This study describes the early investigation of a new VMAT planning module developed by Philips for the Pinnacle TPS.

The Pinnacle solution for VMAT planning, SmartArc, was evaluated for example prostate patients including inverse planning, export, delivery and verification. The quality of verification was investigated as the gantry spacing of the arc used in the final dose calculation on the TPS was varied. The inverse planning was performed using the same objectives as for our established prostate IMRT programme. Verification of the treatment plan was performed using the Delta4 phantom (Scandidos, Sweden) on an Elekta Synergy linac. The robustness of delivery to machine faults such as gantry overshoots, prf drop-outs, beam interruptions, and communication drop-outs was investigated.

SmartArc was able to produce clinically acceptable plans that were at least comparable to the standard IMRT plans. The time for delivery was around 100 seconds compared to 220 seconds for the IMRT plans, mostly due to the separate loading of each step-and-shoot IMRT beam compared to a single button push for VMAT. The quality of the verification was strongly influenced by the gantry spacing of the final dose calculation on Pinnacle, with <60% of pixels with a Γ<1 using a 2mm/2% tolerance for 8° spacing compared to >90% when using 4° for the final gantry spacing. However, the time for delivery increased from ~1.6 min to ~2.2min when the final gantry spacing was reduced. Repeat deliveries with a large range of faults and terminations induced did not degrade the delivered solution highlighted the robustness of the linac delivery.

The SmartArc module developed within the Pinnacle TPS is able to produce clinically acceptable plans, provided a final gantry spacing of ≤4° is employed.

This is a preview of subscription content, log in via an institution to check access.

Access this chapter

Chapter
USD 29.95
Price excludes VAT (USA)
  • Available as PDF
  • Read on any device
  • Instant download
  • Own it forever
eBook
USD 259.00
Price excludes VAT (USA)
  • Available as PDF
  • Read on any device
  • Instant download
  • Own it forever
Softcover Book
USD 329.99
Price excludes VAT (USA)
  • Compact, lightweight edition
  • Dispatched in 3 to 5 business days
  • Free shipping worldwide - see info

Tax calculation will be finalised at checkout

Purchases are for personal use only

Institutional subscriptions

Preview

Unable to display preview. Download preview PDF.

Unable to display preview. Download preview PDF.

Author information

Authors and Affiliations

Authors

Editor information

Editors and Affiliations

Rights and permissions

Reprints and permissions

Copyright information

© 2009 Springer-Verlag Berlin Heidelberg

About this paper

Cite this paper

Rowbottom, C.G., Golby, C., Atherton, S., Mackay, R.I. (2009). Investigation into the Pinnacle SmartArc Module for VMAT Planning. In: Dössel, O., Schlegel, W.C. (eds) World Congress on Medical Physics and Biomedical Engineering, September 7 - 12, 2009, Munich, Germany. IFMBE Proceedings, vol 25/1. Springer, Berlin, Heidelberg. https://doi.org/10.1007/978-3-642-03474-9_203

Download citation

  • DOI: https://doi.org/10.1007/978-3-642-03474-9_203

  • Publisher Name: Springer, Berlin, Heidelberg

  • Print ISBN: 978-3-642-03472-5

  • Online ISBN: 978-3-642-03474-9

  • eBook Packages: EngineeringEngineering (R0)

Publish with us

Policies and ethics