Strömung in natürlichen und künstlichen Organen und Gefäßen

Flow in natural and artificial organs and vessels

  • Published:
Klinische Wochenschrift Aims and scope Submit manuscript

Summary

After introductory remarks about the flow phenomena in the human body (separation, dead waters, stagnation regions) and the model simularity conditions, the flow behaviour in arterial bifurcations at pulsatile flow are considered; they are accompanied by the formation of dead waters and secondary flows and therefore they give the danger of atherosclerosis. Then the flow pattern at arterio-venous anastomosis is interpreted with regard to the danger of thrombogenesis.

At the selected example of ball prosthesis the flow pattern and the pressure loss of artificial heart valves (aortic as well as mitral valves) are described; the special behaviour of ball valves used as mitral valves is discussed critically.

Finally in a view of contribution to the question of the flow-induced blood damage, in extended series of basic tests on models (orifices and perforated discs) the rate of hemolysis was measured depending on the flow; the tests were completed by measurements at rolling pumps, atria of a artificial ventricle and other parts.

In order to try an explanation for the flow-induced hemolysis one starts from the influence of the hemodynamic shear stresses on the damage of the erythrocytes; hereby the effect of the frequently repeated passages of the red blood cells through a stenosis (e.g. regurgitation) should be taken into account.

Zusammenfassung

Nach einer kurzen Einführung in Strömungserscheinungen (Ablösung, Totwasser, Stagnationsbereiche) und in die Modellähnlichkeit werden die Vorgänge in Aderverzweigungen bei pulsierender Strömung mit der Bildung der durch Sekundärströmung bedingten Towasserbereiche als eine Voraussetzung der Atherosklerose betrachtet; anschließend werden Strömungsbilder bei arteriovenösen Anastomosen hinsichtlich ihrer Thrombogenesegefahr gedeutet. Strömungsverlauf und Druckverlust künstlicher Herzklappen werden am Beispiel von Ballprothesen besprochen; hierbei wird das beondere Verhalten der Ballprothesen als Mitralisklappe kritisch erörtert.

In längeren Reihen von Grundsatzversuchen mit Modellen (Lochscheiben, perforierte Scheiben) wurde durch Messung der Hämolyserate in Abhängigkeit vom Strömungszustand die Möglichkeit starker strömungsbedingter Blutschädigung nachgewiesen; die Versuche wurden durch Messungen an Rollenpumpen, Vorhöfen künstlicher Ventrikel u.a. ergänzt. Zur strömungstechnischen Deutung wird von der Wirkung der hämodynamischen Schubspannung auf die Hämolyse ausgegangen; als wesentlich wird auf den Effekt vielfach wiederholten Durchgangs der Erythrozyten durch hämolysebegünstigende Stellen hingewiesen (z.B. Regurgitation).

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Similar content being viewed by others

Literatur

  1. Müller-Mohnssen, H.: Die Strömungsverhältnisse in den Coronararterien usw. Probleme der Coronardurchblutung; Verlag Springer, Göttingen 1958. S. 179

    Google Scholar 

  2. Rudinger, G.: Shock waves in mathematical models of the aorta. J. appl. Mechan.37, Trans. ASME (1970), S. 34

    Google Scholar 

  3. Lambossy, P.: Aperçu historique et critique sur le problème de la propagation des ondes dans un liquide compressible enfermé dans un tube elastique. Helv. Physiol. Acta8, (1950) 209 und9, 145 (1951)

    Google Scholar 

  4. Rosenberg, H.: Instationäre Strömungsvorgänge in Leitungssystemen mit flexibel-elastischen Rohrwänden. Diss. TH Aachen 1963; Forsch.-Ber. Nordrhein-Westfalen Nr. 1450

  5. Lighthill, M.J.: Vicosity effects in sound waves of finite amplitudes. in G.K. Batchelor, H. Bondi: Survey in Mechanics. Cambridge Univ. Press 250 (1956)

  6. Zeller, H., Talukder, N., Lorenz, J.: Model studies of pulsating flow in arterial branches and wave propagation in blood vessels. Agard Conf. Proc.65, 15 (1970). — Versuche zur pulsierenden Strömung durch arterielle Verzweigungen und zur Wellenausbreitung in großen Blutgefäßen. Abhdlgn. Aerodyn. Inst. Aachen20, 41 (1971)

    Google Scholar 

  7. Ravindran, R.: Wellenausbreitung durch nicht-Newtonsche Flüssigkeiten in elastischen Leitungen. Dissert. TH Aachen 1973

  8. Lorenz, J.: Über die Wellenausbreitung in viscoelastischen Leitungen und ihre gasdynamische Analogie. Diss. TH. Aachen 1973

