Abstract
This chapter provides an overview of the safety aspects of application of ultrasound in medicine. It starts with the short history of ultrasound methods and devices as well as basic principles of ultrasound imaging systems. The application of ultrasound in medicine greatly evolved and nowadays it can be divided into two main areas: imaging and therapy. In order to assure a safe and responsible application of ultrasound in medicine one should be aware of physical processes which can be produced in tissue by ultrasound such as temperature rise, cavitation and acoustic streaming. The importance of understanding how these processes can affect the human cell is self-explanatory. In order to better understand the guidelines for testing and quality control of ultrasonic devices it is necessary to give an overview of basic output parameters. Only the most important parameters from the point of safe use of ultrasound are described, e.g. acoustic pressure, acoustic power and intensity. In order to protect the public against inappropriate exposure when ultrasound is used for medical applications, international standards and national regulations are developed. Diagnostic ultrasound imaging is very often the basis for diagnostic decision; therefore it is also necessary to include such systems into a comprehensive quality assurance programme. Ultrasound systems used for therapy have larger intensities though there are additional safety requirements compared to diagnostic systems. The ultrasound intensity, effective radiation area and beam non-uniformity ratio and are parameters which should be monitored.
Access this chapter
Tax calculation will be finalised at checkout
Purchases are for personal use only
Similar content being viewed by others
References
Manbachi A, Cobbold RSC (2011) Development and application of piezoelectric materials for ultrasound generation and detection. Ultrasound 19(4):187
Kremkau FW (2006) Diagnostic ultrasound. Principles and instruments, Chap. 4, 7th edn. Saunders, Philadelphia, USA
Macovski A (1983) Medical imaging systems, Chap. 10. Prentice-Hall, Englewood Cliffs, NJ
IEC 60601 part 2–5, Medical electrical equipment: particular requirements for the safety of ultrasound physiotherapy equipment. International Electrotechnical Commission, Geneva
IEC 60601 part 2–37: Medical electrical equipment: particular requirements for the safety of ultrasound diagnostic and monitoring equipment, 2001 consolidated with amendment 1, 2004. International Electrotechnical Commission, Geneva
IEC 60601 part 2–62 (2013) Medical electrical equipment: particular requirements for the basic safety and essential performance of high intensity therapeutic ultrasound (HITU) equipment. International Electrotechnical Commission, Geneva
Perkins MA (1989) A versatile force balance for ultrasound power measurement. Phys Med Biol 34(11):1645
Farmery MJ, Whittingham TA (1978) A portable radiation-force balance for use with diagnostic ultrasonic equipment. Ultrasound Med Biol 3(4):373–379
Carson PL, Fischella PR, Oughton TV (1978) Ultrasonic power and intensities produced by diagnostic ultrasound equipment. Ultrasound Med Biol 3(4):341–350
Farmery MJ, Whittingham TA (1978) A portable radiation-force balance for use with diagnostic ultrasonic equipment. Ultrasound Med Biol 3:373–379
Duck FA (1989) Output data from European studies. Ultrasound Med Biol 15(Suppl 1):61–64
Duck FA, Martin K (1991) Trends in diagnostic ultrasound exposure. Phys Med Biol 36(11):1423–1432
Henderson J, Willson K, Jago JR, Whittingham TA (1995) A survey of the acoustic outputs of diagnostic ultrasound equipment in current clinical use. Ultrasound Med Biol 21(5):699–705
Whittingham TA (2000) The acoustic output of diagnostic machines. In: ter Haar G, Duck FA (eds) The safe use of ultrasound in medicine diagnosis, 2nd edn. BMUS/BIR, London, UK
FDA (1997) Information for manufacturers seeking marketing clearance of diagnostic ultrasound systems and transducers. Division of Reproductive, Abdominal, Ear, Nose, Throat and Radiological Devices: Office of Device Evaluation. US Department of Health and Human Services: Food and Drug Administration
AIUM/NEMA (1992) UD2: Acoustic output measurement standard for diagonstic equipement (Rockville, MD: American Institute of Ultrasound in Medicine)
Bigelow TA, Church CC, Sandstrom K, Abbott JG, Ziskin MC, Edmonds PD, Herman B, Thomenius KE, Teo TJ (2011) The thermal index: its strengths, weaknesses, and proposed improvements. J Ultrasound Med 30(5):714–734
Apfel RE, Holland CK (1991) Gauging the likelihood of cavitation from short-pulse, low-duty cycle diagnostic ultrasound. Ultrasound Med Biol 17(2):179–185
Curch CC (2002) Spontaneous homogeneous nucleation, inertial cavitation and the safety of diagnostic ultrasound. Ultrasound Med Biol 28(10):1349–1364
Duck FA, Starritt HC, ter Haar GR, Lunt MJ (1989) Surface heating of diagnostic ultrasound transducers. Br J Radiol 62:1005–1013
BMUS (2010) Guidelines for the safe use of diagnostic ultrasound equipment. Ultrasound 18:52–59
Calvert J, Duck F, Clift S, Azaime H (2007) Surface heating by transvaginal transducers. Ultrasound Obstet Gynecol 29:427–432
