Skip to main content

Antisepsis of the Skin by Treatment with Tissue-Tolerable Plasma (TTP): Risk Assessment and Perspectives

  • Conference paper
  • First Online:
Plasma for Bio-Decontamination, Medicine and Food Security

Abstract

The application of tissue tolerable plasma (TTP) is well suited for ­disinfection of living tissue. In particular, when treating chronic wounds, it has ­several advantages in comparison to the classical application of antiseptics, which do not penetrate sufficiently into the tissue or inhibit wound regeneration. The mode of action of the plasma is mainly based on synergetic effects between temperature increase and the formation of free radicals, which destroy the bacteria and fungi.

In the present paper a risk assessment of TTP in dermatology is given. The investigations have been carried out with an atmospheric pressure plasma-jet working with Argon as a discharge medium. It was found that during the plasma treatment of tissue, the antioxidative potential is reduced only in the upper part of the stratum corneum, but not in deeper cell layers. Selecting the optimum parameters of the plasma formation, the UV exposure of the skin is less than in the case of UV irradiation of the sun on a summer day at noon.

If the duration of the plasma treatment of the skin is in the optimal range for wound healing, no thermal damage has to be expected.

Additionally, it could be demonstrated that plasma is able to reach the follicular reservoir for antisepsis where germs are located.

This is a preview of subscription content, log in via an institution to check access.

Access this chapter

Chapter
USD 29.95
Price excludes VAT (USA)
  • Available as PDF
  • Read on any device
  • Instant download
  • Own it forever
eBook
USD 169.00
Price excludes VAT (USA)
  • Available as EPUB and PDF
  • Read on any device
  • Instant download
  • Own it forever
Softcover Book
USD 219.99
Price excludes VAT (USA)
  • Compact, lightweight edition
  • Dispatched in 3 to 5 business days
  • Free shipping worldwide - see info
Hardcover Book
USD 219.99
Price excludes VAT (USA)
  • Durable hardcover edition
  • Dispatched in 3 to 5 business days
  • Free shipping worldwide - see info

Tax calculation will be finalised at checkout

Purchases are for personal use only

Institutional subscriptions

References

  1. Phillips TJ (2001) Current approaches to venous ulcers and compression. Dermatol Surg 27(7):611–621

    Article  Google Scholar 

  2. Sekkat N, Kalia YN, Guy RH (2002) Biophysical study of porcine ear skin in vitro and its comparison to human skin in vivo. J Pharm Sci 91(11):2376–2381

    Article  Google Scholar 

  3. von Woedtke T, Kramer A, Weltmann KD (2008) Plasma sterilization: what are the conditions to meet this claim? Plasma Process Polym 5(6):534–539

    Article  Google Scholar 

  4. Weltmann KD, Kindel E, Brandenburg R, Meyer C, Bussiahn R, Wilke C, von Woedtke T (2009) Atmospheric pressure plasma jet for medical therapy: plasma parameters and risk estimation. Contrib Plasma Phys 49(9):631–640

    Article  ADS  Google Scholar 

  5. Weltmann KD, Kindel E, von Woedtke T, Hahnel M, Stieber M, Brandenburg R (2010) Atmospheric-pressure plasma sources: prospective tools for plasma medicine. Pure Appl Chem 82(6):1223–1237

    Article  Google Scholar 

  6. Zastrow L, Groth N, Klein F, Kockott D, Lademann J, Renneberg R, Ferrero L (2009) The missing link–light-induced (280–1,600 nm) free radical formation in human skin. Skin Pharmacol Physiol 22(1):31–44

    Article  Google Scholar 

  7. Lademann J, Richter H, Alborova A, Humme D, Patzelt A, Kramer A, Weltmann KD, Hartmann B, Ottomann C, Fluhr JW, Hinz P, Hubner G, Lademann O (2009) Risk assessment of the application of a plasma jet in dermatology. J Biomed Opt 14(5):054025

    Article  Google Scholar 

  8. Lademann O, Richter H, Patzelt A, Alborova A, Humme D, Weltmann KD, Hartmann B, Hinz P, Kramer A, Koch S (2010) Application of a plasma-jet for skin antisepsis: analysis of the thermal action of the plasma by laser scanning microscopy. Laser Phys Lett 7(6):458–462

    Article  ADS  Google Scholar 

  9. Hammann A, Huebner NO, Bender C, Ekkernkamp A, Hartmann B, Hinz P, Kindel E, Koban I, Koch S, Kohlmann T, Lademann J, Matthes R, Muller G, Titze R, Weltmann KD, Kramer A (2010) Antiseptic efficacy and tolerance of tissue-tolerable plasma compared with two wound antiseptics on artificially bacterially contaminated eyes from commercially slaughtered pigs. Skin Pharmacol Physiol 23(6):328–332