  9. Teipel, J.: Nichtlineare Wellenausbreitungsvorgänge in elastischen Leitungen. Acta Mech.16, 93 (1973)

    Google Scholar 

  10. Zeller, Lorenz, J.: Analogous treatment of wave propagation in liquid-filled elastic tubes and gas-filled rigid tubes. Euromech. Coll. London 1973

  11. Dropmann, K.: Über die Prädilektionsstellen der Atherosklerose und die hämodyn. Verhältnisse in Krümmerstrecken und an Teilungsstellen. Zs. f. Kreislaufforschg.52, 171 (1963)

    Google Scholar 

  12. Talukder, N., Zeller, H.: Modellversuche zur Strömung in arteriellen Verzweigungen. Dt. Ges. Biomed. Technik Hannover 1974. S. 163

  13. Talukder, N.: Untersuchung zur Strömung in arteriellen Verzweigungen. Diss. TH. Aachen 1974

  14. Fry, D.L.: Acute vascular endothelic changes associated with increased blood velocity gradients. Circulation research22, 165 (1968)

    Google Scholar 

  15. Caro, C.G., Fitzgerald, J.M., Schröter, R.C.: Atheroma and arterial wall shear. Proc. Roy. Soc. London B177, 109 (1971)

    Google Scholar 

  16. Gerlach, R., Lymberopoulos, S.: Modellstudien zur Strömungsführung in verschiedenen Gefäßanastomosen. Dt. Ges. Biomed. Techn. 1974. S. 159. Strömungstechn.-vergleichende Untersuchungen innerer arteriovenöser Anatomosen. Dt. Ges. Biomed. Techn. 1973. S. 101

  17. Reul, H.: Funktions- und Strömungsuntersuchungen an natürlichen und künstlichen Herzklappen usw. Diss. TH. Aachen 1971

  18. Chmiel, H.: Zur Blutrheologie in Medizin und Technik. Habilitationsschrift TH. Aachen 1973

  19. Kalmar, P.: Künstliche Herzklappen. Dt. Ges. Biomed. Techn. Erlangen 1973. S. 207

  20. Kramer, C., Naumann, A., Bleifeld, W.: Effert, S.: Über die Durchströmung künstlicher Herzklappen. Thoraxchirurgie, Vasculäre Chirurgie16, 269 (1968)

    Google Scholar 

  21. Naumann, A.: Strömungsfragen der Medizin. Arb.-Gem. f. Forschg. Nordrh.-Westf. H. 203 (1969), S. 27

    Google Scholar 

  22. Naumann, A.: Über Strömungsvorgänge im menschlichen Körper. Acta medicotechnica18, 46 u. 59 (1970)

    Google Scholar 

  23. Naumann, A., Kramer, C.: Flow investigations on artificial heart valves. AGARD Conf. Proc.65, 4 (1971)

    Google Scholar 

  24. Belhouse, B.J., Talbot, L.: The fluid mechanics of the aortic valve. J. Fluid Mech.35, 721 (1969)

    Google Scholar 

  25. Kramer, C.: Künstliche Herzklappen und mechanische Hämolyse. Habilitationsschr. TH. Aachen 1973

  26. Köhler, J., Kramer, C.: Zum Schließvorgang künstlicher Herzklappen. Acta Medicotechn.20, 36 (1972)

    Google Scholar 

  27. Störmer, B., Mendling, W., Kremer, K., Köhler, J.: Vergleichende Modelluntersuchungen zu künstlichen Herzklappen im Kreislaufsimulator. Dt. Ges. f. Thorax-, Herz- u. Gefäßchirurgie Bad Nauheim 1974

  28. Köhler, J., Lambert, J., Heuser, G.: Strömungsuntersuchungen zum Füllvorgang einer pulsierend fördernden Blutpumpe. Dt. Ges. Biomed. Techn. 1973. S. 121

  29. Köhler, J., Blumenfeld, M., Eckstein, H., Heuser, G.: Modelluntersuchungen zur Strömung in Pumpen und Ventilen zur Blutförderung. Dt. Ges. Biomed. Techn. 1974. S. 165

  30. Köhler, J., Heuser, G.: Verbesserung der Kammerfüllung einer pulsierend fördernden Blutpumpe mit künstlichem Vorhof. Abhdlgn. Aerodyn. Institut TH. Aachen, H. 21 (1974), S. 42

    Google Scholar 

  31. Gehrmann, G., Bleifeld, W., Loogen, F.: Mechanische Hämolyse nach Implantation künstlicher Herzklappen. Zs. f. Kreislaufforschg.55, 25 (1966)

    Google Scholar 

  32. Gehrmann, G., Loogen, F.: Mechanische hämolytische Anämie nach Implantation künstlicher Aortenklappen. Dt. Mediz. Wochenschr.89, 625 (1964)

