Summary
A simple equipment for beveling intracellular microelectrodes has been made by modifying a conventional turntable. An agar disc containing ceric oxide polishing compound which was put on a flat glass disc was found to be suitable for beveling of Pyrex single- and double-barreled micropipette electrodes with a tip diameter larger than 0.1 μm. Although the beveled microelectrodes have not always been shown to give better cell penetration into the frog retinal rods than unbeveled fine ones, better records could usually be obtained when electric currents were injected into cells, probably being due to their lower resistance, noise, tip potentials and coupling resistance between two barrels.
Similar content being viewed by others
References
Brown, K. T., Flaming, D. G.: Beveling of fine micropipette electrodes by a rapid precision method. Science185, 693–695 (1974)
Brown, K. T., Flaming, D. G.: Instrumentation and technique for beveling fine micropipette electrodes. Brain Res.86, 172–180 (1975)
Krippe, B. R., Ogden, T. E.: A technique for beveling fine micropipettes. Electroenceph. clin. Neurophysiol.36, 323–326 (1974)
Tasaki, K., Tsukahara, Y., Ito, S., Wayer, M. J., Yu, W. Y.: A simple, direct and rapid method for filling microelectrodes. Physiol. and Behav.3, 1009–1010 (1968)
Author information
Authors and Affiliations
Rights and permissions
About this article
Cite this article
Tauchi, M., Kikuchi, R. A simple method for beveling micropipettes for intracellular recording and current injection. Pflugers Arch. 368, 153–155 (1977). https://doi.org/10.1007/BF01063469