Identifizierung einiger Sprengstoffe mit Hilfe spezieller GC/MS-Techniken, insbesondere der PPNICI-Methode

Identification of some explosives by special GC/MS-techniques, particularly the PPNICI-method

  • Originalabhandlungen
  • Published:
Fresenius' Zeitschrift für analytische Chemie Aims and scope Submit manuscript

Summary

Twelve representative explosives were analysed by GC/MS using various ionization techniques (EI; CI) and reagent gases (methane, isobutane, ammonia); besides PICI (positive ion CI) the NICI (negative ion CI) method was applied. Samples excluded from gas chromatography were introduced by solid probe.

Under electron impact conditions the investigated substances showed spectra of only limited evidence, thus strongly indicating the use of Chemical Ionization.

From the tested reagent gases, isobutane fitted best (for PICI); the CI-spectra showed some peculiarities of analytical interest, as dimerizations, adduct formations and ions, for which no reasonable explanation can be given.

Like ammonia as reagent gas, which produces a much more intensive quasimolecular ion with some explosives, application of the PPNICI method offers some advantages: suppression of the ionization of irrelevant substances (like plasticizers) in the specimen leads to a slight increase in sensitivity; the superiority of ammonia as reagent gas can clearly be demonstrated by comparing the mass spectra of glycerol trinitrate obtained by the various ionization methods mentioned.

One of the main advantages of the PPNICI method is the fact, that positive and negative CI spectra can be recorded simultaneously during one run, thus minimizing consumption of sample material.

Zusammenfassung

Zwölf repräsentative Sprengstoffe wurden gas-chromatographisch/massenspektrometrisch mit Hilfe verschiedener Ionisierungsmethoden (EI; CI) und Reaktandgase (Methan, Isobutan, Ammoniak) untersucht; neben der PICI (positive ion CI) gelangte dabei die NICI (negative ion CI) zur Anwendung. Bei den nicht gas-chromatographierbaren Verbindungen erfolgte die Probenzuführung über den Direkteinlaß (DE).

Die elektronenstoßinduzierte Fragmentierung der Verbindungen führte im allgemeinen zu nur bedingt aussagekräftigen Spektren, die den Einsatz der CI erforderten.

Von den getesteten Reaktandgasen erwies sich Isobutan (für die PICI) als am geeignetsten; die CI-Spektren zeigten einige für die Analytik interessante Besonderheiten, wie Dimerisierungen, Adduktbildungen und das Auftreten von Ionen, deren Genese aus dem Probenmaterial nicht schlüssig erklärt werden kann.

Das Reaktandgas Ammoniak bietet bei der Analyse einiger der Verbindungen gewissen Vorteile (intensiveres Quasimolekülion), wie auch der Einsatz der NICI: so bedingt die Unterdrückung der Ionisation von Störsubstanzen in der Probensubstanz eine, wenn auch geringfügige, Steigerung der Empfindlichkeit gegenüber der PICI; die Vorteile bei der Verwendung von Ammoniak als Reaktand- bzw. Moderatorgas werden besonders beim direkten Vergleich der mit den verschiedenen Ionisierungsarten erhaltenen Spektren des Nitroglycerins deutlich.

Als spezieller Vorteil der PPNICI-Methode fällt ins Gewicht, daß die positiven und negativen CI-Spektren im Verlauf einer Analyse erhalten werden, der Verbrauch an Probenmaterial also gering gehalten werden kann.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Similar content being viewed by others

