Summary
It is possible to perform continuous quantitativeP O 2 measurements on vasodilated skin by means of surface Pt electrodes according to Clark when the electrode is fixed to the skin with a synthetic plastic material and in situ calibration is performed. A new in situ calibration of theP O 2 electrode is described. At first the skinP O 2 increases with O2 inspiration. After perfusion stop the skinP O 2 shows a linear decrease because of the skin respiration, down to aP O 2 at which hemoglobin liberates chemically bound O2. As thisP O 2 value of hemoglobin is known it is possible to use it for calibrating the electrode. TheP O 2 of normal skin is about 0–7 Torr. After vasodilation obtained by rubbing with a nicotinic acid derivate (Finalgon®, Anasco, Wiesbaden),P O 2 increases to a mean value of 38.1 (±8.1) Torr (n=77). Under these conditions, skinP O 2 reaches arterial values never in adults and rarely in new-born babies.
Similar content being viewed by others
References
Clark, L. C., Jr.: Monitor and control of blood and tissue oxygen tensions. Trans. Amer. Soc. Art. Int. Org.2, 41 (1956).
Ellis, R. A.: In: Handbuch f Haut- und Geschlechtskrankheiten. Epidermal Sub-Layers, Bd. I/1, S. 8. Hrsg.: O. Gans u. G. K. Steigleder. Berlin-Heidelberg-New York 1968.
Evans, N. T. S., Naylor, P. F. D.: The systemic oxygen supply to the surface of human skin. Respirat. Physiol.3, 21 (1967a).
Evans, N. T. S., Naylor, P. F. D.: The oxygen tension gradient across human epidermis. Respirat. Physiol.3, 38 (1967b).
Gleichmann, U., Lübbers, D. W.: Die Messung des Sauerstoffdruckes in Gasen und Flüssigkeiten mit der Platin-Elektrode unter besonderer Berücksichtingung der Messung im Blut. Pflügers Arch. ges. Physiol.271, 431 (1960).
Grunewald, W.: Diffusionsfehler und Eigenverbrauch der Pt-Elektrode beipO2-Messungen im steady state. Pflügers Arch.320, 24 (1970).
Grunewald, W.: Accuracy and errors of thepO2-measurement by means of the platinum electrode and its calibration in vivo. Int. Sympos. on Modern Technology in Physiological Sciences, München 1971 (in press).
Huch, A., Huch, R., Lübbers, D. W.: Quantitative polarographische Sauerstoffdruckmessung auf der Kopfhaut des Neugeborenen. Arch. Gynäk.207, 443 (1969).
Huch, R.: Quantitative kontinuierliche Sauerstoffpartialdruckmessung (pO2-Messung) auf der hyperämisierten Erwachsenenhaut. Dissertation, Marburg 1971.
Kessler, M., Lübbers, D. W.: Aufbau und Anwendungsmöglichkeit verschiedenerP O 2-Elektroden. Pflügers Arch. ges. Physiol.291, 82 (1966).
Loew, P. G., Thews, G.: Die Altersabhängigkeit des arteriellen Sauerstoffdruckes bei der berufstätigen Bevölkerung. Klin. Wschr.40, 1092 (1962).
Pernkopf, E., Patzelt, V.: In: Handbuch für Haut- und Geschlechtskrankheiten. Epidermal Sub-Layers, Bd. I/1, S. 8. Hrsg.: O. Gans u. G. K. Steigleder. Berlin-Heidelberg-New York: Springer 1968.
Rooth, G., Sjöstedt, S., Caligara, F.: Bloodless determination of arterial oxygen tension by polarography. Science Tools, The LKW Instrument J.4, 37 (1957).
Author information
Authors and Affiliations
Additional information
Part of the results have been reported during the Workshop on Oxygen Transport in Tissue, 19–22 July, 1971, in Dortmund and at the “4. Deutsche Kongress für Perinatale Medizin”, 4–6 Nov. 1971, in Berlin. The study was carried out with partial support from the German Research Council (DFG).
Rights and permissions
About this article
Cite this article
Huch, R., Lübbers, D.W. & Huch, A. Quantitative continuous measurement of partial oxygen pressure on the skin of adults and new-born babies. Pflugers Arch. 337, 185–198 (1972). https://doi.org/10.1007/BF00586844
Received:
Issue Date:
DOI: https://doi.org/10.1007/BF00586844