Summary
A system consisting of a miniaturized enzyme reactor and a novel ammonium-selective minielectrode was designed for urea determinations in 10-μl samples. A discontinuous technique was devised in view of a high measurement reproducibility. The parameters of the measuring procedure were carefully optimized. The applicability of the urea analyzer was tested using 122 blood serum samples from patients. The results of the enzymatic/potentiometric method agreed favorably with the reference values determined by the diacetyl monoxime method. The relative residual standard deviation found for such linear correlations (slope 1.0) was ±6% over the 1 to 45 mM urea range.
Zusammenfassung
Ein Meßsystem, bestehend aus einem miniaturisierten Enzymreaktor und einer neuartigen Ammonium-Minielektrode, wurde für Harnstoffbestimmungen in 10-μl-Proben entwickelt. Im Hinblick auf eine hohe Reproduzierbarkeit der Meßwerte wurde ein nichtkontinuierliches Verfahren erarbeitet. Die Parameter des Meßverfahrens wurden sorgfältig optimiert. Die Anwendbarkeit des Harnstoff-Analysators wurde an 122 Blutserumproben von Patienten überprüft. Die Ergebnisse der enzymatisch-potentiometrischen Bestimmungsmethode korrelierten ausgezeichnet mit jenen der Diacetylmonoxim-Referenzmethode. Die relative Streuung um die entsprechenden Regressionsgeraden (Steigung 1.0) betrug ±6% für einen Harnstoff-Konzentrationsbereich von 1–45 mM.
Similar content being viewed by others
References
Z. Stefanac, W. E. Morf, U. Thanei, I. Mostert, R. Dörig, H.-B. Jenny, and W. Simon, in: “Progress in Enzyme and Ion-Selective Electrodes” (D. W. Lübbers et al, eds.). Berlin-Heidelberg-New York: Springer-Verlag. 1981. p. 61.
H. H. Weetall and L. S. Hersh, Biochim. Biophys. Acta185, 464 (1969).
G. Johansson and L. Ögren, Analyt. Chim. Acta84, 23 (1976).
G. Johansson and L. Ögren, in: “Ion-Selective Electrodes” (E. Pungor and I. Buzás, eds.). Akadémiai Kiadó, Budapest. 1977, p. 93.
B. Watson and M. H. Keyes, Analyt. Letters9, 713 (1976).
D. N. Gray, M. H. Keyes, and B. Watson, Analyt. Chemistry49, 1067 A (1977).
U. Thanei-Wyss, Diss. ETH Nr. 7114, Zürich, 1982.
H.-B. Jenny, D. Ammann, R. Dörig, B. Magyar, R. Asper, and W. Simon, Mikrochim. Acta [Wien]1980 II, 125.
W. E. Morf and U. Thanei-Wyss, Mikrochim. Acta [Wien], submitted for publication.
P. C. Meier, W. E. Morf, M. Läubli, and W. Simon, in preparation.
W. Simon, D. Ammann, M. Oehme, and W. E. Morf, Ann. New York Acad. Sci.307, 52 (1978).
W. E. Morf, The Principles of Ion-Selective Electrodes and of Membrane Transport, Studies in Analytical Chemistry, Vol. 2, Amsterdam: Elsevier. 1981; Akadémiai Kiadó, Budapest, 1981.
W. E. Morf and W. Simon, in: “Ion-Selective Electrodes in Analytical Chemistry” (H. Freiser, ed.), New York: Plenum Press. 1978. ch. 3.
R. Dörig, Diss. ETH Zürich, in preparation.
Author information
Authors and Affiliations
Rights and permissions
About this article
Cite this article
Thanei-Wyss, U., Morf, W.E., Lienemann, P. et al. Determination of urea in 10-μl blood serum samples with a urease reactor / ion-selective electrode cell. Mikrochim Acta 81, 135–147 (1983). https://doi.org/10.1007/BF01202688
Received:
Issue Date:
DOI: https://doi.org/10.1007/BF01202688