Polyethylene liner cementation into a well-fixed metal acetabular shell for the management of periacetabular osteolysis: a systematic review

  • General Review
  • Published:
European Journal of Orthopaedic Surgery & Traumatology Aims and scope Submit manuscript

Abstract

Purpose

Although various papers have reported on the clinical performance of cup retention with cementation of a new liner and bone grafting in the management of well-fixed cups with polyethylene wear and periacetabular osteolysis after total hip arthroplasty (THA), no systematic review of this topic has been published to date.

Methods

Medline, EMBASE and Cochrane Library were searched for articles published from January 1999 to January 2019 using “osteolysis” AND “well-fixed”, “osteolysis” AND “retro-acetabular”, “bone graft” AND (“retention” OR “retained” OR “stable”) AND “cup”, and “cemented liner” AND “well-fixed”.

Results

Nine articles were selected for review (186 cases, 76.1 months mean follow-up). The overall revision rate was 11.3% (21 hips) most commonly due to aseptic loosening (9/186 hips), dislocation (8/186 hips), and liner wear progression (2/186 cases). The reported square size of osteolytic lesions ranged from a mean of 465.84 mm2 to a max of 4,770 mm2. Almost all reported lesions treated with bone grafts resolved or did not progress 97% (72/74). All studies indicated improved pain and functional scores at follow-up.

Conclusion

Cementation of a new liner with periacetabular bone grafting provides an alternative option to isolated liner exchange and cup revision for the management of periacetabular osteolysis in well-fixed cups with a disrupted locking mechanism or unavailable exchange liner. Further higher quality studies are required in order to examine if the use of highly cross-linked polyethylene, highly porous-coated cups, hydroxyapatite-coated cups, and small-diameter cups influence the clinical outcome of liner cementation in well-fixed cups with periacetabular osteolysis.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Fig. 1

Similar content being viewed by others

References

  1. Kim YH, Kim JS, Cho SH (1999) Primary total hip arthroplasty with a cementless porous-coated anatomic total hip prosthesis: 10- to 12-year results of prospective and consecutive series. J Arthroplasty 14(5):538–548

    Article  CAS  Google Scholar 

  2. Bozic KJ, Kurtz SM, Lau E, Ong K, Vail TP, Berry DJ (2009) The epidemiology of revision total hip arthroplasty in the United States. J Bone Joint Surg Am 91(1):128–133

    Article  Google Scholar 

  3. Delaunay C, Hamadouche M, Girard J, Duhamel A (2013) SoFCOT Group: What are the causes for failures of primary hip arthroplasties in France? Clin Orthop Relat Res 471(12):3863–3869

    Article  Google Scholar 

  4. Hozack WJ, Mesa JJ, Carey C, Rothman RH. Relationship between polyethylene wear, pelvic osteolysis, and clinical symptomatology in patients with cementless acetabular components. A framework for decision making. J Arthroplasty 1996;11(7):769–72.

  5. Hall A, Eilers M, Hansen R, et al: Advances in acetabular reconstruction in revision total hip arthroplasty: maximizing function and outcomes after treatment of periacetabular osteolysis around the well-fixed shell. J Bone Joint Surg Am. 2013;18;95(18):1709–18.

  6. Maloney WJ, Galante JO, Anderson M et al (1999) Fixation, polyethylene wear, and pelvic osteolysis in primary total hip replacement. Clin Orthop Relat Res 369:157–164

    Article  Google Scholar 

  7. Sandgren B, Crafoord J, Olivecrona H, Garellick G, Weidenhielm L: Risk factors for periacetabular osteolysis and wear in asymptomatic patients with uncemented total hip arthroplasties. ScientificWorldJournal 2014;2014:905818.

  8. Lie SA, Hallan G, Furnes O, Havelin LI, Engesaeter LB (2007) Isolated acetabular liner exchange compared with complete acetabular component revision in revision of primary uncemented acetabular components: a study of 1649 revisions from the Norwegian Arthroplasty Register. J Bone Joint Surg Br 89(5):591–594

    Article  CAS  Google Scholar 

  9. Talmo CT, Kwon YM, Freiberg AA, Rubash HE, Malchau H (2011) Management of polyethylene wear associated with a well-fixed modular cementless shell during revision total hip arthroplasty. J Arthroplasty 26(4):576–581