Literatur

  1. Volk, F., Schubert, H.: Explosivstoffe 16, 2 (1968) zit. u.a. bei [2, 19, 20]

    Google Scholar 

  2. Lit. bei: Vouros, P., Petersen, B. A., Colwell, L., Karger, B. L.: Anal. Chem. 49, 1039 (1977)

    Google Scholar 

  3. Lit. bei: Krull, I. S., Camp, M. J.: Int. Laboratory May/June, 15 (1980) (ausführliche Literaturübersicht)

  4. Koll, M., Wimmer, F., Fischer, K.: Arch. Krim. 130, 1 (1962)

    Google Scholar 

  5. Parker, R. G., McOwen, J. M., Cherolis, J. A.: J. Forens. Sci. 20, 254 (1975)

    Google Scholar 

  6. Peimer, R. E., Washington, W. D., Snow, K. B.: J. Forens. Sci. 25, 398 (1980)

    Google Scholar 

  7. Parker, R. G., Stephenson, M. O., McOwen, J. M., Cherolis, J. A.: J. Forens. Sci. 20, 133 (1975)

    Google Scholar 

  8. Tomita, Y., Mat, H. H., Guilbault, G. G.: Anal. Chem. 51, 1475 (1979)

    Google Scholar 

  9. Farey, M. G., Wilson, S. E.: J. Chromatogr. 114, 220 (1975)

    Google Scholar 

  10. Wheals, B. B.: J. Chromatogr. 122, 85 (1976)

    Google Scholar 

  11. Dalton, R. W., Chandler, C. D., Bolleter, W. T.: J. Chromatogr. Sci. 13, 40 (1975)

    Google Scholar 

  12. Doali, J. O., Juhasz, A. A.: J. Chrom. Sci. 12, 51 (1974)

    Google Scholar 

  13. Waters Assoc. Publ.: Forensic analysis by liquid chromatography. Explosives

  14. Lafleur, A. L., Morriseau, B. D.: Anal. Chem. 52, 1313 (1980)

    Google Scholar 

  15. Christos, T., Spinetti, L.: Rep. Invest. U.S. Bur. Mines RI 7795 (1973)

  16. Quinlin, W. T.: Rep. Atm. Energy Comm. U.S. MHSMP-74-8 (1973)

  17. Henry, R. A.: Application Lab. Rep. No. 7201, Du Pont Instr. Prod. Div., Hitchin.

  18. Fariwar-Mohsenie, M., Ripper, E., Habermann, K. H.: Fresenius Z. Anal. Chem. 296, 152 (1979)

    Google Scholar 

  19. Gillis, R. G., Lacey, M. J., Shannon, J. S.: Org. Mass Spectrom. 9, 359 (1974)

    Google Scholar 

  20. Yinon, J.: Biomed. Mass Spectrom. 1, 393 (1974)

    Google Scholar 

  21. Zitrin, S., Yinon, J.: Org. Mass Spectrom. 11, 388 (1976)

    Google Scholar 

  22. Michnowicz, J. A.: Explosives: Analysis using dual chemical/electron ionization GC/MS. Hewlett-Packard Appl. Note AN 176-8

  23. Hunt, D. F., Sethi, S. K.: Analytical applications of positive and negative ion chemical ionization mass spectrometry. In: High performance mass spectrometry: Chemical Applications (M. L. Gross, ed.), ACS Symposium Series 70 (1978)

  24. Hunt, D. F.: Selective reagents for chemical ionization mass spectrometry. Finnigan Spectra 6, no. 1 (1976)

    Google Scholar 

  25. Gielsdorf, W., Holz, H.: Dt. Apotheker-Ztg. 120, 1353 (1980)

    Google Scholar 

  26. Gielsdorf, W., Paulig, G., Herper, M.: Analysis of 2,5-di-methoxi-4-bromoamphetamme („DOB“) by glc-ms and tlc Microgram XIII, 128 (1980)

    Google Scholar 

  27. Gielsdorf, W., Toffel-Nadolny, P.: J. Clin. Chem. Clin. Biochem. 19, 25 (1981)

    Google Scholar 

  28. Gielsdorf, W., Schubert, K., Allin, K.: Arch. Krim. 166, 21 (1980)

    Google Scholar 

  29. Richter, W. J., Schwarz, H.: Angew. Chem. 90, 449 (1978)

    Google Scholar 

  30. Toffel-Nadolny, P., Gielsdorf, W.: Arzneimittelforsch. (Drug Res.) 31, 719 (1981)

    Google Scholar 

  31. Gielsdorf, W., Schubert, K.: J. Clin. Chem. Clin. Biochem. (im Druck) (1981)

  32. Gielsdorf, W.: Z. Rechtsmed. (im Druck) (1981)

  33. Kauert, G., Drasch, G., v. Meyer, L.: Beitr. gerichtl. Med. 37, 329 (1979)

    Google Scholar 

  34. Yinon, J.: J. Forens. Sci. 25, 401 (1980)

    Google Scholar 

  35. Hunt, D. F., Crow, F. W.: Anal. Chem. 50, 1781 (1978)

    Google Scholar 

  36. Hunt, D. F., Stafford, G. C., Crow, F. W., Russell, J. W.: Anal. Chem. 48, 2098 (1976)

    Google Scholar 

  37. Brandenberger, H.: Present state and future trends of negative ion mass spectrometry in toxicology, forensic and environmental chemistry. In: Recent developments in mass spectrometry in biochemistry and medicine, Vol. 2 (A. Frigerio, ed.). New York: Plenum Publ. Corp. 1979

    Google Scholar 

  38. Frangi-Schnyder, D., Brandenberger, H.: Fresenius Z. Anal. Chem. 290, 153 (1978)

    Google Scholar 

  39. Slaybeck, J. R. B., Kan, M. N.: Biomedical applications of negative ion chemical ionization mass spectrometry. Finnigan Application Report AR 8020

  40. Selected ion monitoring of malathion with MS 30 negative ion chemical ionization. Kratos Data Sheet No. 108

  41. Melton, C. E.: Negative ion mass spectra. In: Mass spectrometry of organic ions (F. W. McLafferty, ed.). New York, London: Acad. Press 1963

    Google Scholar 

  42. Bruins, A. P.: Biomed. Mass. Sprectrom. 7 (10), 454 (1980)

    Google Scholar 

  43. Dillard, J. G.: Negative ions. In: Biomedical applications of mass spectrometry, First Suppl. Vol. (G. R. Waller and O. C. Dermer, eds.). New York: John Wiley & Sons 1980

    Google Scholar 

Download references

Author information

Authors and Affiliations

Authors

Additional information

Auszugsweise vorgetragen auf dem 4. Symposium für Toxikologie. Berlin, 20.–22. Mai 1981

Rights and permissions

Reprints and permissions

About this article

Cite this article

Gielsdorf, W. Identifizierung einiger Sprengstoffe mit Hilfe spezieller GC/MS-Techniken, insbesondere der PPNICI-Methode. Z. Anal. Chem. 308, 123–128 (1981). https://doi.org/10.1007/BF00469440

Download citation

  • Received:

  • Issue Date:

  • DOI: https://doi.org/10.1007/BF00469440

Key words

Navigation