Quality assurance in diagnostic radiology: a guide prepared following a workshop held in Neuherberg, Federal Republic of Germany, 20–24 October 1980, and organized jointly by Institute of Radiation Hygiene, Federal Health Office, Neuherberg, Federal Republic of Germany, Society for Radiation and Environmental Research, Neuherberg, Federal Republic of Germany and World Health Organization, Geneva, Switzerland, 1982
Hangiandreou NJ, Stekel SF, Tradup DJ, Gorny KR, King DM (2011) Four-year experience with a clinical ultrasound quality control program. Ultrasound Med Biol 37(8):1350
American College of Radiology (ACR) (2011) ACR technical standard for diagnostic medical physics performance monitoring of real time ultrasound equipment. ACR, Reston, VA
American Institute of Ultrasound in Medicine (AIUM) (2008) Routine quality assurance for diagnostic ultrasound equipment. AIUM, Laurel, MD
Goodsitt MM, Carson PL, Witt S, Hykes DL, Kofler JM (1998) Real-time B-mode ultrasound quality control test procedures. Report of AAPM Ultrasound Task Group No. 1. Med Phys 25:1385–1406
International Electrotechnical Commission, IEC 62736-1 (2011) Ultrasonics—quality control of diagnostic medical ultrasound systems
Institute of Physics and Engineering in Medicine (IPEM) (2010) Quality assurance of ultrasound imaging systems, Report No 102, IPEM, York
Kollmann C, Dekorte C, Dudley NJ, Gritzmann N, Martin K, Evans DH (2012) Guideline for technical quality assurance (TQA) of ultrasoud devices (B-Mode)—Version 1.0 (July 2012). EFSUMB Technical Quality Assurance Group. Ultraschall Med 33:544–549
Dudley N, Russell S, Ward B, Hoskins P (2014) The BMUS guidelines for regular quality assurance testing of ultrasound scanners. Ultrasound 22:6–7
Samei E, Badano A, Chakraborty D, Compton K, Cornelius C, Corrigan K, Flynn MJ, Hemminger B, Hangiandreou N, Johnson J, Moxley-Stevens DM, Pavlicek W, Roehrig H, Rutz L, Shepard J, Uzenoff RA, Wang J, Willis CE (2005) Assessment of display performance for medical imaging systems: executive summary of AAPM TG18 report. Med Phys 32(4):1205–1225
Shaw A, Hodnett M (2008) Calibration and measurement issues for therapeutic ultrasound. Ultrasonics 48:234–252
Gutiérrez MI, Calás H, Ramos A, Vera A, Leija L (2012) Acoustic field modeling for physiotherapy ultrasound applicators by using approximated functions of measured non-uniform radiation distributions. Ultrasonics 52(6):767–777
IEC 61689 (2007) Ultrasonics—physiotherapy systems—field specifications and methods of measurement in the frequency range 0.5 MHz to 5 MHz
Hekkenberg RT, Oosterbaan WA, van Beekum WT (1986) Evaluation of ultrasound therapy devices. Physiotherapy 72:390–395
Pye SD, Milford C (1994) The performance of ultrasound physiotherapy machines in Lothian region, Scotland. Ultrasound Med Biol 20(4):347–359
Pye SD (1996) Ultrasound therapy equipment—does it perform? Physiotherapy 82(1):39–44
Duck F (2009) Acoustic dose and acoustic dose rate. Ultrasound Med Biol 35(10):1679–1685
Miller D, Smith N (2012) Overview of therapeutic ultrasound applications and safety considerations. J Ultrasound Med 31(4):623–634
Pye S, Hildersley K, Somer E, Munro V (1994) A simple calorimeter for monitoring the output power of ultrasound therapy machines. Physiotherapy 80(4):219–223
Zeqiri B, Barrie J (2008) Evaluation of a novel solid-state method for determining the acoustic power generated by physiotherapy ultrasound transducers. Ultrasound Med Biol 34:1513–1527
Zeqiri B, Gélat PN, Barrie J, Bickley CJ (2007) A novel pyroelectric method of determining ultrasonic output power: device concept, modelling and preliminary studies. IEEE Trans Ultrason Ferroelectr Freq Control 54(11):2318–2330
Zeqiri B, Žauhar G, Hodnett M, Barrie J (2011) Progress in developing a thermal method for measuring the output power of medical ultrasound transducers that exploits the pyroelectric effect. Ultrasonics 51:420–424
Zeqiri B, Žauhar G, Rajagopal S, Pounder A (2012) Systematic evaluation of a secondary method for measuring diagnostic-level medical ultrasound transducer output power based on a large-area pyroelectric sensor. Metrologia 49(3):368–381
Butterworth I, Barrie J, Zeqiri B, Žauhar G, Parisot B (2012) Exploiting thermochromic materials for the rapid quality assurance of physiotherapy ultrasound treatment heads. Ultrasound Med Biol 38(5):767–776
Žauhar G, Radojčić ĐS, Dobravac D, Jurković S (2015) Quantitative testing of physiotherapy ultrasound beam patterns within a clinical environment using a thermochromic tile. Ultrasonics 58:6–10
Author information
Authors and Affiliations
Corresponding author
Editor information
Editors and Affiliations
Rights and permissions
Copyright information
© 2018 Springer Nature Singapore Pte Ltd.
About this chapter
Cite this chapter
Žauhar, G., Diklić, A., Jurković, S. (2018). Inspection and Testing of Diagnostic Ultrasound Devices. In: Badnjević, A., Cifrek, M., Magjarević, R., Džemić, Z. (eds) Inspection of Medical Devices . Series in Biomedical Engineering. Springer, Singapore. https://doi.org/10.1007/978-981-10-6650-4_6
Download citation
DOI: https://doi.org/10.1007/978-981-10-6650-4_6
Published:
Publisher Name: Springer, Singapore
Print ISBN: 978-981-10-6649-8
Online ISBN: 978-981-10-6650-4
eBook Packages: EngineeringEngineering (R0)