Abstract
An internal solid contact sensor (ISCS) for the determination of methacycline hydrochloride (MC·Cl), Pt/PPy/PVC(MC-PT), is described, based on the use of conducting poly(pyrrole) (PPy) as solid contact material and MC-phosphotungstate (PT) as the ion exchanger and dibutyl phthalate (DBP) as the plasticizer. A direct potentiometric method for the quantitative analysis of MC·Cl is also described. Under the condition of pH 2.7, the linear concentration range, slope (25˚C) and detection limit of the sensor are 6.4 × 1.0-6 - 3.0 × 1.0-3 M, 52.4 ± 0.2 mV/decade and 4.4 × 1.0-6 M, respectively. The response time is >5 s. The determinations of MC·Cl in tablets were carried out by direct potentiometry. The average recovery and relative standard deviation are 100.1 and 0.7% (n = 4), respectively.
References
L. J. Nagels and I. Poels, Trends Anal. Chem., 2000, 19(7), 410.
W. A. Moats, J. Agric. Food Chem., 2000, 48, 2244.
J. S. Yu, E. M. Sena, and W. Paulus, Cancer Res., 1996, 56, 5423.
M. Marble, Immunotherapy Weekly, 1999, 7(26), 10.
X. X. Sun, X. Zhang, and H. Y. Aboul-Enein, IL. FARMACO, 2004, 59, 307.
X. X. Sun and H. Y. Aboul-Enein, Talanta, 2002, 58, 387.
U. S. Pharmacopoeia, XXIII, 1995, The United States Phamacopoeia Convetion Inc., Rockville, MD.
Chinese Pharmacopoeia, Part II, 2000, Commission of the Ministry of Health Beijing.
X. X. Sun, Fenxi Ceshi Xuebao, 2003, 4, 1.
H. Y. Aboul-Enein, X. X. Sun, and C. J. Sun, Sensors, 2002, 2, 424.
R. P. Buck and E. Lindner, Pure Appl. Chem., 1994, 66(12), 2527.
Author information
Authors and Affiliations
Rights and permissions
About this article
Cite this article
Sun, X.X., Aboul-Enein, H.Y. Quantitative Analysis of Methacycline Hydrochloride by Direct Potentiometry Using the Internal Solid Contact Sensor. ANAL. SCI. 23, 231–233 (2007). https://doi.org/10.2116/analsci.23.231
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.2116/analsci.23.231