Summary
Twelve representative explosives were analysed by GC/MS using various ionization techniques (EI; CI) and reagent gases (methane, isobutane, ammonia); besides PICI (positive ion CI) the NICI (negative ion CI) method was applied. Samples excluded from gas chromatography were introduced by solid probe.
Under electron impact conditions the investigated substances showed spectra of only limited evidence, thus strongly indicating the use of Chemical Ionization.
From the tested reagent gases, isobutane fitted best (for PICI); the CI-spectra showed some peculiarities of analytical interest, as dimerizations, adduct formations and ions, for which no reasonable explanation can be given.
Like ammonia as reagent gas, which produces a much more intensive quasimolecular ion with some explosives, application of the PPNICI method offers some advantages: suppression of the ionization of irrelevant substances (like plasticizers) in the specimen leads to a slight increase in sensitivity; the superiority of ammonia as reagent gas can clearly be demonstrated by comparing the mass spectra of glycerol trinitrate obtained by the various ionization methods mentioned.
One of the main advantages of the PPNICI method is the fact, that positive and negative CI spectra can be recorded simultaneously during one run, thus minimizing consumption of sample material.
Zusammenfassung
Zwölf repräsentative Sprengstoffe wurden gas-chromatographisch/massenspektrometrisch mit Hilfe verschiedener Ionisierungsmethoden (EI; CI) und Reaktandgase (Methan, Isobutan, Ammoniak) untersucht; neben der PICI (positive ion CI) gelangte dabei die NICI (negative ion CI) zur Anwendung. Bei den nicht gas-chromatographierbaren Verbindungen erfolgte die Probenzuführung über den Direkteinlaß (DE).
Die elektronenstoßinduzierte Fragmentierung der Verbindungen führte im allgemeinen zu nur bedingt aussagekräftigen Spektren, die den Einsatz der CI erforderten.
Von den getesteten Reaktandgasen erwies sich Isobutan (für die PICI) als am geeignetsten; die CI-Spektren zeigten einige für die Analytik interessante Besonderheiten, wie Dimerisierungen, Adduktbildungen und das Auftreten von Ionen, deren Genese aus dem Probenmaterial nicht schlüssig erklärt werden kann.
Das Reaktandgas Ammoniak bietet bei der Analyse einiger der Verbindungen gewissen Vorteile (intensiveres Quasimolekülion), wie auch der Einsatz der NICI: so bedingt die Unterdrückung der Ionisation von Störsubstanzen in der Probensubstanz eine, wenn auch geringfügige, Steigerung der Empfindlichkeit gegenüber der PICI; die Vorteile bei der Verwendung von Ammoniak als Reaktand- bzw. Moderatorgas werden besonders beim direkten Vergleich der mit den verschiedenen Ionisierungsarten erhaltenen Spektren des Nitroglycerins deutlich.
Als spezieller Vorteil der PPNICI-Methode fällt ins Gewicht, daß die positiven und negativen CI-Spektren im Verlauf einer Analyse erhalten werden, der Verbrauch an Probenmaterial also gering gehalten werden kann.
Similar content being viewed by others
Literatur
Volk, F., Schubert, H.: Explosivstoffe 16, 2 (1968) zit. u.a. bei [2, 19, 20]
Lit. bei: Vouros, P., Petersen, B. A., Colwell, L., Karger, B. L.: Anal. Chem. 49, 1039 (1977)
Lit. bei: Krull, I. S., Camp, M. J.: Int. Laboratory May/June, 15 (1980) (ausführliche Literaturübersicht)
Koll, M., Wimmer, F., Fischer, K.: Arch. Krim. 130, 1 (1962)
Parker, R. G., McOwen, J. M., Cherolis, J. A.: J. Forens. Sci. 20, 254 (1975)
Peimer, R. E., Washington, W. D., Snow, K. B.: J. Forens. Sci. 25, 398 (1980)
Parker, R. G., Stephenson, M. O., McOwen, J. M., Cherolis, J. A.: J. Forens. Sci. 20, 133 (1975)
Tomita, Y., Mat, H. H., Guilbault, G. G.: Anal. Chem. 51, 1475 (1979)
Farey, M. G., Wilson, S. E.: J. Chromatogr. 114, 220 (1975)
Wheals, B. B.: J. Chromatogr. 122, 85 (1976)
Dalton, R. W., Chandler, C. D., Bolleter, W. T.: J. Chromatogr. Sci. 13, 40 (1975)
Doali, J. O., Juhasz, A. A.: J. Chrom. Sci. 12, 51 (1974)
Waters Assoc. Publ.: Forensic analysis by liquid chromatography. Explosives
Lafleur, A. L., Morriseau, B. D.: Anal. Chem. 52, 1313 (1980)
Christos, T., Spinetti, L.: Rep. Invest. U.S. Bur. Mines RI 7795 (1973)
Quinlin, W. T.: Rep. Atm. Energy Comm. U.S. MHSMP-74-8 (1973)
Henry, R. A.: Application Lab. Rep. No. 7201, Du Pont Instr. Prod. Div., Hitchin.