    Article  Google Scholar 

  10. Maloney WJ, Herzwurm P, Paprosky W, Rubash HE, Engh CA (1997) Treatment of pelvic osteolysis associated with a stable acetabular component inserted without cement as part of a total hip replacement. J Bone Joint Surg Am 79(11):1628–1634

    Article  CAS  Google Scholar 

  11. Beaulé PE, Ebramzadeh E, Le Duff M, Prasad R, Amstutz HC (2004) Cementing a liner into a stable cementless acetabular shell: the double-socket technique. J Bone Joint Surg Am 86(5):929–934

    Article  Google Scholar 

  12. Schmalzried TP, Fowble VA, Amstutz HC: The fate of pelvic osteolysis after reoperation. No recurrence with lesional treatment. Clin Orthop Relat Res 1998;(350):128–37.

  13. Maloney WJ, Paprosky W, Engh CA, Rubash H (2001) Surgical treatment of pelvic osteolysis. Clin Orthop Relat Res 393:78–84

    Article  Google Scholar 

  14. Restrepo C, Ghanem E, Houssock C, Austin M, Parvizi J, Hozack WJ (2009) Isolated polyethylene exchange versus acetabular revision for polyethylene wear. Clin Orthop Relat Res 467(1):194–198

    Article  Google Scholar 

  15. LaPorte DM, Mont MA, Pierre-Jacques H, Peyton RS, Hungerford DS (1998) Technique for acetabular liner revision in a nonmodular metal-backed component. J Arthroplasty 13(3):348–350

    Article  CAS  Google Scholar 

  16. Hofmann AA, Prince EJ, Drake FT, Hunt KJ (2009) Cementation of a polyethylene liner into a metal acetabular shell: a biomechanical study. J Arthroplasty 24(5):775–782

    Article  Google Scholar 

  17. Atkins D, Eccles M, Flottorp S et al (2004) Systems for grading the quality of evidence and the strength of recommendations I: critical appraisal of existing approaches The GRADE Working Group. BMC Health Serv Res 4(1):38

    Article  Google Scholar 

  18. Hankins DA, Fletcher IE, Prieto F et al (2019) Critical evaluation of the methodologic quality of the Top 50 cited articles relating to knee dislocation and multiligamentous knee injury. Orthop J Sports Med 7(11):2325967119880505

    Article  Google Scholar 

  19. Kang P, Yang J, Zhou Z, Shen B, Pei F (2012) Retention of a well-fixed acetabular component in the setting of acetabular osteolysis. Int Orthop 36(5):949–954

    Article  Google Scholar 

  20. Narkbunnam R, Amanatullah DF, Electricwala AJ, Huddleston JI 3rd, Maloney WJ, Goodman SB (2017) Outcome of 4 surgical treatments for wear and osteolysis of cementless acetabular components. J Arthroplasty 32(9):2799–2805

    Article  Google Scholar 

  21. Rivkin G, Kandel L, Qutteineh B, Liebergall M, Mattan Y (2015) Long term results of liner polyethylene cementation technique in revision for peri-acetabular osteolysis. J Arthroplasty 30(6):1041–1043

    Article  Google Scholar 

  22. Stamenkov R, Neale SD, Kane T, Findlay DM, Taylor DJ, Howie DW (2014) Cemented liner exchange with bone grafting halts the progression of periacetabular osteolysis. J Arthroplasty 29(4):822–826

    Article  Google Scholar 

  23. Deheshi BM, Allen DJ, Kim PR (2008) Treatment of retroacetabular osteolysis with calcium sulfate and retention of original components. J Arthroplasty 23(8):1240.e9–12

    Article  Google Scholar 

  24. Yoon TR, Seon JK, Song EK, Chung JY, Seo HY, Park YB (2005) Cementation of a metal-inlay polyethylene liner into a stable metal shell in revision total hip arthroplasty. J Arthroplasty 20(5):652–657

    Article  Google Scholar 

  25. Lim SJ, Lee KH, Park SH, Park YS (2014) Medium-term results of cementation of a highly cross-linked polyethylene liner into a well-fixed acetabular shell in revision hip arthroplasty. J Arthroplasty 29(3):634–637

    Article  Google Scholar 

  26. Park MS, Yoon SJ, Lee JR (2015) Outcomes of polyethylene liner cementation into a fixed metal acetabular shell with minimum follow-up of 7 years. Hip Int 25(1):61–66

