Abstract
Pacemaker malfunction can present in several ways including failure of output, pacing failure, undersensing, oversensing, pacing at a slower or faster rate than the preset rate, pacemaker-mediated tachycardia, pacemaker syndrome, cross talk, or displacement of lead. Analysis of surface electrocardiogram is very useful in the identification and differential diagnosis of various malfunction. Proper analysis of the size, direction, and rate of the pacing spike and its relation with the intrinsic QRS and paced QRS can easily identify and differentiate various pacemaker malfunctions. Electrocardiogram can also help in identifying concomitant myocardial infarction. All these issues are discussed with representative electrocardiograms. Summary and MCQs at the end of the chapter help in rapid revision and self-assessment.
Access this chapter
Tax calculation will be finalised at checkout
Purchases are for personal use only
References
Vijayraman P, Ellenbogen KA. Bradyarrhythmias and pacemakers. In: Fuster V, Walsh RA, Harrington RA, editors. Hurst’s The Heart. New York: Mc Graw Hill; 2011. p. 1025–57.
Olshansky B, Chang MK, Pogowizd SM, Goldschlager N. Cardiac pacing and pacemaker rhythms. In: Olshansky B, Chung MK, Pogowizd SM, Goldschlager N, editors. Arrhythmia Essentials. Burlington: Jones Bartlett Learning; 2012. p. 240–60.
Swerdlow CD, Wang PJ, Zipes DP. Pacemakers and implantable cardioverter-defibrillators. In: Mann DP, Zipes DP, Libby P, Bonow RO, editors. Braunwald’s Heart Disease. Philadelphia: Elsevier; 2015. p. 721–42.
Haisty WK, Lim TH, Wagner GS. Artificial cardiac pacemakers. In: Wagner GS, editor. Marriot’s Practical Electrocardiography. Philadelphia: Wolters Kluwer; 2001. p. 423–44.
Singh R, Mittal SR, Mathur D. Status of surface electrocardiography in evaluating pacemaker malfunctions. Indian Journal of Electrocardiology. 1996;2:7–10.
Sethi KK, Rajagopal S. Diagnosis of pacemaker malfunction. Indian Journal of Electrocardiology. 1989;1:4–8.
Author information
Authors and Affiliations
Appendices
Summary
-
Hysteresis. This phenomenon is seen in on-demand intermittent pacing. Following an on-demand paced beat, the next paced beat is delayed by few milliseconds, while the pacemaker waits for the next spontaneous cardiac impulse to come. This interval is thus slightly longer than the interval set for regular pacing. This is a normal function and not a pacemaker malfunction.
-
Electromagnetic interference. Any electrical stimulation in the body can change the pacemaker function to VVO mode. This can happen during electrocautery, lithotripsy, diathermy, radio-frequency catheter ablation, transcutaneous electric nerve stimulation, or spinal cord stimulation [5]. This is a normal protection against failure of pacing due to oversensing.
-
Pacing failure. When there is no pacing stimulus in spite of the absence of an expected spontaneous P or QRS, it is called pacing failure. This can be due to oversensing of a noncardiac electric potential, tall T or P waves, lead connection failure, or lead fracture. Magnet application changes pacing to VOO mode with appearance of regular pacing spikes. Magnet application does not restore pacing spikes in the setting of device malfunction.
-
Failure to capture. The presence of a pacemaker stimulus (pacing spike) which is not followed by a paced P wave or QRS complex is known as failure to capture. It is usually due to lead displacement or elevated pacing threshold at the electrode-myocardial interface. Elevation of threshold can be due to fibrosis, ischemia, or infarction around the tip of the pacing lead. Loose screw, insulation break, lead fracture, or battery depletion can also result in failure to capture. Combination of failure to pace and capture usually indicates a pacing system failure.
-
Cross talk. Oversensing of the atrial pacing stimulus by the ventricular lead is known as “cross talk.” It results in inhibition of ventricular pacing.
-
Battery depletion. Automatic decrease in the pacing rate or change of pacemaker function to VOO mode suggests battery depletion.
1.1 Differentiation of Pacemaker Malfunction
-
(A)
Pacemaker spike
-
(1).
Absent pacemaker spikes
-
(a)
QRS faster than the preset pacing rate:
-
(a)
It happens if patient’s intrinsic heart rate is faster than the pacemaker rate (Figs. 15.20 and 15.21). Applying magnet over the pulse generator converts the pacemaker to asynchronous mode. Appearance of regular spike with intermittent capture suggests normal function (Figs. 15.20 and 15.21).
-
(b)
QRS slower than the preset pacing rate:
-
Dry pocket (fibrosis around the pulse generator), battery exhaustion, or lead fracture can reduce the pacing rate.
-
-
(2).
Intermittent absence of pacemaker spikes
-
Lead fracture—fractured ends make intermittent contact. It is not corrected by a magnet.
-
Oversensing of myopotentials or extraneous electric signals.
It is corrected by use of magnet.
-
-
(3).