    Google Scholar 

  33. Kramer, C.: Studies on flow induced mechanical hemolysis. AGARD Conf. Proc.65, 5 (1970)

    Google Scholar 

  34. Kramer, C., Sand, P., Bleifeld, W.: Blutströmung und mechanische Hämolyse. Zs. f. Biomed. Techn.16, 164 (1971)

    Google Scholar 

  35. Bleifeld, W., Kramer, C., Pelzer, H.: Strömungsdynamische Einflußgrößen der mechanischen Hämolyse. Thoraxchirurgie. Vaseuläre Chirurgie21, 130 (1973)

    Google Scholar 

  36. Gerhardt, H.J., Kramer, C.: Investigations of damping in turbulent blood flow. Euromech. Coll. London 1973

  37. Holzhüter, H., Angelkort, B., Kenzel, E., Lambert, J., Pelzer, H., Köhler, J.: Neues Meßverfahren zur quantitativen Erfassung der Thrombozyten — Destruktion in extrakorporalen Kreisläufen. Dt. Ges. Biomed. Techn. 1974. S. 177

  38. Nevaril, C.G., Hellums, J.D., Alfrey, C.P., Lynch, E.C.: Physical effects in red blood cell trauma. J. Amer. Inst. Chem. Eng.15, 707 (1969)

    Google Scholar 

  39. Williams, A.R.: Shear-induced fragmentation of human erythrocytes. Biorheology10, 303 (1972)

    Google Scholar 

  40. Taylor, G.J.: The formation of emulsions in definable fields of flow. Proc. Roy. Soc. A146, 501 (1934)

    Google Scholar 

  41. Schmid-Schönbein, H., Wells, R.: Fluid drop-like transition of erythrocytes under shear. Science18, 288 (1969)

    Google Scholar 

  42. Williams, A.R.: Viscoelasticity of the human erythrocyte membrane. Biorheology11, 313 (1973)

    Google Scholar 

  43. Rand, R.: Mechanical properties of the red blood cell membrane. Biophys. Journ.4, 308 (1964)

    Google Scholar 

  44. Blackshear, P.L.: Mechanical hemolysis in the flowing blood; in Y.C. Fung, N. Perrone, M. Anliker: Biomechanics. Prentice Hall 1972. S. 501.

  45. Benner, K.U., Frede, K.E., Lauterjung, K.L.: In vitro investigation with a new method for continuous and quantitative measurements. Pflügers Arch.310, 142 (1970)

    Google Scholar 

  46. Lambert, J., Egberts, M., Holzhüter, H., Wenzel, E., Faude, N.: Ermittelung der zeitlichen Druckänderung und Blutschädigung im Eingriffsbereich einer Rollenpumpe. Dt. Ges. Biomed. Techn. 1974 S. 173

  47. Lambert, J., Naumann, A.: Zur Strömungsmechanik und Blutschädigung in Hochdruckrollenpumpen. Abhdlgn. Aerodyn. Institut Aachen, H.21, 48 (1974)

    Google Scholar 

  48. Heuser, G., Köhler, J., Lambert, J., Bleifeld, W.: Ein Modell für Detailstudien zur Strömungsmechanik und Blutschädigung im Ballonbereich bei Anwendung der intraaortalen Ballonpulsation. Dt. Ges. Biomed. Techn. 1974 S. 173

  49. Dintonfass, L., Yu, J.S., Grace, C.: Dynamic blood coagulation in Theoretical and Clinical Hemorheology. Springer Verlag 1971. S. 303

  50. Schneider, M.D.: Nature of thrombus in a hollow fiber oxygenator. Amer. Journ. Vet. Res.34, 151 (1973)

    Google Scholar 

  51. Turina, M.: Die Rolle des Blutfilters in der Verhütung des Lungenschadens nach Ganzkörperperfusion. Thoraxchirurgie20, 122 (1972)

    Google Scholar 

Download references

Author information

Authors and Affiliations

Authors

Additional information

Vortrag auf der 108. Versammlung der Gesellschaft Deutscher Naturforscher und Ärzte, Berlin, 16.–19. September 1974.

Ord. em. Professor für Strömungslehre. Herrn Dipl.-Ing. J. Lambert sei für die Unterstützung bei der Abfassung dieses Berichtes gedankt.

Rights and permissions

Reprints and permissions

About this article

Cite this article

Naumann, A. Strömung in natürlichen und künstlichen Organen und Gefäßen. Klin Wochenschr 53, 1007–1019 (1975). https://doi.org/10.1007/BF01614302

Download citation

  • Issue Date:

  • DOI: https://doi.org/10.1007/BF01614302

Key words

Schlüsselwörter

Navigation