    Article  Google Scholar 

  27. Koh KH, Moon YW, Lim SJ, Lee HI, Shim JW, Park YS (2011) Complete acetabular cup revision versus isolated liner exchange for polyethylene wear and osteolysis without loosening in cementless total hip arthroplasty. Arch Orthop Trauma Surg 131(11):1591–1600

    Article  Google Scholar 

  28. DeLee JG, Charnley J (1976) Radiological demarcation of cemented sockets in total hip replacement. Clin Orthop Relat Res 121:20–32

    Google Scholar 

  29. Smith TM, Berend KR, Lombardi AV Jr, Mallory TH, Russell JH (2005) Isolated liner exchange using the anterolateral approach is associated with a low risk of dislocation. Clin Orthop Relat Res 441:221–226

    Article  Google Scholar 

  30. Griffin WL, Fehring TK, Mason JB, McCoy TH, Odum S, Terefenko CS (2004) Early morbidity of modular exchange for polyethylene wear and osteolysis. J Arthroplasty 19(7 Suppl 2):61–66

    Article  Google Scholar 

  31. Engh CA Jr, Hopper RH Jr, Huynh C, Ho H, Sritulanondha S, Engh CA Sr (2012) A prospective, randomized study of cross-linked and non-cross-linked polyethylene for total hip arthroplasty at 10-year follow-up. J Arthroplasty 27(8 Suppl):2-7.e1

    Article  Google Scholar 

  32. Stilling M, Rahbek O, Søballe K (2009) Inferior survival of hydroxyapatite versus titanium-coated cups at 15 years. Clin Orthop Relat Res 467(11):2872–2879

    Article  Google Scholar 

  33. Bobyn JD, Stackpool GJ, Hacking SA, Tanzer M, Krygier JJ (1999) Characteristics of bone ingrowth and interface mechanics of a new porous tantalum biomaterial. J Bone Joint Surg Br 81(5):907–914

    Article  CAS  Google Scholar 

  34. Malahias MA, Kostretzis L, Greenberg A, Nikolaou VS, Atrey A, Sculco PK (2020) Highly porous titanium acetabular components in primary and revision total hip arthroplasty: a systematic review. J Arthroplasty 35(6):1737–1749

    Article  Google Scholar 

  35. Berry DJ, von Knoch M, Schleck CD, Harmsen WS (2005) Effect of femoral head diameter and operative approach on risk of dislocation after primary total hip arthroplasty. J Bone Joint Surg Am 87(11):2456–2463

    PubMed  Google Scholar 

  36. Ritter MA, Harty LD, Keating ME, Faris PM, Meding JB (2001) A clinical comparison of the anterolateral and posterolateral approaches to the hip. Clin Orthop Relat Res 385:95–99

    Article  Google Scholar 

  37. Claus AM, Totterman SM, Sychterz CJ, Tamez-Peña JG, Looney RJ, Engh CA Sr (2004) Computed tomography to assess pelvic lysis after total hip replacement. Clin Orthop Relat Res 422:167–174

    Article  Google Scholar 

  38. Puri L, Wixson RL, Stern SH, Kohli J, Hendrix RW, Stulberg SD (2002) Use of helical computed tomography for the assessment of acetabular osteolysis after total hip arthroplasty. J Bone Joint Surg Am 84(4):609–614

    Article  Google Scholar 

  39. Paprosky WG, Perona PG, Lawrence JM: Acetabular defect classification and surgical reconstruction in revision arthroplasty. A 6-year follow-up evaluation. J Arthroplasty 1994;9(1):33–44.

  40. Saleh KJ, Holtzman J, Gafni ASaleh L, et al (2001) Development, test reliability and validation of a classification for revision hip arthroplasty. J Orthop Res 19(1):50–56

    Article  CAS  Google Scholar 

Download references

Funding

This research did not receive any specific grant from funding agencies in the public, commercial, or not-for-profit sectors.

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Seong J. Jang.

Ethics declarations

Conflict of interest

All authors declare that they have no conflict of interest. No benefits in any form have been received or will be received from a commercial party related directly or indirectly to the subject of this article.

Additional information

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Rights and permissions

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Malahias, MA., Ma, QL., Jang, S.J. et al. Polyethylene liner cementation into a well-fixed metal acetabular shell for the management of periacetabular osteolysis: a systematic review. Eur J Orthop Surg Traumatol 32, 1459–1468 (2022). https://doi.org/10.1007/s00590-021-03130-w

Download citation

  • Received:

  • Accepted:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s00590-021-03130-w

Keywords

Navigation