    Article  Google Scholar 

  10. Lademann O (2011) Antisepsis of the follicular reservoir by treatment with tissue-tolerable plasma (TTP). Laser Phys Lett 1-5. doi:10.1002/lapl.201010123

    Google Scholar 

  11. Lange-Asschenfeldt B, Alborova A, Kruger-Corcoran D, Patzelt A, Richter H, Sterry W, Kramer A, Stockfleth E, Lademann J (2009) Effects of a topically applied wound ointment on epidermal wound healing studied by in vivo fluorescence laser scanning microscopy analysis. J Biomed Opt 14(5):054001

    Article  Google Scholar 

  12. Meyer W, Schwarz R, Neurand K (1978) The skin of domestic mammals as a model for the human skin, with special reference to the domestic pig. Curr Probl Dermatol 7:39–52

    Google Scholar 

  13. Foest R, Kindel E, Ohl A, Stieber M, Weltmann KD (2005) Non-thermal atmospheric pressure discharges for surface modification. Plasma Phys Control Fusion 47:B525–B536

    Article  Google Scholar 

  14. Weltmann KD, Brandenburg R, von Woedtke T, Ehlbeck J, Foest R, Stieber M, Kindel E (2008) Antimicrobial treatment of heat sensitive products by miniaturized atmospheric pressure plasma jets (APPJs). J Phys D Appl Phys 41(19):194008

    Google Scholar 

  15. Lademann J, Jacobi U, Surber C, Weigmann HJ, Fluhr JW (2009) The tape stripping procedure – evaluation of some critical parameters. Eur J Pharm Biopharm 72(2):317–323

    Article  Google Scholar 

  16. Williamson PT, Roth JF, Haddingham T, Watts A (2000) Expression and purification of recombinant neurotensin in Escherichia coli. Protein Exp Purif 19(2):271–275

    Article  Google Scholar 

  17. Martschick A, Teichmann A, Richter H, Schanzer S, Antoniou C, Sterry W, Lademann J (2007) Analysis of the penetration profiles of topically applied substances by laser scanning microscopy. Laser Phys Lett 4(5):395–398

    Article  ADS  Google Scholar 

  18. Kandarova H, Richter H, Liebsch M, Lademann J (2007) Stratum corneum architecture of reconstructed human skin models monitored by fluorescent confocal laser scanning microscopy. Laser Phys Lett 4(4):308–311

    Article  ADS  Google Scholar 

  19. Lademann J, Jacobi U, Richter H, Otberg N, Weigmann HJ, Meffert H, Schaefer H, Blume-Peytavi U, Sterry W (2004) In vivo determination of UV-photons entering into human skin. Laser Phys 14(2):234–237

    Google Scholar 

  20. Fluhr JW, Sassning S, Lademann O, Darvin ME, Schanzer S, Kramer A, Richter H, Sterry W, Lademann J (2011) The influence of skin treatment with tissue tolerable plasma on the antioxidant concentration in human skin. Experimental Dermatology. DOI: 10.1111/j.1600-0625.2011.01411.x

    Google Scholar 

  21. Lademann O, Kramer A, Richter H, Patzelt A, Meinke MC, Röwert-Huber J, Czaika V, Weltmann K-D, Hartmann B, Koch S (2011) Antisepsis of the follicular reservoir by treatment with tissue-tolerable plasma (TTP). Laser Phys Lett 8(4):313–317

    Article  Google Scholar 

Download references

Acknowledgments

This work was realized within the framework of the multi-disciplinary research cooperation of “Campus PlasmaMed”, particularly within the project “PlasmaWound”. The authors acknowledge that this work was supported by a grant funded by the German Ministry of Education and Research (BMBF, Grant No, 13 N9779).

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Jürgen Lademann .

Editor information

Editors and Affiliations

Rights and permissions

Reprints and permissions

Copyright information

© 2012 Springer Science+Business Media B.V.

About this paper

Cite this paper

Lademann, J. et al. (2012). Antisepsis of the Skin by Treatment with Tissue-Tolerable Plasma (TTP): Risk Assessment and Perspectives. In: Machala, Z., Hensel, K., Akishev, Y. (eds) Plasma for Bio-Decontamination, Medicine and Food Security. NATO Science for Peace and Security Series A: Chemistry and Biology. Springer, Dordrecht. https://doi.org/10.1007/978-94-007-2852-3_21

Download citation

Publish with us

Policies and ethics