Spinal cord compression caused by the rod of a Harrington instrumentation device: A late complication in scoliosis surgery

  • Case Report
  • Published:
European Spine Journal Aims and scope Submit manuscript

Abstract

We report a unique complication caused by the rod of a Harrington instrumentation device, which resulted in spinal stenosis and myelopathy. A literature review revealed no previous causes of direct spinal cord impingement caused by the rod of a Harrington device. In this case, years after the initial operation, the rod penetrated the lamina at the junction between a thoracic and a lumbar curve, causing spinal stenosis and myelopathy. We conclude that regular control of the position of the device and awareness of possible late neurological complications are necessary to identify such complications as early as possible.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Similar content being viewed by others

References

  1. Been HD, Kalkman CJ, Traast HS, Ongerboer de Visser BW (1994) Neurologic injury after insertion of laminar hooks during Cotrel-Dubousset instrumentation. Spine 19:1402–1405

    Google Scholar 

  2. Bernard TN, Johnston CE, Roberts JM, Burke SW (1983) Late complications due to wire breakage in segmental spinal instrumentation. J Bone Joint Surg [Am] 65:1339–1345

    Google Scholar 

  3. Court-Brown CM, McMaster MJ (1982) Pseudarthrosis: a late cause of paraparesis after scoliosis surgery. J Bone Joint Surg [Am] 64:1246–1248

    Google Scholar 

  4. Curtis RS, Dickson JH, Harrington PR, Wendell DE (1979) Results of Harrington instrumentation in the treatment for severe scoliosis. Clin Orthop 144: 128–134

    Google Scholar 

  5. Dove J (1988) Segmental wiring for spinal deformity. A morbidity report. Spine 14:229–231

    Google Scholar 

  6. Eismont FJ, Simeone FA (1981) Bone overgrowth (hypertrophy) as a cause of late paraparesis after scoliosis fusion. J Bone Joint Surgery [Am] 63: 1016–1019

    Google Scholar 

  7. Guillemin F, Czomy A, Pouewl J (1991) Sciatic nerve compression by hematoma: case report of a late complication of Harrington's operation. Spine 16:237–239

    Google Scholar 

  8. Hales DD, Dawson EG, Delamarter R (1989) Late neurological complications of Harrington-Rod instrumentation. J Bone Joint Surg [Am] 71:1053–1057

    Google Scholar 

  9. Johnston CE, Happel LT, Norris R, Burke SW, King AG, Roberts JM (1986) Delayed paraplegia complicating sublaminar segmental spinal instrumentation. J Bone Joint Surg [Am] 68: 556–563

    Google Scholar 

  10. Kornberg M, Herndon WA, Rechtine GR (1985) Lumbar nerve root compression at the site of hook insertion: late complication of Harrington rod instrumentation for scoliosis. Spine 10: 853–855

    Google Scholar 

  11. MacEwen GD, Bunnell WP, Sriram K (1975) Acute neurological complications in the treatment of scoliosis. J Bone Joint Surg [Am] 57:404–408

    Google Scholar 

  12. Michel CR, Lalain JJ (1985) Late results of Harrington's operation. Spine 10:414–419

    Google Scholar 

  13. Moskowitz A, Moe JH, Winter RB, Binner H (1980) Long-term follow-up of scoliosis fusion. J Bone Joint Surg [Am] 62:364–375

    Google Scholar 

  14. Savini R, Di Silvestre M, Gargiulo G (1990) Late paraparesis due to pseudarthrosis after posterior spinal fusion. J Spine Disord 3:427–432

    Google Scholar 

  15. Schlenzka D, Poussa M, Muschik M (1993) Operative treatment of adolescent idiopathic thoracic scoliosis. Clin Orthop 297:155–160

    Google Scholar 

  16. Thompson GH, Wilber RG, Shaffer JW, Scoles PV, Nash CL (1985) Segmental spinal instrumentation in idiopathic scoliosis. A preliminary report. Spine 10:623–630

    Google Scholar 

  17. Wilber RG, Thompson GH, Shaffer JW, Brown RH, Nash CL (1984) Postoperative neurological deficits in segmental spinal instrumentation. J Bone Joint Surg [Am] 66:1178–1187

    Google Scholar 

Download references

Author information

Authors and Affiliations

Authors

Rights and permissions

Reprints and permissions

About this article

Cite this article

Krödel, A., Rehmet, J.C. & Hamburger, C. Spinal cord compression caused by the rod of a Harrington instrumentation device: A late complication in scoliosis surgery. Eur Spine J 6, 208–210 (1997). https://doi.org/10.1007/BF01301439

Download citation

  • Received:

  • Revised:

  • Accepted:

  • Issue Date:

  • DOI: https://doi.org/10.1007/BF01301439

Key words

Navigation