Dependence of Electromagnetic Flow Meter Sensitivity upon Cell Distribution and Orientation

  • Conference paper
Theoretical and Clinical Hemorheology
  • 89 Accesses

Abstract

Electromagnetic flowmeters are commonly used for measuring blood flow and there have been varying reports of their dependence upon erythrocyte concentration. We have shown recently, in carefully controlled experiments, that flowmeter sensitivity is indeed changed by haematocrit (DENNIS and WYATT, 1969). We have found that when the flow is turbulent the sensitivity is hardly affected, but with laminar flow the sensitivity is reduced with increasing haematocrit. A typical result is shown in Fig. 1, where we see that in this particular case, with a flow of 250 ml min-1 in a 6 mm dia. tube, the sensitivity falls by 4% as the haematocrit is increased from 0 to 66%.

This is a preview of subscription content, log in via an institution to check access.

Access this chapter

Chapter
USD 29.95
Price excludes VAT (USA)
  • Available as PDF
  • Read on any device
  • Instant download
  • Own it forever
eBook
USD 84.99
Price excludes VAT (USA)
  • Available as PDF
  • Read on any device
  • Instant download
  • Own it forever
Softcover Book
USD 109.99
Price excludes VAT (USA)
  • Compact, lightweight edition
  • Dispatched in 3 to 5 business days
  • Free shipping worldwide - see info

Tax calculation will be finalised at checkout

Purchases are for personal use only

Institutional subscriptions

Preview

Unable to display preview. Download preview PDF.

Unable to display preview. Download preview PDF.

References

  • DENNIS, J., WYATT, D.G.: Circulation Res. 24, 875 (1969).

    PubMed  CAS  Google Scholar 

  • SHERCLIFF, J.A.: The Theory of Electromagnetic Flow Measurement, Cambridge: University Press (1962).

    Google Scholar 

  • VELICK, S.E., GORIN, H.G.: J. Gen. Physiol. 23, 753 (1939).

    Article  Google Scholar 

Download references

Author information

Authors and Affiliations

Authors

Editor information

Editors and Affiliations

Rights and permissions

Reprints and permissions

Copyright information

© 1971 Springer-Verlag Berlin · Heidelberg

About this paper

Cite this paper

Wyatt, D.G. (1971). Dependence of Electromagnetic Flow Meter Sensitivity upon Cell Distribution and Orientation. In: Hartert, H.H., Copley, A.L. (eds) Theoretical and Clinical Hemorheology. Springer, Berlin, Heidelberg. https://doi.org/10.1007/978-3-642-65156-4_37

Download citation

  • DOI: https://doi.org/10.1007/978-3-642-65156-4_37

  • Publisher Name: Springer, Berlin, Heidelberg

  • Print ISBN: 978-3-642-65158-8

  • Online ISBN: 978-3-642-65156-4

  • eBook Packages: Springer Book Archive

Publish with us

Policies and ethics