Summary
An automated continuous monitoring method for total mercury at sub-ppb level based on cold vapor atomic-absorption-“continuous micro flow analysis“ is developed. The method uses continuous micro flow sample partial digestion, reduction and extraction in small bore tubes at low flow rates of reagents. Tin(II) chloride in sodium hydroxide solution was used as a reducing reagent. A mixed solution of potassium peroxodisulphate and copper(II) ion was used as oxidizing and catalytic reagent, respectively. The method was successfully applied to the completely continuous monitoring of total mercury in both wastewater and saline water without a heating digestion bath.
Similar content being viewed by others
References
Rujicka J, Hansen EH (1986) Anal Chim Acta 179:1–58
Korenaga T (1987) Bunseki:245–252
Valcarcel M, Luque de Castro MD (1987) Flow injection analysis. Elis Horwood, Chichester
Goto M (1983) Trends Anal Chem 2:92–94
Goto M, Shiroeda T, Ishii D (1981) Bunseki Kagaku 30:403–407
Goto M, Shibakawa T, Arita T, Ishii D (1982) Anal Chim Acta 140:179–185
Goto M, Murofushi S, Ishii D (1988) Bunseki Kagaku 37:47–51
Goto M, Nihimura M, Tominaga T, Ishii D (1988) Bunseki Kagaku 37:52–55
Goto M, Kumagai S, Ishii D (1988) Anal Sci 4:87–90
Hatch WR, Ott WL (1968) Anal Chem 40:2085–2087
Bailey BW, Lo FC (1971) Anal Chem 43:1525–1526
El-Awady AA, Miller RB, Carter MJ (1976) Anal Chem 48:110–116
Coyle P, Hartley T (1981) Anal Chem 53:354–356
Oda CE, Ingle JD Jr (1981) Anal Chem 53:2030–2033
Fujita M, Takabatake E (1983) Anal Chem 55:454–457
Fang Z, Xu S, Wang X, Zhang S (1986) Anal Chim Acta 179:325–340
Sone K, Tanaka M (1958) Qualitative analytical chemistry II. Kyorizu Shupan Co, Tokyo
Umezaki Y, Iwamoto K (1971) Jpn Anal 20:173–179
Magos L (1971) Analyst 96:847–853
Korenaga T, Yamada E, Hara Y, Sakamoto H, Chohji T, Nakagawa C, Ikatsu H, Izawa M, Goto M (1987) Bunseki Kagaku 36:194–198
Toffaletti J, Savory J (1975) Anal Chem 47:2091–2095
Agemian H, Chau ASY (1978) Anal Chem 50:13–16
Author information
Authors and Affiliations
Rights and permissions
About this article
Cite this article
Goto, M., Munaf, E. & Ishii, D. Continuous micro flow monitoring method for total mercury at sub-ppb level in wastewater and other waters using cold vapor atomic absorption spectrometry. Z. Anal. Chem. 332, 745–749 (1988). https://doi.org/10.1007/BF01129768
Received:
Issue Date:
DOI: https://doi.org/10.1007/BF01129768