Regular pacing spike
-
(a)
Rate
-
Rate slower than the preset rate
-
Intermittent—hysteresis
-
Regular—battery depletion, sensing of the T wave, sensing of an echo beat
-
-
Rate faster than the preset rate—pacemaker-mediated tachycardia (PMT)
-
-
(b)
Amplitude – Large—unipolar unit
-
Small—bipolar unit
-
Smaller than at the time of implantation—lead fracture resulting in high lead impedance
-
-
(c)
Duration – Broad fusions with intrinsic QRS and direct current leakage into the heart due to a short circuit in output circuit
-
(d)
Axis – Negative in leads II, III, and aVF—pacing from the right ventricular apex
-
Positive in leads II, III, and aVF—pacing from the right ventricular outflow
-
Change from previous record—displacement
-
-
(e)
Changing direction and amplitude (Figs. 15.22 and 15.23). Insulation fracture
-
(a)
-
(4).
Delayed spike after a sensed QRS—hysteresis
-
(1).
-
(B)
Distance between the pacing spike and the paced QRS
-
(C)
QRS complex
-
(a)
Configuration. Left bundle branch block configuration—right ventricular pacing
Right bundle branch block configuration—left ventricular pacing, biventricular pacing
Recent change in QRS configuration—displacement
-
(b)
Axis. Left upper quadrant—pacing from right ventricular apex
Inferior—pacing from right ventricular outflow
-
(c)
Fusion or pseudo fusion beats do not indicate sensing failure.
-
(a)
-
(D)
Sensing failure [ 6 ]
-
(a)
Ventricular ectopics arising at a distance from the tip of the lead
-
(b)
Low amplitude of QRS of intracardiac electrogram due to infarction, inflammation, or fibrosis
-
(c)
Lead displacement, even microdisplacement
-
(d)
Myocardial perforation
-
(e)
Lead fracture
-
(f)
Fault in pulse generator
-
(a)
-
(E)
Failure to capture (pacemaker spike without subsequent QRS)
-
(a)
Raised threshold—Hypokalemia, hypocalcemia, flecainide
Ischemia, infarction, fibrosis, inflammation
-
(b)
Lead displacement
-
(c)
Myocardial perforation
-
(a)
MCQs
Q1. Change in the axis of the pacemaker spike as compared to previous ECG suggests:
-
(a)
Insulation break
-
(b)
Lead fracture
-
(c)
Lead displacement
-
(d)
All
Q2. Increase in the distance between the pacing spike and onset of the paced QRS as compared to previous ECG could be due to:
-
(a)
Ischemia
-
(b)
Infarction
-
(c)
Fibrosis
-
(d)
All
Q3. Change in the QRS configuration from LBBB to RBBB pattern suggests:
-
(a)
Lead displacement into right ventricular outflow tract
-
(b)
Lead fracture
-
(c)
Perforation of interventricular septum
-
(d)
Insulation break
Q4. Failure to sense ventricular ectopics with normal sensing of intrinsic cardiac impulse needs:
-
(a)
Replacement of pulse generator
-
(b)
Change of lead
-
(c)
Programming of pulse generator
-
(d)
None
Q5. Presence of a pacemaker spike with failure of ventricular depolarization can be due to:
-
(a)
Hypokalemia
-
(b)
Myocardial perforation
-
(c)
Lead fracture
-
(d)
Battery exhaustion
Q6. Lead impedance less than 300 ohms suggests:
-
(a)
Insulation break
-
(b)
Lead fracture
-
(c)
Microdisplacement
-
(d)
Myocardial perforation
Q7. Lead impedance more than 1200 ohms suggests:
-
(a)
Insulation break
-
(b)
Lead fracture
-
(c)
Poor electrode location
-
(d)
Macro displacement
Q8. Lead impedance is not increased by:
-
(a)
Lead fracture
-
(b)
Poor connection of the lead to the pulse generator
-
(c)
Loose set screws
-
(d)
Lead displacement
Q9. Diaphragmatic stimulation suggests:
-
(a)
Oversensing
-
(b)
Insulation rupture
-
(c)
Gross change in lead position
-
(d)
Myocardial perforation
Q10. Intermittent loss of pacing/sensing suggests:
-
(a)
Lead displacement
-
(b)
Myocardial perforation
-
(c)
Lead fracture
-
(d)
Pseudomalfunction
Q11. Delayed spike after sensing of a normal QRS suggests:
-
(a)
Hysteresis
-
(b)
Sensing malfunction
-
(c)
Pacing malfunction
-
(d)
All
Q12. Asystole produced by exercising muscles close to the implant site suggests:
-
(a)
Sensing failure
-
(b)
Pacing failure
-
(c)
Oversensing
-
(d)
All
Q13. Which cause of low-voltage QRS causes undersensing?
