Abstract
A simple method of temperature stabilization on cold stage of 4 K Gifford-McMahon (GM) cryocooler has been proposed. A copper pot connected to compressor unit by a stainless steel capillary tube is mounted on the 4 K stage of a GM cryocooler. Depended on the temperature of pot, pressurized helium can naturally go into or out of the pot. The utilization of high volumetric specific heat of pressurized helium in the pot at 4 K region produces a good temperature stability on the 4 K stage. The periodic temperature fluctuation of the 4 K stage is ∼ 0.5 K (peak-to- peak) typically, but the high pressure pot described in the present paper reduces the temperature fluctuation effectively down to ∼ 0.05 K (peak-to-peak). The method is safe to deal with, and the system is easy to operate. This paper shows the experimental details, and discusses the temperature stabilization effect of the pot and other advantages of the method.
Access this chapter
Tax calculation will be finalised at checkout
Purchases are for personal use only
Preview
Unable to display preview. Download preview PDF.
References
Hashimoto, T., Li, R., Kuriyama, T. and Nagao, M., “Recent progress in application of magnetic regenerator material” ,Cryogenic Engineering (Journal of the Cryogenic Society of Japan),Vol. 31 (1996), pp. 131–149.
Satoh, T., Onishi, A., Li, R., Asami, H. and Kanazawa, Y., “Development of 1.5 W4 K G-M cryocooler with magnetic regenerator material” , Adv. Cryog. Eng., Vol. 41, Plenum Press, New York (1996), in press.
Li, R., Onishi, A., Satoh, T. and Kanazawa, Y., “Influence of valve open timing and interval on performance of 4 K Gifford-McMahon cycle cryocooler” , Adv. Cryog. Eng., Vol. 41, Plenum Press, New York (1996), in press.
Plambeck, R., Thatte, N. and Sykes, P., “A4K Gifford-McMahon refrigerator for astronomy”, Proc. 7th International Cryocooler Conference, Air Force Phillips Laboratory Report, PL-CP-93-1001, Kirtland, AF, NM (1993), pp. 401–415.
Takahashi, M., Hatakeyama, H., Kuriyama, T., Nakagome, H., Kawabe, R., Iwashita, H., McCulloch, G., Shibata, K. and Ukita, S., “A compact 150 GHz SIS receiver cooled by 4 K GM refrigerator”, Proc. 7th International Cryocooler Conference, Air Force Phillips Laboratory Report, PL-CP-93-1001, Kirtland, AF, NM (1993), pp. 495–507.
Fujimoto, S., Kazami, K., Takada, Y., Yoshida, T., Ogata, H. and Kado, H., “Cooling of SQUIDs using a Gifford-McMahon cryocooler containing magnetic regenerative material to measure biomagnetism”, Cryogenics, Vol. 35 (1995), pp. 143–148.
Kaiser, G., Seidel, P. and Thurk, M., “Noise reduction of cryo-refrigerators” , Adv. Cryog. Eng., Vol. 39, Plenum Press, New York (1994), pp. 1281–1285.
Kaiser, G., Dorrer, L., Matthes, A., Seidel, P., Schmidl, F., Schneidewind, H. and Thurk, M., “Cooling of HTSC Josephson junctions and SQUIDs with cryo-refrigerators”, Cryogenics, Vol. 34, ICEC Supplement (1994), pp. 891–894.
Author information
Authors and Affiliations
Editor information
Editors and Affiliations
Rights and permissions
Copyright information
© 1997 Springer Science+Business Media New York
About this chapter
Cite this chapter
Li, R., Onishi, A., Satoh, T., Kanazawa, Y. (1997). Temperature Stabilization on Cold Stage of 4 K G-M Cryocooler. In: Ross, R.G. (eds) Cryocoolers 9. Springer, Boston, MA. https://doi.org/10.1007/978-1-4615-5869-9_87
Download citation
DOI: https://doi.org/10.1007/978-1-4615-5869-9_87
Publisher Name: Springer, Boston, MA
Print ISBN: 978-1-4613-7691-0
Online ISBN: 978-1-4615-5869-9
eBook Packages: Springer Book Archive