Abstract
Purpose
There are many reports of corneal complications caused by argon laser peripheral iridotomy. In this study, we investigated whether cooling the anterior ocular segment during laser iridotomy prevented corneal damage.
Methods
A space for cooling the anterior ocular segment by perfusion with ice-cold water was made between the cornea and a contact lens. Dutch pigmented rabbits were excessively irradiated by an argon green laser, aiming at the peripheral iris. We used a contact lens without a cooling system as a control. Temperature in the anterior chamber and intraocular pressure were also monitored throughout the experiment.
Results
During laser treatment, the temperature without the cooling system rose to a maximum of 44.5°C in the anterior chamber, whereas use of the cooling system consistently kept this temperature at 11.1°–16.1°C. Although most eyes treated without cooling showed corneal damage, damage was seen in only a few or in no eyes cooled during laser treatment.
Conclusions
Argon laser treatment using contact lenses with a cooling system drastically reduced the corneal damage induced by argon laser peripheral iridotomy. This technique may be acceptable for clinical use, considering its technical simplicity and low incidence of treatment-related complications. Jpn J Ophthalmol 2007;51:317–324 © Japanese Ophthalmological Society 2007
Similar content being viewed by others
References
RK Abraham (1981) ArticleTitleProtocol for single-session argon laser iridectomy for angle-closure glaucoma Int Ophthalmol Clin 21 145–166 Occurrence Handle1:STN:280:DyaL3M7mslegug%3D%3D Occurrence Handle10.1097/00004397-198102110-00016 Occurrence Handle7194324
R Ritch (1982) ArticleTitleArgon laser treatment for medically unresponsive attacks of angle-closure glaucoma Am J Ophthalmol 94 197–204 Occurrence Handle1:STN:280:DyaL383otFGmsQ%3D%3D Occurrence Handle10.1016/0002-9394(82)90075-7 Occurrence Handle7114142
RL Stamper MF Lieberman MV Drake (1999) Becker-Shaffer's diagnosis and therapy of the glaucomas EditionNumber7th ed Mosby St. Louis
C Hong Y Kitazawa T Tanishima (1983) ArticleTitleInfluence of argon laser treatment of glaucoma on corneal endothelium Jpn J Ophthalmol 27 567–574 Occurrence Handle1:STN:280:DyaL2c7jvFGktA%3D%3D Occurrence Handle6668749
AL Schwartz NF Martin PA Weber (1988) ArticleTitleCorneal decompensation after argon laser iridectomy Arch Ophthalmol 106 1572–1574 Occurrence Handle1:STN:280:DyaL1M%2FkvFOksg%3D%3D Occurrence Handle10.1001/archopht.1988.01060140740047 Occurrence Handle3056353
RW Zabel IM MacDonald G Mintsioulis (1991) ArticleTitleCorneal endothelial decompensation after argon laser iridotomy Can J Ophthalmol 26 367–373 Occurrence Handle1:STN:280:DyaK387htVKqsw%3D%3D Occurrence Handle1764642
KR Wilhelmus (1992) ArticleTitleCorneal edema following argon laser iridotomy Ophthalmic Surg 23 533–537 Occurrence Handle1:STN:280:DyaK38znsFCmug%3D%3D Occurrence Handle1508483
S Jeng JS Lee SC Huang (1991) ArticleTitleCorneal decompensation after argon laser iridectomy—a delayed complication Ophthalmic Surg 22 565–569 Occurrence Handle1:STN:280:DyaK38%2FntlWhsQ%3D%3D Occurrence Handle1961612
H De Guillebon R Pfister J Govignon O Pomerantzeff CL Schepens (1971) ArticleTitleCorneal temperature measurements during retinal photocoagulation Arch Ophthalmol 85 712–717 Occurrence Handle1:STN:280:DyaE3M3mtlSntA%3D%3D Occurrence Handle10.1001/archopht.1971.00990050714012 Occurrence Handle5562793
RL Cooper IJ Constable (1981) ArticleTitlePrevention of corneal burns during high-energy laser iridotomy Am J Ophthalmol 91 534–535 Occurrence Handle1:STN:280:DyaL3M7otFeguw%3D%3D Occurrence Handle10.1016/0002-9394(81)90246-4 Occurrence Handle7194582
