Skip to main content

In Vivo Testing of a Roller-Screw Type Electric Total Artificial Heart

  • Conference paper

Abstract

Work was begun on the development of an electric motor drive total artificial heart at the Pennsylvania State University in 1978. The first systems developed utilized a low-speed high-torque brushless DC motor rotating a triple-track cam mechanism. This mechanism translated a rotary force into a rectilinear motion to actuate alternately the sac type blood pumps located on either end of the motor drive mechanism. The prototype of this system weighed slightly over 1 kg. Implantation of the cam type electric motor driven artificial heart began in 1983 and 222-day calf survival was obtained late in that year [1]. In 1984, work was begun on a roller-screw type electric motor driven artificial heart. The roller-screw device had the advantages of being smaller, lighter, and less expensive to produce. Extensive in vitro testing of the roller-screw device was performed from 1984 through 1986. In mid-1986 implantation of the roller-screw device was begun in calves. Over the next year the device was implanted in six animals with a maximum survival time of greater than 85 days.

This is a preview of subscription content, log in via an institution.

Buying options

Chapter
USD   29.95
Price excludes VAT (USA)
  • Available as PDF
  • Read on any device
  • Instant download
  • Own it forever
eBook
USD   39.99
Price excludes VAT (USA)
  • Available as EPUB and PDF
  • Read on any device
  • Instant download
  • Own it forever
Softcover Book
USD   54.99
Price excludes VAT (USA)
  • Compact, lightweight edition
  • Dispatched in 3 to 5 business days
  • Free shipping worldwide - see info

Tax calculation will be finalised at checkout

Purchases are for personal use only

Learn about institutional subscriptions

Preview

Unable to display preview. Download preview PDF.

Unable to display preview. Download preview PDF.

References

  1. Rosenberg G, Snyder A, Landis DL, Geselowitz D, Donachy JH, Pierce WS (1984) An electric motor-driven total artificial heart: seven months survival in the calf. Trans Am Soc Artif Intern Organs 30:69–74

    PubMed  CAS  Google Scholar 

  2. Rosenberg G, Snyder A, Weiss W, Landis D, Geselowitz D, Pierce W (1982) A roller screw drive for implantable blood pumps. Trans Am Soc Artif Intern Organs 28:123

    PubMed  CAS  Google Scholar 

  3. Rosenberg G, Donachy JH, Geselowitz D, Snyder A, Landis DL, Pierce WS (1982) Recent progress of the artificial heart. Eur Rev Biomed Technol 4:283–284

    Google Scholar 

  4. Richenbacher WE, Rosenberg G, Landis DL, Weiss WJ, Donachy JH, Pierce WS (1983) Development of an implantable electric motor artificial heart. Surg Forum 34:250–253

    Google Scholar 

  5. Rosenberg G, Cleary TJ, Snyder AJ, Landis DL, Gaselowitz DB, Pierce WS (1985) A totally implantable artificial heart design. Transactions of the American society of mechanical engineers, 85-WA/DE-11

    Google Scholar 

  6. Rosenberg G, Snyder A, Pierce WS, Geselowitz D (1982) Engineering development of the mechanical heart. In: McNeil BJ, Cravalho EG (eds) Critical issues in medical technology. Cravalho. Auburn House, Boston, pp 381–397

    Google Scholar 

  7. Gaines WE, Donachy JH, Rosenberg G, Landis DL, Pierce WS (1984) Studies leading to an artificial heart for clinical application. Contemp Surg 24:41–48

    Google Scholar 

  8. Jorge E, Snyder AJ, Rosenberg G, Fehr D, Geselowitz DB, Pierce WS (1986) Subcutaneous access port for the compliance chamber of the electric ventricular assist device and total artificial heart. In: Nose Y, Kjellstrand C, Ivanovich P (eds) Progress in artificial organ. Isao, Anaheim, pp 537–540

    Google Scholar 

  9. Lee S, Rosenberg G, Donachy JH, Wisman CB, Pierce WS (1984) The compliance problem: a major obstacle in the development of implantable blood pumps. Am Soc Artif Intern Organs 8(1):82–90

    Article  CAS  Google Scholar 

  10. Reid JS, Rosenberg G, Pierce WS (1985) Transmission of water through a biocompatible polyurethane - application to circulatory assist devices. J Biomed Mater Res 19(9):1181–1202

    Article  PubMed  CAS  Google Scholar 

  11. Wisman CB, Rosenberg G, Weiss WJ, Landis DL, Donachy JH, Snyder AJ, Richenbacher WE, Pierce WS (1983) The development and successful application of an intrathoracic compliance chamber for the implantable electric motor-driven ventricular assist pump. Surg Forum 34:253–256

    Google Scholar 

  12. Phillips WM, Donachy JH, Rosenberg G, Pierce WS (1980) The use of segmented polyurethanes in ventricular assist devices and artificial hearts. In: Szycher M, Robinson W (eds) Synthetic biomedical polymers; concepts and practice. Technomic, Westport, pp 39–57

    Google Scholar 

  13. Snyder A, Rosenberg G, Landis D (1985) Indirect estimation of circulatory pressures for control of an electric motor-driven total artificial heart. In: Langrana NA (ed) 1985 advances in bioengineering. The American society of mechanical engineers, New York, pp 87–88

    Google Scholar 

  14. Snyder AJ, Rosenberg G, Landis DL, Weiss W, Pierce WS (1984) Introductory lecture on control. Presented at the second world symposium on the artificial heart, July 3, Berlin, pp 167–190

    Google Scholar 

  15. Magovern JA, Rosenberg G, Pierce WS (1986) The development and current status of a total artificial heart. Artif Organs 10(5):357–363

    Article  PubMed  CAS  Google Scholar 

Download references

Authors

Editor information

Editors and Affiliations

Rights and permissions

Reprints and permissions

Copyright information

© 1989 Springer-Verlag Berlin Heidelberg

About this paper

Cite this paper

Rosenberg, G. et al. (1989). In Vivo Testing of a Roller-Screw Type Electric Total Artificial Heart. In: Unger, F. (eds) Assisted Circulation 3. Springer, Berlin, Heidelberg. https://doi.org/10.1007/978-3-642-74404-4_34

Download citation

  • DOI: https://doi.org/10.1007/978-3-642-74404-4_34

  • Publisher Name: Springer, Berlin, Heidelberg

  • Print ISBN: 978-3-642-74406-8

  • Online ISBN: 978-3-642-74404-4

  • eBook Packages: Springer Book Archive

Publish with us

Policies and ethics