Abstract
A novel micropipette holder has been designed to overcome the problem of drift enountered when experimenting on small preparations using patchclamp techniques. In order to increase the stiffness and thermal stability of the system, the holder is designed such that the micropipette is clamped to a rigid quartz rod, and isolated from stresses arising in the connectors. The pressure line, made from stiff tubing, and electrical leads are firmly attached to the table and micromanipulator. This design was able to hold the micropipette tip stable to with 100 nm for more than 1 h, even with changes of 26.7 kPa (200 mmHg) in the pressure line.
Similar content being viewed by others
References
Hamill OP, Marty A, Neher E, Sakmann B, Sigworth FJ (1981) Improved patch-clamp techniques for high-resolution current recording from cells and cell-free membrane patches. Pflügers Arch 391:85–100
Sokabe M, Sachs F (1990) The structure and dynamics of patch-clamped membranes: a study by differential interference microscopy. J Cell Biol 111:599–606
Weast RC. Handbook of chemistry and physics. CRC Cleveland, Ohio
Author information
Authors and Affiliations
Rights and permissions
About this article
Cite this article
Sachs, F. A low drift micropipette holder. Pflugers Arch. 429, 434–435 (1995). https://doi.org/10.1007/BF00374160
Received:
Accepted:
Issue Date:
DOI: https://doi.org/10.1007/BF00374160