KS Lim SS Wickremasinghe MF Cordeiro C Bunce PT Khaw (2005) ArticleTitleAccuracy of intraocular pressure measurements in New Zealand white rabbits Invest Ophthalmol Vis Sci 46 2419–2423 Occurrence Handle10.1167/iovs.04-0610 Occurrence Handle15980230
M Yamaguchi M Imai H Ohashi Y Hosokawa H Tachibana H Ito (1986) ArticleTitleEnhanced myocardial protection by systemic deep hypothermia In children undergoing total correction of tetralogy of Fallot Ann Thorac Surg 41 639–646 Occurrence Handle1:STN:280:DyaL283ktFSlsQ%3D%3D Occurrence Handle10.1016/S0003-4975(10)63079-2 Occurrence Handle3718042
DS Prough DA Stump RC Roy et al. (1986) ArticleTitleResponse of cerebral blood flow to changes in carbon dioxide tension during hypothermic cardiopulmonary bypass Anesthesiology 64 576–581 Occurrence Handle1:STN:280:DyaL287osVOitw%3D%3D Occurrence Handle10.1097/00000542-198605000-00006 Occurrence Handle3083727
M Matsumoto Y Iida T Sakabe T Sano T Ishikawa K Nakakimura (1997) ArticleTitleMild and moderate hypothermia provide better protection than a burst-suppression dose of thiopental against ischemic spinal cord injury in rabbits Anesthesiology 86 1120–1127 Occurrence Handle1:CAS:528:DyaK2sXjs1Shuro%3D Occurrence Handle10.1097/00000542-199705000-00016 Occurrence Handle9158362
J Rinkoff R Machemer T Hida D Chandler (1986) ArticleTitleTemperature-dependent light damage to the retina Am J Ophthalmol 102 452–462 Occurrence Handle1:STN:280:DyaL2s%2Fitlahuw%3D%3D Occurrence Handle10.1016/0002-9394(86)90073-5 Occurrence Handle3766660
NM Jabbour CL Schepens SM Buzney (1988) ArticleTitleLocal ocular hypothermia in experimental intraocular surgery Ophthalmology 95 1687–1690 Occurrence Handle1:STN:280:DyaL1M7lvF2gtg%3D%3D Occurrence Handle10.1016/S0161-6420(88)32956-8 Occurrence Handle3231436
JD Zilis D Chandler R Machemer (1990) ArticleTitleClinical and histologic effects of extreme intraocular hypothermia Am J Ophthalmol 109 469–473 Occurrence Handle1:STN:280:DyaK3c3jtlartA%3D%3D Occurrence Handle10.1016/S0002-9394(14)74615-X Occurrence Handle2330950
K Tamai E Toumoto A Majima (1997) ArticleTitleLocal hypothermia protects the retina from ischaemic injury in vitrectomy Br J Ophthalmol 81 789–794 Occurrence Handle1:STN:280:DyaK1c%2FosV2itg%3D%3D Occurrence Handle10.1136/bjo.81.9.789 Occurrence Handle9422935 Occurrence Handle1722302
K Tamai A Majima C Yan (1995) ArticleTitleEffects of local hypothermia on uveal blood flow and postoperative inflammation in vitrectomy Jpn J Ophthalmol 39 43–48 Occurrence Handle1:STN:280:DyaK2MzntFaitw%3D%3D Occurrence Handle7643482
Y Kitazawa E Maekawa S Sasaki T Tokoro M Mochizuki S Ito (1999) ArticleTitleCooling effect on excimer laser photorefractive keratectomy J Cataract Refract Surg 25 1349–1355 Occurrence Handle1:STN:280:DyaK1MvktFKjtw%3D%3D Occurrence Handle10.1016/S0886-3350(99)00207-2 Occurrence Handle10511934
HA Stein AG Salim RM Stein A Cheskes (1999) ArticleTitleCorneal cooling and rehydration during photorefractive keratectomy to reduce postoperative corneal haze J Refract Surg 15 IssueIDSuppl 2 S232–S233 Occurrence Handle1:STN:280:DyaK1M3hvFSmtg%3D%3D Occurrence Handle10202729
H Fujishima Y Yagi I Toda J Shimazaki K Tsubota (1995) ArticleTitleIncreased comfort and decreased inflammation of the eye by cooling after cataract surgery Am J Ophthalmol 119 301–306 Occurrence Handle1:STN:280:DyaK2M7nslagug%3D%3D Occurrence Handle10.1016/S0002-9394(14)71171-7 Occurrence Handle7872390
Author information
Authors and Affiliations
Corresponding author
About this article
Cite this article
Kataoka, T., Zako, M., Takeyama, M. et al. Cooling Prevents Induction of Corneal Damage by Argon Laser Peripheral Iridotomy. Jpn J Ophthalmol 51, 317–324 (2007). https://doi.org/10.1007/s10384-007-0460-3
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s10384-007-0460-3