Fariwar-Mohsenie, M., Ripper, E., Habermann, K. H.: Fresenius Z. Anal. Chem. 296, 152 (1979)
Gillis, R. G., Lacey, M. J., Shannon, J. S.: Org. Mass Spectrom. 9, 359 (1974)
Yinon, J.: Biomed. Mass Spectrom. 1, 393 (1974)
Zitrin, S., Yinon, J.: Org. Mass Spectrom. 11, 388 (1976)
Michnowicz, J. A.: Explosives: Analysis using dual chemical/electron ionization GC/MS. Hewlett-Packard Appl. Note AN 176-8
Hunt, D. F., Sethi, S. K.: Analytical applications of positive and negative ion chemical ionization mass spectrometry. In: High performance mass spectrometry: Chemical Applications (M. L. Gross, ed.), ACS Symposium Series 70 (1978)
Hunt, D. F.: Selective reagents for chemical ionization mass spectrometry. Finnigan Spectra 6, no. 1 (1976)
Gielsdorf, W., Holz, H.: Dt. Apotheker-Ztg. 120, 1353 (1980)
Gielsdorf, W., Paulig, G., Herper, M.: Analysis of 2,5-di-methoxi-4-bromoamphetamme („DOB“) by glc-ms and tlc Microgram XIII, 128 (1980)
Gielsdorf, W., Toffel-Nadolny, P.: J. Clin. Chem. Clin. Biochem. 19, 25 (1981)
Gielsdorf, W., Schubert, K., Allin, K.: Arch. Krim. 166, 21 (1980)
Richter, W. J., Schwarz, H.: Angew. Chem. 90, 449 (1978)
Toffel-Nadolny, P., Gielsdorf, W.: Arzneimittelforsch. (Drug Res.) 31, 719 (1981)
Gielsdorf, W., Schubert, K.: J. Clin. Chem. Clin. Biochem. (im Druck) (1981)
Gielsdorf, W.: Z. Rechtsmed. (im Druck) (1981)
Kauert, G., Drasch, G., v. Meyer, L.: Beitr. gerichtl. Med. 37, 329 (1979)
Yinon, J.: J. Forens. Sci. 25, 401 (1980)
Hunt, D. F., Crow, F. W.: Anal. Chem. 50, 1781 (1978)
Hunt, D. F., Stafford, G. C., Crow, F. W., Russell, J. W.: Anal. Chem. 48, 2098 (1976)
Brandenberger, H.: Present state and future trends of negative ion mass spectrometry in toxicology, forensic and environmental chemistry. In: Recent developments in mass spectrometry in biochemistry and medicine, Vol. 2 (A. Frigerio, ed.). New York: Plenum Publ. Corp. 1979
Frangi-Schnyder, D., Brandenberger, H.: Fresenius Z. Anal. Chem. 290, 153 (1978)
Slaybeck, J. R. B., Kan, M. N.: Biomedical applications of negative ion chemical ionization mass spectrometry. Finnigan Application Report AR 8020
Selected ion monitoring of malathion with MS 30 negative ion chemical ionization. Kratos Data Sheet No. 108
Melton, C. E.: Negative ion mass spectra. In: Mass spectrometry of organic ions (F. W. McLafferty, ed.). New York, London: Acad. Press 1963
Bruins, A. P.: Biomed. Mass. Sprectrom. 7 (10), 454 (1980)
Dillard, J. G.: Negative ions. In: Biomedical applications of mass spectrometry, First Suppl. Vol. (G. R. Waller and O. C. Dermer, eds.). New York: John Wiley & Sons 1980
Author information
Authors and Affiliations
Additional information
Auszugsweise vorgetragen auf dem 4. Symposium für Toxikologie. Berlin, 20.–22. Mai 1981
Rights and permissions
About this article
Cite this article
Gielsdorf, W. Identifizierung einiger Sprengstoffe mit Hilfe spezieller GC/MS-Techniken, insbesondere der PPNICI-Methode. Z. Anal. Chem. 308, 123–128 (1981). https://doi.org/10.1007/BF00469440
Received:
Issue Date:
DOI: https://doi.org/10.1007/BF00469440