Abstract
A photometer is described for the measurement of light-scattering in high-quality, optical bulk glass. The photometer operates continuously over the wavelength range 0.5 to 1μm and can measure scattering coefficients down to a level of 5×10−7 cm−1. A photon counting technique is used for light measurement and calibration is by reflection of the incident beam from an ideal diffusing screen. Measurements with high-purity benzene are in good agreement with computed and previous experimental values. Results of scattering from two samples of optical glass are given.
Similar content being viewed by others
References
W. A. Gambling andH. Edels,Brit. J. Appl. Phys. 8 (1957) 481–482.
J. K. Nakamura andS. E. Schwarz,Appl. Optics 7 (1968) 1073–1078.
R. R. Alfano andN. Ockman,J. Opt. Soc. Am. 58 (1968) 90–95.
G. A. Morton,Appl. Optics 7 (1968) 1–10.
F. Grum andG. W. Luckey,ibid 7 (1968) 2289–2294.
W. H. Martin,Trans. Roy. Soc. Canada 17 III (1923) 151–155.
G. Dezelić,J. Chem.Phys. 45 (1966) 185–191.
Author information
Authors and Affiliations
Rights and permissions
About this article
Cite this article
Laybourn, P.J.R., Dakin, J.P. & Gambling, W.A. A photometer to measure light scattering in optical glass. Opto-electronics 2, 36–42 (1970). https://doi.org/10.1007/BF01436153