-
(a)
Myocardial infarction
-
(b)
Pericardial effusion
-
(c)
Emphysema
-
(d)
Obesity
Q14. Amplitude of intracardiac R wave signal is not reduced in:
-
(a)
Lead fracture
-
(b)
Microdisplacement
-
(c)
Myocardial perforation
-
(d)
Emphysema
Q15. Fusion beat suggests:
-
(a)
Sensing failure
-
(b)
Pacing failure
-
(c)
Oversensing
-
(d)
Nil
Q16. Regular bradycardia in a patient with permanent pacemaker is not due to:
-
(a)
T wave sensing
-
(b)
Battery depletion
-
(c)
Pacing failure
-
(d)
Myopotential inhibition
Q17. Pacemaker-mediated tachycardia can occur in:
-
(a)
VVI pacing
-
(b)
DDD pacing
-
(c)
Biventricular pacing
-
(d)
All
Q18. Marked variation in spike amplitude can occur in:
-
(a)
Lead fracture
-
(b)
Insulation break
-
(c)
Oversensing of extracorporeal signals
-
(d)
Battery depletion
Q19. Tachycardia in a patient with a DDD pacemaker can be due to:
-
(a)
Atrial tachycardia
-
(b)
Ventricular tachycardia
-
(c)
Pacemaker-mediated tachycardia
-
(d)
All
Q20. Deterioration in cardiac status after VVI pacing suggests:
-
(a)
Pacing failure
-
(b)
Pacemaker syndrome
-
(c)
Myocardial ischemia
-
(d)
Unmasking of cardiac failure
Q21. Cannon waves in a patient with implanted pacemaker suggest which mode of pacing:
-
(a)
AAI
-
(b)
VVI
-
(c)
DDD
-
(d)
Biventricular
Q22. Pacemaker stimulus falling on the T wave of an intrinsic QRS suggests:
-
(a)
Sensing failure
-
(b)
Pacing failure
-
(c)
Failure to capture
-
(d)
None
Q23. Failure of appearance of a pacing stimulus in spite of a long pause suggests:
-
(a)
Sensing failure
-
(b)
Pacing Failure
-
(c)
Failure of capture
-
(d)
Hysteresis
Q24. Pacemaker stimulus not followed by a paced QRS suggests:
-
(a)
Sensing failure
-
(b)
Pacing failure
-
(c)
Failure of capture
-
(d)
Electromagnetic interference
Q25. Atrial pacing stimulus followed by a P wave but not followed by ventricular pacing spike and paced QRS suggests:
-
(a)
Hysteresis
-
(b)
Failure of capture
-
(c)
Electromagnetic interference
-
(d)
Cross talk
Q26. Hysteresis suggests:
-
(a)
Failure of capture
-
(b)
Failure to pace
-
(c)
Normal function
-
(d)
Failure to sense
Q27. Automatic decrease in the pacing rate of a pacemaker suggests:
-
(a)
Hysteresis
-
(b)
Electromagnetic interference
-
(c)
Failure to sense
-
(d)
Battery depletion
Q28. Battery depletion can produce:
-
(a)
Failure to sense
-
(b)
Failure to pace
-
(c)
Change of VVI to VOO mode
-
(d)
Hysteresis
Q29. Pacemaker syndrome can occur in:
-
(a)
VVI pacing
-
(b)
DDD pacing
-
(c)
DDDR pacing
-
(d)
All
Q30. Pathology at the tissue-lead interface can produce:
-
(a)
Failure to sense
-
(b)
Failure to capture
-
(c)
Failure to pace
-
(d)
All
Q31. Lead fracture can cause:
-
(a)
Hysteresis
-
(b)
Cross talk
-
(c)
Electromagnetic interference
-
(d)
Sensing failure
Q32. Insulation break can cause:
-
(a)
Sensing failure
-
(b)
Pacing Failure
-
(c)
Failure of capture
-
(d)
Hysteresis
Q33. Magnet application over pulse generator changes:
-
(a)
VVI to VOO mode
-
(b)
VOO mode to VVI mode
-
(c)
VVI to DDD mode
-
(d)
DDD to VVI mode
Q34. Hysteresis is suggested when a paced beat occurs:
-
(a)
Before its scheduled time
-
(b)
At its scheduled time
-
(c)
Slightly later than the scheduled time
-
(d)
Much after the scheduled time
Q35. Electromagnetic interference changes:
-
(a)
VOOO to VVI mode
-
(b)
VVI to VOO mode
-
(c)
DDD to VVI mode
-
(d)
None
Answers
(1) c; (2) d; (3) c; (4) d; (5) a, b; (6) a; (7) b; (8) d; (9) d; (10) c; (11) a; (12) c; (13) a; (14) d; (15) d; (16) d; (17) a; (18) a; (19) d; (20) b; (21) b; (22) a; (23) b; (24) c; (25) d; (26) c; (27) d; (28) c; (29) a; (30) a, b; (31) d; (32) a, c; (33) a; (34) c; (35) b
Rights and permissions
Copyright information
© 2023 The Author(s), under exclusive license to Springer Nature Singapore Pte Ltd.
About this chapter
Cite this chapter
Mittal, S. (2023). Pacemaker Malfunctions. In: Insights into Electrocardiograms with MCQs. Springer, Singapore. https://doi.org/10.1007/978-981-99-0127-2_15
Download citation
DOI: https://doi.org/10.1007/978-981-99-0127-2_15
Published:
Publisher Name: Springer, Singapore
Print ISBN: 978-981-99-0126-5
Online ISBN: 978-981-99-0127-2
eBook Packages: MedicineMedicine (R0)