HA-coated hip implants: A 10 year follow-up

(May '87 – Dec '98)

  • Sotest
  • The use of hydroxyapatite in orthopaedic surgery: a 10-year experience 1986–1999
  • Published:
European Journal of Orthopaedic Surgery & Traumatology Aims and scope Submit manuscript

Summary

We started using HA-coated hip prostheses on May 1997 and our personal experience reaches to date 2075 HA hips of which 1898 were primary hips. This study was based upon a consecutive series of 134 consecutive HA cases implanted from May '87 until December '88, at the Clinique Médico-Chirurgicale of Bruay-Labuissière, France. No patient was lost to follow up. 26 patients deceased and of the 104 remaining cases, 58 complete dossiers were available for addressing clinical and radiological scored results. Within the 58 over 10 year cases cohort, mean age was 56 years. Main aetiology was osteoarthritis in 81% and necrosis in 17%, 1 case was a revision one. Femoral component was HA Omnifit stem in all the cases, fitted with Arc 2f cup. No case of loosening nor any mechanical failure was observed within this over 10 yr of FU series, since 4 reoperations were related to 2 stiffnesses and 2 recurrent dislocations with neither retrieval of acetabular nor femoral component.

As regards radiographic findings, femoral lysis over 10 yr was none in 48%, limited to neck area in 28%, proximally extended in 24%, as no case of distal lysis was observed. 88% of cups had no lysis at 10 yr of FU. All our 58 “10 yr+” cases belonged to the “ingrowth confirmed” group. Mean values of HHS scores were preoperatively 33.34 ± 12.46 out of 100 pts preoperatively up to 97.6 pts at 10 year. Thus results were “excellent” in 86%, “good” in 9%, “fair” in 5%, and “poor” in 0%. Mean values according to Merle d'Aubigne score (MDA) were 7.98 ± 1.74 pts out of 18 preoperatively up to 16.90 pts at 10 yr, leading to 84% of “excellent”, 11% of “good”, 5% of “fair” and 0% of “poor” results.

When considering the cumulative survival analysis since 1987, the results at 10 year, based upon the 1922 primary cases reaches 0.9889 ± 0.0061 when considering all types of failures (Mechanical loosening + Clinical failure + Radiological failure). While considering a 10 year follow-up period as a major milestone in terms of hip results, the results of this series may look very encouraging and lead us to remain confident in bioactive coatings.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Similar content being viewed by others

Author information

Authors and Affiliations

Authors

Rights and permissions

Reprints and permissions

About this article

Cite this article

Epinette, J.A. HA-coated hip implants: A 10 year follow-up. Eur J Orthop Surg Traumatol 9, 83–85 (1999). https://doi.org/10.1007/BF01695733

Download citation

  • Issue Date:

  • DOI: https://doi.org/10.1007/BF01695733

Key words

Navigation