Summary
A small angle stepping motor was used for construction of a micropositioner.
Linear movements are produced by direct coupling of the rotor axis to a high precision microdrive. The linearly moving system is constructed from stainless steal prismatic guides with hardened surfaces and permits precise steps in the 100 nm range.
Extreme reduction of the moving masses and minimal friction of the radial thrust bearing enables strong acceleration of the electrode. During simultaneous measurements of step performance motoneurons in the frog spinal cord, CA1 cells of hippocampal brain slices and glia cells in tissue culture were punctured with single electrodes (tip<1 μm) and double barrelled ion-sensitive microelectrodes (Ø 1,5–2 μm). In all three preparations, cell penetration could be performed by means of both types of electrodes with a high yield when the step velocity reached or exceeded 4 mm/s. Steps with lower velocity resulted in less successful cell penetrations and were accompanied by typical dimpling effects. The results indicate that a critical velocity is required for cell puncturing with a minimum of damage.
Similar content being viewed by others
References
Benninger C, Kadis J, Prince DA (1980) Extracellular calcium and potassium changes in hippocampal slices. Brain Res 187:165–182
Brown KT, Flaming DG (1977) New microelectrode techniques for intracellular work in small cells. Neuroscience 2:813–827
Bührle Ch-Ph, Sonnhof U (1981) Intracellular activities of K+, Na+, Ca2+ and Cl− in motoneurons and glia cells of the frog spinal cord. Pflügers Arch 389:R94
Bührle Ch-Ph, Sonnhof U (1980) Ion fluxes across the membrane of motoneurons during the action of glutamate demonstrated by measurements of intra and extracellular ion activities (aNa e/+i , aCa e/2+i , aK e/+i , aCl e/−i ). In: Proceedings of the Satellite Symposium on ion-selective microelectrodes and their use in excitable tissues of the XVIII. Int. Congress of Physiol. Sciences, Prague, Plenum Press, New York, 1981
Ellis GW (1962) Piezoelectric micromanipulators Science 138:84–91
Fish RM, Bryan JS, McReynolds JS, Ries JJ (1971): A mechanical microelectrode pulsing device to facilitate the penetration of small cells. IEEE Trans Bio Med Eng 18:240–241
Fromm M, Weskamp P, Hegel U (1980) Versatile piezoelectric driver for cell puncture. Pflügers Arch 384:69–73
Lassen UV, Rasmussen BE (1978) Use of microelectrodes for measurements of membrane potentials. In: Giebisch G, Tosteson DC, Ussing HH (eds) Membrane transport in biology, Ch 5 of Vol 1 Springer, Berlin, Heidelberg, New York, p 913
Peters M, Tetzel HD (1979) A high-speed piezoelectric microelectrode advancer. Pflügers Arch 379:R59
Prazma J (1978) Penetration of cells membrane by the piezoelectric driver. Experientia 34:1387
Ransom BR, Neale E, Henkart M, Bullock PN, Nelson PG (1977) Mouse spinal cord in cell culture. I. Morphology and intrinsic neuronal electrophysiologic properties. J Neurophysiol 40:1132–1150
Sonnhof U, Bührle Ch-Ph (1980) An analysis of glutamate induced ion fluxes across the membrane of spinal motoneurones of the frog. Adv Biochem Psychopharmacol 27:195–204
Sonnhof U, Grafe P, Krumnikl J, Linder M, Schindler L (1975) Inhibitory postsynaptic actions of taurine, GABA and other amino acids on motoneurons of the isolated frog spinal cord. Brain Res 100:327–341
Tupper JT, Rikmenspoel R (1969) Piezoelectric device for glass microelectrodes. Rev Sci Instrum 10:851–852
Author information
Authors and Affiliations
Rights and permissions
About this article
Cite this article
Sonnhof, U., Förderer, R., Schneider, W. et al. Cell puncturing with a step motor driven manipulator with simultaneous measurement of displacement. Pflugers Arch. 392, 295–300 (1982). https://doi.org/10.1007/BF00584314
Received:
Accepted:
Issue Date: