epithelium in rabbits. J Paediatr Surg 10:583–587
Dalhamn T (1956) Mucous flow and ciliary activity in the trachea of healthy rats and rats exposed to respiratory irritant gases. Acta Physiol Scand [Suppl] 131:60
Craven DE, Goularte TA, Make BJ (1984) Contaminated condensate in mechanical ventilator circuits as a risk factor for nosocomial pneumonia. Am Rev Respir Dis 129:625
Christopher KL, Saravolatz LD, Bush TL, Conway WL (1983) The potential role of respiratory theray equipment in cross infection. Am Rev Respir Dis 128:271
Harris RL, Riley HD (1967) Reactions to aerosol medication in infants and children. JAMA 210:953
Im SWK, Fung JPH (1982) Unusual dissemination of pseudomonads by ventilators. Anaesthesia 37:1074
Emergency Care Research Institute (1983) Heat and moisture exchangers. In: Health Device 12:155
Weeks DB, Ramsey FM (1983) Laboratory investigations of six artificial noses for use during endotracheal anaesthesia. Anaesth Analg 62:758–763
Walker AKY, Bethune DW (1976) A comparative study of condenser humidifiers. Anaesthesia 31:1086–1093
Mebius C (1983) A comparative evaluation of disposable humidifiers. Acta Anaesthesiol Scand 27:403–409
Cohen IL, Weinberg PF, Fein IA, Rowinski GS (1988) ET tube occlusion associated with the use of heat and moisture exchangers in the intensive care unit. Crit Care Med 16:277–279
Gallagher J, Strangeways JEM, Allt-Graham J (1987) Contamination control in long term ventilaton: a clinical study using a heat and moisture-exchanging filters. Anaesthesia 42:476–481
Condon EU (1967) Handbook of physics. Churchill Livingstone, New York
Weeks DB (1975) Humidification during anaesthesia. NY State J Med 75:1216–1218
Weeks DB (1986) A laboratory evaluation of recently available heatand-moisture exchangers. Anaesthesiol Rev 13:33–36
Dery R (1973) The evolution of heat and moisture in the respiratory tract during anaesthesia with a non rebreathing system. Can Anaesth Soc J 20:296–309
Chalon J, Markham JP, Ali MM, Ramanathan S, Turndorf H (1984) The Pall Ultipor breathing circuit filter: an efficient heat and moisture exchanger. Anaesth Analg 63:566–570
Boys JE, Howells TH (1972) humidification in anaesthesia. A review of the present situation. Br J Anaesth 44:879
Burton JDK (1962) Effects of dry anaesthetic gases on the respiratory mucous membrane. Lancet I:234
Lowe GD (1987) Efficiency of heat and moisture exchangers: a reply. Anaesthesia 42:1118
Weeks DB (1981) Evaluation of disposable humidifier for use during anaesthesia. Anaesthesiology 54:337–340
Author information
Authors and Affiliations
Rights and permissions
About this article
Cite this article
Jackson, C., Webb, A.R. An evaluation of the heat and moisture exchange performance of four ventilator circuit filters. Intensive Care Med 18, 264–268 (1992). https://doi.org/10.1007/BF01706470
Received:
Accepted:
Issue Date:
DOI: https://doi.org/10.1007/BF01706470